Spatial Frequency
 
Measure of resolution (lp/mm or lp/cm).
A dimension of the Fourier transformation space (k-space representation of an image), having units of inverse distance. Higher values of spatial frequencies correspond to finer detail in the image.
spacer
MRI Resources 
Manufacturers - Movies - Databases - Quality Advice - IR - MRI Reimbursement
 
Time Resolved Imaging of Contrast KineticsInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(TRICKS) Time resolved imaging of contrast kinetics is a MRI technique, which increases the temporal resolution of dynamic contrast enhanced magnetic resonance angiography (CE-MRA) sequences. The K-space is divided into regions by increasing the sampling rate at the lower spatial frequencies and by reducing the sampling rate at the higher spatial frequencies. Since the time duration between two frames is shortened, it can be observed how frequently and how quickly the images are repeated at the exact same location.
TRICKS is particularly useful for dynamic vascular studies with high temporal resolution. TRICKS improves the calculation of the contrast bolus arrival and improves the characterization of arterio-venous malformations (AVMs).

See also Automatic Bolus Detection, MRA, Cardiac MRI.
spacer

• View the DATABASE results for 'Time Resolved Imaging of Contrast Kinetics' (2).Open this link in a new window

 
Further Reading:
  Basics:
Optimal k-Space Sampling for Dynamic Contrast-Enhanced MRI with an Application to MR Renography
Thursday, 5 November 2009   by www.ncbi.nlm.nih.gov    
MRI Resources 
Services and Supplies - Abdominal Imaging - Pediatric and Fetal MRI - Distributors - Service and Support - Liver Imaging
 
Turbo Field EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(TFE) Turbo field echo is a gradient echo pulse sequence with data acquisition after an initial 180° (similar to IR) preparation pulse for contrast enhancement. The difference between a FFE and TFE other than the speed of the sequence is that the image is acquired while approaching steady state (the echoes are collected during the time in which the tissues are experiencing T1 relaxation).
The contrast is prepared one time, which means the contrast is changing while the echoes are collected and can be manipulated by selecting the type and timing of the prepulse. A delay time is given before the actual image acquisition. To achieve T1 contrast the 180° prepulse is followed by an operator selected delay time, that results in no signal from the targeted tissue. So when the echoes are acquired, no signal is present, additional RF spoiling is performed to optimize for T1 contrast. The delay chosen corresponds to when T1 relaxation reaches and suppresses T1 signal or optimizes the difference between tissues. Contrast for these sequences are enhanced when K-space is filled using a centric or low-high ordering. A TFE can be acquired with a 2D or 3D technique and with or without T1, T2 weighting.
See Ultrafast Gradient Echo Sequence, TurboFLASH and Magnetization Prepared Rapid Gradient Echo (MPRAGE).
spacer

• View the DATABASE results for 'Turbo Field Echo' (6).Open this link in a new window

 
Further Reading:
  Basics:
Sequence for Philips(.pdf)
   by www.droid.cuhk.edu.hk    
Pediatric and Adult Cochlear Implantation1
2003   by radiographics.rsnajnls.org    
Searchterm 'K-Space' was also found in the following services: 
spacer
News  (1)  Resources  (2)  Forum  (10)  
 
Ultrafast Gradient Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Ultrafast Gradient Echo Sequence Timing Diagram In simple ultrafast GRE imaging, TR and TE are so short, that tissues have a poor imaging signal and - more importantly - poor contrast except when contrast media enhanced (contrast enhanced angiography). Therefore, the magnetization is 'prepared' during the preparation module, most frequently by an initial 180° inversion pulse.
In the pulse sequence timing diagram, the basic ultrafast gradient echo sequence is illustrated. The 180° inversion pulse is executed one time (to the left of the vertical line), the right side represents the data collection period and is often repeated depending on the acquisition parameters.
See also Pulse Sequence Timing Diagram, there you will find a description of the components.
Ultrafast GRE sequences have a short TR,TE, a low flip angle and TR is so short that image acquisition lasts less than 1 second and typically less than 500 ms. Common TR: 3-5 msec, TE: 2 msec, and the flip angle is about 5°. Such sequences are often labeled with the prefix 'Turbo' like TurboFLASH, TurboFFE and TurboGRASS.
This allows one to center the subsequent ultrafast GRE data acquisition around the inversion time TI, where one of the tissues of interest has very little signal as its z-magnetization is passing through zero.
Unlike a standard inversion recovery (IR) sequence, all lines or a substantial segment of k-space image lines are acquired after a single inversion pulse, which can then together be considered as readout module. The readout module may use a variable flip angle approach, or the data acquisition may be divided into multiple segments (shots). The latter is useful particularly in cardiac imaging where acquiring all lines in a single segment may take too long relative to the cardiac cycle to provide adequate temporal resolution.
If multiple lines are acquired after a single pulse, the pulse sequence is a type of gradient echo echo planar imaging (EPI) pulse sequence.

See also Magnetization Prepared Rapid Gradient Echo (MPRAGE) and Turbo Field Echo (TFE).
spacer

• View the DATABASE results for 'Ultrafast Gradient Echo Sequence' (13).Open this link in a new window

MRI Resources 
Portals - Quality Advice - NMR - Health - Breast Implant - Absorption and Emission
 
ViewForum -
related threads
 
The ensemble of raw data points collected during the signal readout. For example, during Cartesian sampling, normally used to refer to data collected with a fixed value of the phase encoding gradient. Also referred to as one line of k-space. In projection reconstruction, the line is radial oriented, while in spiral imaging, it is a spiral.
spacer

• View the DATABASE results for 'View' (160).Open this link in a new window


• View the NEWS results for 'View' (20).Open this link in a new window.
 
Further Reading:
  Basics:
Image Characteristics and Quality
   by www.sprawls.org    
MRI Resources 
Anatomy - Diffusion Weighted Imaging - Resources - Functional MRI - Research Labs - Knee MRI
 
previous      56 - 60 (of 62)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10 11 12 13]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]