Steven Ford
Mon. 7 Mar.11, 16:07
[Reply (6 of 8) to: '6-1.5T MAGNETS, DIFFERING GRADIENTS' started by: 'Elise Gough' on Wed. 23 Feb.11]

Category:
Applications and Examinations |
6-1.5T MAGNETS, DIFFERING GRADIENTS |
We maintain a lot of magnets. The leading cause of image quality problems is applications related. Nobody can possibly know all the nuances of pulse sequences by various vendors, software levels, etc. I've heard many times that a certain machine is no good, when in fact the sequences are inefficiently set up. Look there first.
Usually the techs welcome good training. If they don't want to be trained, then you have a problem, but it sounds like this issue is caused by other factors.
If there are differences in baseline quality between the machines, then compensate for that by other means. Signal starvation is easily remedied; keep the quality as consistent as you can and let time be the variable, if it comes down to that.
|
| View the whole thread | | |
Elise Gough
Mon. 7 Mar.11, 15:56
[Reply (5 of 8) to: '6-1.5T MAGNETS, DIFFERING GRADIENTS' started by: 'Elise Gough' on Wed. 23 Feb.11]

Category:
Applications and Examinations |
6-1.5T MAGNETS, DIFFERING GRADIENTS |
Thanks so much for the input. NOBODY in my group seems to have a handle on why images vary in quality from 8 different magnets. I'm being pressed to take position of imaging specialist and thru my own research we have gradient strengths from 50 slew rate up to 120. Radiologists complain neuro stinks on the 50 slew rate 1.5T magnet and it is the most pounded on magnet in our group doing hospital inpatients. So many variables including tech expertise.
Elise Gough RT(R)(CT)(MR)
|
| View the whole thread | | |
Elise Gough
Fri. 4 Mar.11, 13:56
[Reply (2 of 8) to: '6-1.5T MAGNETS, DIFFERING GRADIENTS' started by: 'Elise Gough' on Wed. 23 Feb.11]

Category:
Applications and Examinations |
6-1.5T MAGNETS, DIFFERING GRADIENTS |
Thank you for your response. Our group is strict about FOV, thickness/gap parameters being the same across all magnets. One magnet has slew rate of 77, FRFSE T2, TE 85,TR 3500 Classic Fat supressed images which are pristine, vs magnets of 120 slew rate same parameters look signal starved. So gradient strength makes a speed difference, not an image quality difference?
Elise Gough RT(R)(CT)(MR)
|
| View the whole thread | | |
Robert Patten
Thu. 3 Mar.11, 22:08
[Start of: 'Building 3d Volumes from MRI DICOM' 4 Replies]

Category:
General |
Building 3d Volumes from MRI DICOM |
I have a number of MRI and CTs since a bad accident. I am a Mechanical Engineer with an electrical, power and computer background.
I was able to render the CT to 3D using Onis (not sure if you've seen this program), but have been unable to manage it with any of the MRIs.
The MRIs were taken on a Phillips Intera 1.5T
I can render a volume but only (1) of the (3) 2D images are clear before it is rendered and the resulting 3D image is just noise. I've tried various software packages and spent a lot of time with the software. Perhaps I don't have the required image files.
Maybe someone knows the answer.
Thanks,
Bob
|
| View the whole thread | Reply to this thread (login or register first) | |
Steven Ford
Thu. 3 Mar.11, 20:28
[Reply (1 of 8) to: '6-1.5T MAGNETS, DIFFERING GRADIENTS' started by: 'Elise Gough' on Wed. 23 Feb.11]

Category:
Applications and Examinations |
6-1.5T MAGNETS, DIFFERING GRADIENTS |
I assume that you mean a t2 fat suppressed sequence. Differing gradient strengths have only an indirect effect on these images. The fat saturation sequences require additional pulses which take time to execute; stronger gradient systems can execute these pulses faster.
If you see different results, it can be caused by a number of factors; if you can describe the differences, that would be helpful. Generally speaking, the quality of the magnet homogeneity makes a big difference. If the small FOV scans (wrist) look different from magnet to magnet, that's probably not the cause.
You should ask your MRI applications specialist about this, and pay attention to the TE and bandwidth. Are the FOV and number of steps the same?
|