'blipped Echo Planar Imaging'
SEARCH FOR
search it again
blipped Echo Pl...
2
3
5
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
Result : Searchterm
'
blipped Echo Planar Imaging
'
found in
1
term [
] and
0
definition [
], (+
1
Boolean[
] results
1 - 2
(of 2)
Result Pages :
[
1
]
MRI Resources
Jobs pool
-
Cardiovascular Imaging
-
Service and Support
-
Shielding
-
Image Quality
-
Liver Imaging
Blipped Echo Planar Imaging
See
Echo Planar Imaging
and
Blips
.
•
Share the entry '
Blipped Echo Planar Imaging
':
Further Reading:
News & More:
New Imaging Method Makes Brain Scans 7 Times Faster
Sunday, 9 January 2011
by
www.dailytech.com
MRI Resources
MRI Physics
-
Shoulder MRI
-
IR
-
Quality Advice
-
Spine MRI
-
Hospitals
Blipped
Phase Encoding
A strategy for incrementing the position of the
k-space trajectory
of an
echo
planar
imaging
(EPI)
pulse sequence
.
Echo
planar
imaging
(EPI) uses a constant
gradient
amplitude
in one direction. This, combined with an
oscillating gradient system
in the
frequency encoding
direction, produces a zigzag trajectory in
k-space
. In the
blipped
phase encoding
variant of EPI, the
k-space
position in the
phase
encoded direction is incremented by
gradient
'blips' of the appropriate area. These, when timed to occur during the reversals of the read-out
gradient
, produce a rectilinear path in
k-space
.
The artifacts in an EPI image can arise from both
hardware
and sample imperfections. These are most easily understandable from examination of the
k-space trajectory
involved, which is either a zigzag form (when using a constant
phase encoding gradient
) or a rastered zigzag (when the
phase encoding
is performed with small gradients at the end of each scan line, so-called '
blipped
' EPI).
Further Reading:
Basics:
Chapter 2 - Principles of Magnetic Resonance Imaging
by
www.fmrib.ox.ac.uk
MRI Resources
Colonography
-
Mobile MRI Rental
-
Calculation
-
Patient Information
-
Brain MRI
-
Anatomy
1 - 2
(of 2)
Result Pages :
[
1
]
Random Page
top
Share This Page
MR-TIP
Community
User
Pass
Register (free)
Member
Services
New acceleration techniques will
:
reduce scan times
cause artifacts
increase expenses
be useful if you have a lot of experience
doesn't do much
never heard of
Contribute
Results
Look
Ups
Greeks
Symbols
Units & M.
Radiology
Ultrasound
MRI
Give MR-TIP a Feedback
Help Page
Submit a
Link/Resource
(login or register first)
MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.
Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP •
Radiology-TIP
•
Medical-Ultrasound-Imaging
•
Copyright © 2003 - 2024
SoftWays
. All rights reserved. [ 23 November 2024]
Terms of Use
|
Privacy Policy
|
Advertising
[last update: 2024-02-26 03:41:00]