Turbo Field EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(TFE) Turbo field echo is a gradient echo pulse sequence with data acquisition after an initial 180° (similar to IR) preparation pulse for contrast enhancement. The difference between a FFE and TFE other than the speed of the sequence is that the image is acquired while approaching steady state (the echoes are collected during the time in which the tissues are experiencing T1 relaxation).
The contrast is prepared one time, which means the contrast is changing while the echoes are collected and can be manipulated by selecting the type and timing of the prepulse. A delay time is given before the actual image acquisition. To achieve T1 contrast the 180° prepulse is followed by an operator selected delay time, that results in no signal from the targeted tissue. So when the echoes are acquired, no signal is present, additional RF spoiling is performed to optimize for T1 contrast. The delay chosen corresponds to when T1 relaxation reaches and suppresses T1 signal or optimizes the difference between tissues. Contrast for these sequences are enhanced when K-space is filled using a centric or low-high ordering. A TFE can be acquired with a 2D or 3D technique and with or without T1, T2 weighting.
See Ultrafast Gradient Echo Sequence, TurboFLASH and Magnetization Prepared Rapid Gradient Echo (MPRAGE).
spacer
 
• Share the entry 'Turbo Field Echo':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Fast Field Echo
    • Balanced Turbo Field Echo
    • Fast Spoiled Gradient Echo
    • Steady State Free Precession
    • Ultrafast Gradient Echo Sequence
 
Further Reading:
  Basics:
Sequence for Philips(.pdf)
   by www.droid.cuhk.edu.hk    
Pediatric and Adult Cochlear Implantation1
2003   by radiographics.rsnajnls.org    
MRI Resources 
Corporations - Non-English - Stent - Hospitals - Manufacturers - Developers
 
Hahn Echo
 
Production of spin echo by repeated RF pulses. First observed using equal (90°) RF pulses, now commonly used to describe refocusing of transverse magnetization by a 180° RF pulse. By choosing long echo delay times, the spins in a Hahn echo first dephase for a long time, then rephase, which makes the Hahn pulse sequence more susceptible to diffusion effects.
spacer

• View the DATABASE results for 'Hahn Echo' (2).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
MRI Resources 
Hospitals - PACS - Education - Musculoskeletal and Joint MRI - MRCP - Pediatric and Fetal MRI
 
Magnetization Prepared Rapid Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(MP-GRE / MPRAGE / MP-RAGE) A fast 3D gradient echo pulse sequence using a magnetization preparation pulse like TurboFLASH. Only one segment or partition of a 3D data record is obtained per inversion preparation pulse. After the acquisition, for all rows a delay time (TD) is used to prevent saturation effects.
MPRAGE is designed for rapid acquisition with T1 weighted dominance. Fast gradient echoes are characterized by their rapid sampling time, high signal intensity and image contrast while approaching steady state (the echo is collected during the time when tissues are experiencing T1 relaxation). The rapid speed of the acquisition makes it an excellent alternative to breath-hold abdominal imaging, neuro, dynamic bolus, MR angiography and cardiac imaging.

See Gradient Echo Sequence.
 
Images, Movies, Sliders:
 Brain MRI Sagittal T1 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Magnetization Prepared Rapid Gradient Echo' (3).Open this link in a new window

Searchterm 'delay time' was also found in the following service: 
spacer
News  (1)  
 
Prepulse
 
Prepulses are prior to the excitation. To optimize, this is usually as short as possible. The time from the prepulse to the center of K-space (K0) is the prepulse delay time, also known as the inversion time (TI). Prepulses are more effective if the flip angle is kept as short as possible to values such as 10 to 30 ms.
spacer

• View the DATABASE results for 'Prepulse' (5).Open this link in a new window

 
Further Reading:
  Basics:
Techniques of Fat Suppression(.pdf)
   by cds.ismrm.org    
MRI Resources 
Research Labs - Lung Imaging - Cochlear Implant - Spectroscopy - Brain MRI - Service and Support
 
Delay Alternating with Nutation for Tailored ExcitationInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(DANTE) A technique used to place a saturation band over e.g. the myocardium. This technique includes spatial modulation of magnetization complementary and delays alternating with nutations for tailored excitation, followed by the application of a cine or real-time imaging. Because the saturated magnetization pattern moves with the atoms of the tissue, the cardiac motion shows up as deformations in the grid pattern in the resulting imaging sequence.
spacer
MRI Resources 
Calculation - Spectroscopy - Brain MRI - Spine MRI - NMR - Pediatric and Fetal MRI
 
     1 - 5 (of 13)     next
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 31 January 2025]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]