 'Aliasing Artifact' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Aliasing Artifact' found in 1 term [] and 10 definitions [], (+ 4 Boolean[] results
previous     11 - 15 (of 15)     
Result Pages : [1]  [2 3]
MRI Resources 
Online Books - Devices - Cochlear Implant - Movies - Image Quality - MRI Accidents
 
Wrap Around ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
A wrap around artifact (also called backfolding artifact or aliasing artifact) is produced by inadequate sampling or digitization. Wrap around artifacts result from digitizing fewer than two samples per period in a periodic function. Aliasing can occur in MR imaging whenever the scanned area extends beyond the field of view. These areas extending beyond the field of view boundaries are aliased back into the image to appear at artifactual locations.
 
Images, Movies, Sliders:
 Breast MRI Transverse T1 Pre Contrast 001  Open this link in a new window
    
 
spacer
 
• Related Searches:
    • Field of View
    • Foldover Suppression
    • Foldover Artifact
    • Oversampling
    • Phase Encoding
 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Aliasing Artifact' was also found in the following service: 
spacer
Ultrasound  (4) Open this link in a new window
Sensitivity EncodingInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(SENSE) A MRI technique for relevant scan time reduction. The spatial information related to the coils of a receiver array are utilized for reducing conventional Fourier encoding. In principle, SENSE can be applied to any imaging sequence and k-space trajectories. However, it is particularly feasible for Cartesian sampling schemes. In 2D Fourier imaging with common Cartesian sampling of k-space sensitivity encoding by means of a receiver array enables to reduce the number of Fourier encoding steps.
SENSE reconstruction without artifacts relies on accurate knowledge of the individual coil sensitivities. For sensitivity assessment, low-resolution, fully Fourier-encoded reference images are required, obtained with each array element and with a body coil.
The major negative point of parallel imaging techniques is that they diminish SNR in proportion to the numbers of reduction factors. R is the factor by which the number of k-space samples is reduced. In standard Fourier imaging reducing the sampling density results in the reduction of the FOV, causing aliasing. In fact, SENSE reconstruction in the Cartesian case is efficiently performed by first creating one such aliased image for each array element using discrete Fourier transformation (DFT).
The next step then is to create a full-FOV image from the set of intermediate images. To achieve this one must undo the signal superposition underlying the fold-over effect. That is, for each pixel in the reduced FOV the signal contributions from a number of positions in the full FOV need to be separated. These positions form a Cartesian grid corresponding to the size of the reduced FOV.
The advantages are especially true for contrast-enhanced MR imaging such as dynamic liver MRI (liver imaging) , 3 dimensional magnetic resonance angiography (3D MRA), and magnetic resonance cholangiopancreaticography (MRCP).
The excellent scan speed of SENSE allows for acquisition of two separate sets of hepatic MR images within the time regarded as the hepatic arterial-phase (double arterial-phase technique) as well as that of multidetector CT.
SENSE can also increase the time efficiency of spatial signal encoding in 3D MRA. With SENSE, even ultrafast (sub second) 4D MRA can be realized.
For MRCP acquisition, high-resolution 3D MRCP images can be constantly provided by SENSE. This is because SENSE resolves the presence of the severe motion artifacts due to longer acquisition time. Longer acquisition time, which results in diminishing image quality, is the greatest problem for 3D MRCP imaging.
In addition, SENSE reduces the train of gradient echoes in combination with a faster k-space traversal per unit time, thereby dramatically improving the image quality of single shot echo planar imaging (i.e. T2 weighted, diffusion weighted imaging).
spacer

• View the DATABASE results for 'Sensitivity Encoding' (12).Open this link in a new window

 
Further Reading:
  News & More:
Image Characteristics and Quality
   by www.sprawls.org    
MRI Resources 
Anatomy - Claustrophobia - Examinations - Case Studies - Non-English - MRI Technician and Technologist Schools
 
Signal to Noise Ratio
 
(SNR or S/N) The signal to noise ratio is used in MRI to describe the relative contributions to a detected signal of the true signal and random superimposed signals ('background noise') - a criterion for image quality.
One common method to increase the SNR is to average several measurements of the signal, on the expectation that random contributions will tend to cancel out. The SNR can also be improved by sampling larger volumes (increasing the field of view and slice thickness with a corresponding loss of spatial resolution) or, within limits, by increasing the strength of the magnetic field used. Surface coils can also be used to improve local signal intensity. The SNR will depend, in part, on the electrical properties of the sample or patient being studied. The SNR increases in proportion to voxel volume (1/resolution), the square root of the number of acquisitions (NEX), and the square root of the number of scans (phase encodings). SNR decreases with the field of view squared (FOV2) and wider bandwidths. See also Signal Intensity and Spin Density.

Measuring SNR:
Record the mean value of a small ROI placed in the most homogeneous area of tissue with high signal intensity (e.g. white matter in thalamus). Calculate the standard deviation for the largest possible ROI placed outside the object in the image background (avoid ghosting/aliasing or eye movement artifact regions).
The SNR is then:
Mean Signal/Standard Deviation of Background Noise
 
Images, Movies, Sliders:
 Brain MRI Images T1  Open this link in a new window
      

 
spacer

• View the DATABASE results for 'Signal to Noise Ratio' (48).Open this link in a new window


• View the NEWS results for 'Signal to Noise Ratio' (2).Open this link in a new window.
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
  News & More:
Picture-Perfect Particles Enhance MRI Signal
Friday, 13 April 2012   by cen.acs.org    
Researchers design 'intelligent' metamaterial to make MRIs affordable and accessible
Tuesday, 5 November 2019   by phys.org    
Metamaterials boost sensitivity of MRI machines
Thursday, 14 January 2016   by www.eurekalert.org    
Optimizing Musculoskeletal MR
   by rad.usuhs.mil    
MRI Resources 
Libraries - Cardiovascular Imaging - Pathology - Nerve Stimulator - Coils - Stent
 
Moire Fringes (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Moire fringes, moire
DESCRIPTION
Superimposed signals of different phases
REASON
Interferences
HELP
Surface coil, shimming
A moiré pattern is an interference pattern created for example when two grids are overlaid at an angle, or when they have slightly different mesh sizes. The human visual system creates an imaginary pattern of roughly horizontal dark and light bands, the moiré pattern that appears to be superimposed on the lines.
In MRI, the appearance of moiré fringes can be caused by a variety of reasons e.g., inhomogeneity of the main magnetic field caused by a defect shielding (interference with RF pulses), interferences produced by aliasing, and interferences of echoes from different excitation modes (with different echo times).
mri safety guidance
Image Guidance
Take spin echo-based techniques, or a surface coil. This artifact is often sensitive to shimming or susceptibility gradients.
spacer
 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Moiré pattern
   by en.wikipedia.org    
Moire Fringes
   by www.mritutor.org    
Searchterm 'Aliasing Artifact' was also found in the following service: 
spacer
Ultrasound  (4) Open this link in a new window
Undersampling
 
Undersampling is the decrease in data to increase image acquisition speed (shorter scan times without loss of quality - increased productivity - reduced cost of equipment). There are different strategies to decrease data without losing quality (e.g. reduction of the FOV in one or more spatial directions - RFOV).
Reduction in data normally is associated with an increase in aliasing (degradation of the SNR through backfolding of the entire noise spectrum), or with other artifact caused by missing data, which results in fine lines.
spacer

• View the DATABASE results for 'Undersampling' (6).Open this link in a new window

MRI Resources 
MR Guided Interventions - Sequences - Pediatric and Fetal MRI - Cardiovascular Imaging - RIS - Stent
 
previous      11 - 15 (of 15)     
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]