cardiac and respiratory synchronization
Movement of the heart causes blurring and ghosting in the images. The artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
These artifacts can be reduced by using cardiac synchronization: triggering, gating or retrospective triggering. Maximum reduction can be achieved by using triggering in combination with flow compensation, respiratory triggering or breath hold and regional saturation techniques.
See also Motion Artifact. | | | | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Quick Overview Please note that there are different common names for this artifact.
This artifact appears as a bright spot (or zipper line) in the image center. Central point artifacts are caused by a DC offset in the hardware. MRI scanners normally offer a software compensation (DC correction, baseline correction) for prevention.
Image Guidance
Take care for a constant temperature. If the problem increases or keeps on existing, it should be addressed to the service. | | | | • View the DATABASE results for 'Central Point Artifact' (4).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Quick Overview
NAME
Cerebro spinal fluid pulsation
Pulsatile cerebro spinal fluid flow produces ghost artifacts that are superimposed in the image.
Image Guidance
| | | | • View the DATABASE results for 'Cerebro Spinal Fluid Pulsation Artifact' (3).
| | | | |
| | | Searchterm 'Image Guidance' was also found in the following service: | | | | |
| | |
| |
|
Quick Overview Please note that there are different common names for this artifact.
DESCRIPTION
Black or bright band
During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
Image Guidance
| | | | • View the DATABASE results for 'Chemical Shift Artifact' (7).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Quick Overview
DESCRIPTION
Incorrectly interpretation or display
HELP
Save to a new or different media
An artifact on a MR image can appear when data read off an optical disc became corrupted. There two effects can be seen.
Firstly, Bands where the intensity has been incorrectly interpreted or secondly areas, where the horizontal position of the pixels has been incorrectly displayed. It is important to be able to differentiate between artifacts caused during a MRI scan, from those caused by the associated hardware of an imaging system.
Image Guidance
Review the hardware and store the images on a new disc. | | | | | Further Reading: | Basics:
|
|
| |
| | | | |
| | | |
|
| |