Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Gradient Echo' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Gradient Echo' found in 20 terms [] and 114 definitions []
previous     96 - 100 (of 134)     next
Result Pages : [1 2 3 4]  [5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
Searchterm 'Gradient Echo' was also found in the following services: 
spacer
News  (1)  Resources  (2)  Forum  (6)  
 
Flip Angle
 
(FA) The flip angle a is used to define the angle of excitation for a field echo pulse sequence. It is the angle to which the net magnetization is rotated or tipped relative to the main magnetic field direction via the application of a RF excitation pulse at the Larmor frequency. It is also referred to as the tip angle, nutation angle or angle of nutation.
The radio frequency power (which is proportional to the square of the amplitude) of the pulse is proportional to a through which the spins are tilted under its influence. Flip angles between 0° and 90° are typically used in gradient echo sequences, 90° and a series of 180° pulses in spin echo sequences and an initial 180° pulse followed by a 90° and a 180° pulse in inversion recovery sequences.
spacer
 
• Related Searches:
    • Abdominal Imaging
    • Echo
    • In Phase
    • Out of Phase
    • Net Magnetization Vector
 
Further Reading:
  Basics:
What MRI Sequences Produce the Highest Specific Absorption Rate (SAR), and Is There Something We Should Be Doing to Reduce the SAR During Standard Examinations?
Thursday, 16 April 2015   by www.ajronline.org    
Mapping of low flip angles in magnetic resonance(.pdf)
Saturday, 1 January 2011   by www.hal.inserm.fr    
  News & More:
A practical guideline for T1 reconstruction from various flip angles in MRI
Saturday, 1 October 2016   by journals.sagepub.com    
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Mobile MRI Rental - Hospitals - Image Quality - Mass Spectrometry - Education - Libraries
 
Flow Related Enhancement
 
(FRE) Flow related enhancement could be seen most for blood flow, but also for other liquids with some MR imaging techniques, as an increase in intensity due to the washout of saturated spins. FRE provides positive contrast ("bright blood") of vascular details in time of flight MRA as well as the physiologic characterization of blood flow.
If stationary spins within the scanned region experience only an incomplete T1 relaxation between the repeated radio frequency (RF) excitations, this results in fewer signal of the stationary tissue (compared to inflowing blood with completely relaxed spins). The degree of the flow related enhancement is proportional to the blood flow velocity and the used repetition time. The use of flow compensation (gradient moment nulling) improves the FRE especially in gradient echo sequences.
 
Images, Movies, Sliders:
 TOF-MRA Circle of Willis Inverted MIP  Open this link in a new window
    

 Circle of Willis, Time of Flight, MIP  Open this link in a new window
    
SlidersSliders Overview

 
spacer

• View the DATABASE results for 'Flow Related Enhancement' (10).Open this link in a new window

 
Further Reading:
  Basics:
Conventional MRI and MR Angiography of Stroke
2012   by www.mc.vanderbilt.edu    
MRI Resources 
Chemistry - MRI Accidents - Used and Refurbished MRI Equipment - Jobs - Absorption and Emission - Bioinformatics
 
Flow VoidForum -
related threads
 
The occurrence of a low signal in regions of flow. For a spin echo sequence, this is caused in part by a lack of refocusing of blood, which is excited by the 90° pulse but not by the 180° pulse. For a gradient echo sequence, this is caused by the dephasing of blood signal.
spacer

• View the DATABASE results for 'Flow Void' (5).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifacts, Flow void and Signal void
   by www.neuroradiologycases.com    
Flow comp off: An easy technique to confirm CSF flow within syrinx and aqueduct
Wednesday, 2 January 2013   by medind.nic.in    
DRIVE: shorter scan time, brighter fluid
   by clinical.netforum.healthcare.philips.com    
Searchterm 'Gradient Echo' was also found in the following services: 
spacer
News  (1)  Resources  (2)  Forum  (6)  
 
Gradient Field EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
spacer

• View the DATABASE results for 'Gradient Field Echo' (4).Open this link in a new window

 
Further Reading:
  News & More:
Principles, Techniques, and Applications of T2*- based MR Imaging and Its Special Applications1
September 2009   by pubs.rsna.org    
MRI Resources 
Directories - Education pool - Education - Pathology - Bioinformatics - Calculation
 
Gradient Moment Nulling
 
Gradient moment nulling used as motion artifact suppression technique (MAST) reduces constant velocity motion distortion in standard spin echo or gradient echo pulse sequences. It is an adjustment to zero at the echo time (TE) of the net moments of the amplitude of the waveform of the magnetic field gradients with time. The zeroth moment is the area under the curve. The first moment is the 'center of gravity' etc.
The aim is to minimize the phase shifts acquired by the transverse magnetization of excited nuclei moving along the gradients (including the effect of refocusing radio frequency pulses), particularly for the reduction of image artifacts due to motion.
spacer

• View the DATABASE results for 'Gradient Moment Nulling' (7).Open this link in a new window

 
Further Reading:
  Basics:
Motion Compensation in MR Imaging
   by ccn.ucla.edu    
MRI Resources 
Equipment - Knee MRI - Nerve Stimulator - MRI Training Courses - Pathology - Liver Imaging
 
previous      96 - 100 (of 134)     next
Result Pages : [1 2 3 4]  [5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 20 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]