Point Resolved SpectroscopyInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PRESS) Point resolved spectroscopy is a multi echo single shot technique to obtain spectral data. PRESS is a 90°-180°-180° (slice selective pulses) sequence. The 90° radio frequency pulse rotates the spins in the yx-plane, followed by the first 180° pulse (spin rotation in the xz-plane) and the second 180° pulse (spin rotation in the xy-plane), which gives the signal.
With the long echo times used in PRESS, there is a better visualization of metabolites with longer relaxation times. Many of the metabolites depicted by stimulated echo technique are not seen on point resolved spectroscopy, but PRESS is less susceptible to motion, diffusion, and quantum effects and has a better SNR than stimulated echo acquisition mode (STEAM).
spacer
 
• Share the entry 'Point Resolved Spectroscopy':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Echo Time
    • Pulse, 90°
    • Spectroscopic Imaging Techniques
    • Multi Echo Imaging
    • Pulse, 180°
 
Further Reading:
  Basics:
The Basics of MRI
   by www.cis.rit.edu    
  News & More:
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
MRI Resources 
Spine MRI - Used and Refurbished MRI Equipment - MRI Technician and Technologist Career - Implant and Prosthesis pool - Absorption and Emission - MRI Training Courses
 
Stimulated Echo Acquisition Mode
 
(STEAM) A single volume spectroscopy method. The stimulated echo acquisition mode techniques is similar to Point Resolved Spectroscopy (PRESS). The STEAM sequence use a 90°-90°-90° pulse sequence instead of 90°-180°-180° pulses.
spacer

• View the DATABASE results for 'Stimulated Echo Acquisition Mode' (2).Open this link in a new window

 
Further Reading:
  News & More:
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
MRI Resources 
Jobs - Homepages - PACS - Contrast Agents - Non-English - Diffusion Weighted Imaging
 
Volume Excitation Stimulated Echoes
 
(VEST) The 'Volume Excitation Stimulated Echoes' sequence is a method similar to point resolved spectroscopy with stimulated echoes and volume selective excitation.
spacer
MRI Resources 
RIS - MRI Technician and Technologist Schools - Spectroscopy pool - Artifacts - Coils - Mass Spectrometry
 
     1 - 3 (of 3)     
Result Pages : [1]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]