The number of data points collected in one, two or all three directions. Normally used for the 2D in plane sampling. The display matrix may be different from the acquisition matrix, although the latter determines the resolution. Measurement time may be saved by not acquiring raw data lines corresponding to high resolution. Not measured rows are filled with zeroes prior to the image calculation. A square image is the result of an interpolation in phase encoding direction.
See also Zero Filling.
Image Guidance
| |  | | | | | | | |
|  | |  |  |  |
| |
|
Quick Overview
NAME
Metal, susceptibility
Ferromagnetic metal will cause a magnetic field inhomogeneity, which in turn causes a local signal void, often accompanied by an area of high signal intensity, as well as a distortion of the image.
They create their own magnetic field and dramatically alter precession frequencies of protons in the adjacent tissues. Tissues adjacent to ferromagnetic components become influenced by the induced magnetic field of the metal hardware rather than the parent field and, therefore, either fail to precess or do so at a different frequency and hence do not generate useful signal. Two components contribute to susceptibility artifact, induced magnetism in the ferromagnetic component itself and induced magnetism in protons adjacent to the component. Artifacts from metal may have varied appearances on MRI scans due to different type of metal or configuration of the piece of metal.
The biocompatibility of metallic alloys, stainless steel, cobalt chrome and titanium alloy is based on the presence of a constituent element within the alloy that has the ability to form an adherent oxide coating that is stable, chemically inert and hence biocompatible. In relation to imaging titanium alloys are less ferromagnetic than both cobalt and stainless steel, induce less susceptibility artifact and result in less marked image degradation.
Image Guidance
| |  | | • View the DATABASE results for 'Metal Artifact' (2).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
|  | |  |  |  |
| |
|
Quick Overview
NAME
Moire fringes, moire
DESCRIPTION
Superimposed signals of different phases
A moiré pattern is an interference pattern created for example when two grids are overlaid at an angle, or when they have slightly different mesh sizes. The human visual system creates an imaginary pattern of roughly horizontal dark and light bands, the moiré pattern that appears to be superimposed on the lines.
In MRI, the appearance of moiré fringes can be caused by a variety of reasons e.g., inhomogeneity of the main magnetic field caused by a defect shielding ( interference with RF pulses), interferences produced by aliasing, and interferences of echoes from different excitation modes (with different echo times).
Image Guidance
| |  | | | Further Reading: | Basics:
|
|
| |
|  |  | Searchterm 'Image Guidance' was also found in the following service: | | | | |
|  |  |
| |
|
Quick Overview
Please note that there are different common names for this artifact.
NAME
Motion, phase encoded motion, instability, smearing
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic
Patient motion is the largest physiological effect that causes artifacts, often resulting from involuntary movements (e.g. respiration, cardiac motion and blood flow, eye movements and swallowing) and minor subject movements.
Movement of the object being imaged during the sequence results in inconsistencies in phase and amplitude, which lead to blurring and ghosting. The nature of the artifact depends on the timing of the motion with respect to the acquisition. Causes of motion artifacts can also be mechanical vibrations, cryogen boiling, large iron objects moving in the fringe field (e.g. an elevator), loose connections anywhere, pulse timing variations, as well as sample motion. These artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
| |  | | • View the DATABASE results for 'Motion Artifact' (24).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
|  | |  |  |  |
| |
|

Odin Medical Technologies, Ltd. was founded in Israel, in 1996. Odin Medical Technologies, Inc. was established in early 2000 in the United States as a wholly owned subsidiary. Odin Medical Technologies develops and manufactures intraoperative magnetic resonance imaging ( MRI) image guidance systems designed to improve neurosurgical interventions with a special focus on minimally invasive procedures.
As of July 25, 2006, Odin Medical Technologies, Ltd. was acquired by Medtronic, Inc.
MRI Scanners:
| |  | | • View the DATABASE results for 'Odin Medical Technologies, Inc.' (2).
| | | | |
|  | |  |  |
|  | | |
|
| |