5 Gauss Line
 
This line specifies the perimeter around a MR scanner within which the static magnetic fields are higher than five gauss. Five gauss and below are considered 'safe' levels of static magnetic field exposure for the general public.
Portable devices requiring a separation distance between the device and the MR magnet, should not be considered 'MR Safe', 'MR Compatible', or intended for use in the MR environment. Typically the 5 gauss line is the only location where the static magnetic field strength is specified around a MR scanner. Therefore, labeling specifying a separation distance between the MR magnet and the device to ensure safe or proper operation of the device should be avoided.

See also the related poll result: 'In 2010 your scanner will probably work with a field strength of'
spacer
• For this and other aspects of MRI safety see our InfoSheet about MRI Safety.
• Patient-related information is collected in our MRI Patient Information.

 
• Share the entry '5 Gauss Line':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Magnetic Fringe Field
    • Permanent Magnet
    • 5 Gauss Fringe Field
    • 5 Gauss Limit
    • MRI Safety
 
Further Reading:
  News & More:
Boston Scientific and Biophan in MRI Collaboration
Friday, 21 November 2003   by www.medimaging.net    
World's First Portable MRI Cleared by FDA
Monday, 17 February 2020   by www.medgadget.com    
MRI Safety Resources 
Nerve Stimulator - Claustrophobia - Stimulator pool - Implant and Prosthesis pool - Guidance
 
ContraindicationsForum -
related threadsMRI Resource Directory:
 - Safety -
 
The principal contraindications of the MRI procedure are mostly related to the presence of metallic implants in a patient. The risks of MRI scans increase with the used field strength. In general, implants are becoming increasingly MR safe and an individual evaluation is carried out for each case.
mri safety guidance
MRI Safety Guidance
Some patients should not be examined in MRI machines, or come closer than the 5 Gauss line to the system.

Absolute Contraindications for the MRI scan:
•
electronically, magnetically, and mechanically activated implants
•
ferromagnetic or electronically operated active devices like automatic cardioverter defibrillators
•
metallic splinters in the eye
•
ferromagnetic haemostatic clips in the central nervous system (CNS)

Patients with absolute contraindications should not be examined or only with special MRI safety precautions. Patients with an implanted cardiac pacemaker have been scanned on rare occasions, but pacemakers are generally considered an absolute contraindication. Relative contraindications may pose a relative hazard, and the type and location of an implant should be assessed prior to the MRI examination.

Relative Contraindications for the MRI scan:
•
other pacemakers, e.g. for the carotid sinus
•
lead wires or similar wires (MRI Safety risk)
•
prosthetic heart valves (in high fields, if dehiscence is suspected)
•
haemostatic clips (body)
•

Osteosynthesis material is usually anchored so well in the patients that no untoward effect will result. Another effect on metal parts in the patient's body is the heating of these parts through induction. In addition, image quality may be severely degraded. The presence of other metallic implants such as surgical clips etc. should be made known to the MRI operators. Most of these materials are non-magnetic, but if magnetic, they can pose a hazard.

See also MRI safety, Pregnancy, Claustrophobia and Tattoos.
Radiology-tip.comradRadiation Safety,  As Low As Reasonably Achievable
spacer
Medical-Ultrasound-Imaging.comUltrasound Safety
spacer

• View the DATABASE results for 'Contraindications' (11).Open this link in a new window

 
Further Reading:
  Basics:
MRI in Patients with Implanted Devices: Current Controversies
Monday, 1 August 2016   by www.acc.org    
Newer Heart Devices Safe During MRI
Monday, 23 August 2004   by www.hospimedica.com    
Physics of MRI Safety
   by www.aapm.org    
FDA Releases New Guidance On Establishing Safety, Compatibility Of Passive Implants In MR Environments
Tuesday, 16 December 2014   by www.meddeviceonline.com    
  News & More:
Women with permanent make-up tattoos suffer horrific facial burns after going in for MRI scans - which create an electric current in the ink
Monday, 4 July 2016   by www.dailymail.co.uk    
Positive diagnosis for neural therapeutic implants
Tuesday, 19 April 2016   by medicalxpress.com    
Codman Neuro develops new MRI-resistant programmable valve for treatment of hydrocephalus
Tuesday, 22 September 2015   by www.news-medical.net    
MRI Resources 
Image Quality - Jobs pool - Jobs - Pediatric and Fetal MRI - MR Guided Interventions - Online Books
 
Insulin PumpMRI Resource Directory:
 - Safety -
 
mri safety guidance
MRI Safety Guidance
Not necessarily a contraindication, but the examination may damage or impair it. An insulin pump can be disconnected easily for a period of time. Remove the pump while outside the 5 Gauss line, because the pump batteries and motor are magnetic.
spacer

• View the DATABASE results for 'Insulin Pump' (3).Open this link in a new window

MRI Resources 
Examinations - MRI Accidents - Distributors - Corporations - RIS - MRI Technician and Technologist Career
 
MR Guided InterventionsMRI Resource Directory:
 - MR Guided Interventions -
 
Possible MR guided interventions are breast MRI, liver and bone biopsies. Open MRI is advantageous to guide interventional procedures because this type of MRI scanner provides a better patient access.
Before the MRI guided biopsy is started, all the necessary precautions ensuring normal sterile and safe conditions for an intervention have to be performed. This includes the special safety issues concerning MRI surroundings and special equipment, like non-magnetic biopsy needles, localization tools (e.g. wires to mark lesions prior to surgery with breast MRI guidance) etc.

See also 5 Gauss Line, Contraindications, Low Field MRI, MR Compatibility, Magnetic Resonance Guided Focused Ultrasound and Computer Aided Detection.
spacer

• View the DATABASE results for 'MR Guided Interventions' (8).Open this link in a new window

 
Further Reading:
  Basics:
NeuroArm Robot Promises to Revolutionise MRI-guided Neurosurgery
Wednesday, 7 November 2007   by www.medgadget.com    
Magnetic resonance imaging guided musculoskeletal interventions
at 0.23T
2002
Physics of MRI Safety
   by www.aapm.org    
  News & More:
AI analysis finds younger AFib patients benefit from MRI-guided ablation treatments
Friday, 25 August 2023   by www.eurekalert.org    
Theranostic nano-platform for MRI-guided synergistic therapy against breast cancer
Monday, 26 September 2022   by phys.org    
Magnetic seeds used to heat and kill cancer
Tuesday, 1 February 2022   by www.sciencedaily.com    
What is the effect of MRI with targeted biopsies on the rate of patients discontinuing active surveillance? A reflection of the use of MRI in the PRIAS study
Thursday, 8 April 2021   by www.docwirenews.com    
Modeling of Active Shimming of Metallic Needles for Interventional MRI
Monday, 29 June 2020   by pubmed.ncbi.nlm.nih.gov    
Magnetic Resonance Imaging Guided Confirmatory Biopsy for Initiating Active Surveillance of Prostate Cancer
Wednesday, 11 September 2019   by jamanetwork.com    
FDA clears ViewRay's next-gen, MRI-guided radiation therapy device
Tuesday, 28 February 2017   by www.fiercebiotech.com    
Siemens, U. of Twente Biopsy Robot Promises Greater Precision, Less Cost
Friday, 22 January 2016   by www.meddeviceonline.com    
Magnetic resonance-guided motorized transcranial ultrasound system for blood-brain barrier permeabilization along arbitrary trajectories in rodents
Thursday, 24 December 2015   by www.ncbi.nlm.nih.gov    
New MRI-Guided Catheter Shows Major Potential for Stroke Treatment
Tuesday, 29 December 2015   by www.radiology.ucsf.edu    
Polish study on MRI-ultrasound for targeted prostate biopsy wins CEM award
Tuesday, 12 November 2013   by medicalxpress.com    
C4 Imaging Announces FDA 510(k) Clearance of its Positive-Signal MRI Marker - Siriusâ„¢
Friday, 6 December 2013   by www.digitaljournal.com    
MRI Resources 
Pediatric and Fetal MRI - Societies - Quality Advice - Functional MRI - Calculation - Databases
 
MRI SafetyMRI Resource Directory:
 - Safety -
 
There are different types of contraindications that would prevent a person from being examined with an MRI scanner. MRI systems use strong magnetic fields that attract any ferromagnetic objects with enormous force. Caused by the potential risk of heating, produced from the radio frequency pulses during the MRI procedure, metallic objects like wires, foreign bodies and other implants needs to be checked for compatibility. High field MRI requires particular safety precautions. In addition, any device or MRI equipment that enters the magnet room has to be MR compatible. MRI examinations are safe and harmless, if these MRI risks are observed and regulations are followed.

Safety concerns in magnetic resonance imaging include:
•
the magnetic field strength;
•
possible 'missile effects' caused by magnetic forces;
•
the potential for heating of body tissue due to the application of the radio frequency energy;
•
the effects on implanted active devices such as cardiac pacemakers or insulin pumps;
•
magnetic torque effects on indwelling metal (clips, etc.);
•
the audible acoustic noise;
•
danger due to cryogenic liquids;
•
the application of contrast medium;
mri safety guidance
MRI Safety Guidance
It is important to remember when working around a superconducting magnet that the magnetic field is always on. Under usual working conditions the field is never turned off. Attention must be paid to keep all ferromagnetic items at an adequate distance from the magnet. Ferromagnetic objects which came accidentally under the influence of these strong magnets can injure or kill individuals in or nearby the magnet, or can seriously damage every hardware, the magnet itself, the cooling system, etc.. See MRI resources Accidents.
The doors leading to a magnet room should be closed at all times except when entering or exiting the room. Every person working in or entering the magnet room or adjacent rooms with a magnetic field has to be instructed about the dangers. This should include the patient, intensive-care staff, and maintenance-, service- and cleaning personnel, etc..
The 5 Gauss limit defines the 'safe' level of static magnetic field exposure. The value of the absorbed dose is fixed by the authorities to avoid heating of the patient's tissue and is defined by the specific absorption rate. Leads or wires that are used in the magnet bore during imaging procedures, should not form large-radius wire loops. Leg-to-leg and leg-to-arm skin contact should be prevented in order to avoid the risk of burning due to the generation of high current loops if the legs or arms are allowed to touch. The patient's skin should not be in contact with the inner bore of the magnet.
The outflow from cryogens like liquid helium is improbable during normal operation and not a real danger for patients.
The safety of MRI contrast agents is tested in drug trials and they have a high compatibility with very few side effects. The variations of the side effects and possible contraindications are similar to X-ray contrast medium, but very rare. In general, an adverse reaction increases with the quantity of the MRI contrast medium and also with the osmolarity of the compound.

See also 5 Gauss Fringe Field, 5 Gauss Line, Cardiac Risks, Cardiac Stent, dB/dt, Legal Requirements, Low Field MRI, Magnetohydrodynamic Effect, MR Compatibility, MR Guided Interventions, Claustrophobia, MRI Risks and Shielding.
Radiology-tip.comradRadiation Safety,  Ionizing Radiation
spacer
Medical-Ultrasound-Imaging.comUltrasound Safety,  Absorbed Dose
spacer

• View the DATABASE results for 'MRI Safety' (42).Open this link in a new window


• View the NEWS results for 'MRI Safety' (13).Open this link in a new window.
 
Further Reading:
  Basics:
MRI Safety
2001   by www.fda.gov    
What MRI Sequences Produce the Highest Specific Absorption Rate (SAR), and Is There Something We Should Be Doing to Reduce the SAR During Standard Examinations?
Thursday, 16 April 2015   by www.ajronline.org    
Contrast Agents: Safety Profile
   by www.clinical-mri.com    
  News & More:
How safe is 7T MRI for patients with neurosurgical implants?
Thursday, 17 November 2022   by healthimaging.com    
Newer Heart Devices Safe During MRI
Monday, 23 August 2004   by www.hospimedica.com    
Study: Face Masks Unsafe in MRI Machines
Wednesday, 13 July 2022   by www.laboratoryequipment.com    
COVID-19: Attention shifts to MRI infection control
Thursday, 9 July 2020   by https://www.auntminnieeurope.com/index.aspx?sec=ser§sub=def§pag=dis§ItemID=619012    
FDA Releases New Guidance On Establishing Safety, Compatibility Of Passive Implants In MR Environments
Tuesday, 16 December 2014   by www.meddeviceonline.com    
Modern Implantable Heart Devices Safe For Use In MRI Scans
Wednesday, 16 March 2005   by www.sciencedaily.com    
MRI Resources 
Intraoperative MRI - Shielding - Blood Flow Imaging - Chemistry - Contrast Enhanced MRI - Universities
 
     1 - 5 (of 18)     next
Result Pages : [1]  [2]  [3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]