Adana Osco Barras
Tue. 3 Jan.12, 21:09
[Reply (1 of 2) to: 'Hyperintense blood vessels in SE' started by: 'Peruvumba Jayakumar' on Thu. 29 Dec.11]
Category:
Sequences and Imaging Parameters |
Hyperintense blood vessels in SE |
Prolonged T1 relaxation times, decreased T1 tissue contrast, and increased susceptibility rnare typical effects at high field MRI.rnDue to the longer T1 times at 3T, the signal rnintensity inside the vessels is preserved even in smaller vessels. Try TR 700, TE 10.rnInversion recovery sequences are very well suited for optimal gray to white matter contrast, but enhancement after administration of gadolinium may not be visible.
|
| View the whole thread | | |
Reader Mail
Wed. 12 Dec.07, 10:07
[Reply (1 of 2) to: 'double ir physic' started by: 'soontorn siriserussa' on Sun. 2 Dec.07]
Category:
Sequences and Imaging Parameters |
double ir physic |
Different types of double inversion recovery (DIR, 2IR) sequences are used to improve the suppression of blood signal (black blood technique) or to null the signals from two different tissue types (e.g. white matter and cerebrospinal fluid).
The black blood technique (used in cardiovascular MRI) works with two inversion pulses, where the first pulse is nonselective and the second pulse is slice-selective. TI is set to a value at which the signal of the recovering inverted blood is zero (http://www.mr-tip.com/serv1.php?type=db1&dbs=Double%20Inversion).
The second technique (also named gray matter only) is used in brain imaging to improve the detection of lesions, for example in the diagnostic of multiple sclerosis. Two 180° pulses with different TI are used to suppress two different types of tissue simultaneously.
Hope this helps
|
| View the whole thread | | |
Tom Bellini
Mon. 14 Jun.04, 01:57
[Reply (1 of 3) to: 'Black Blood MRA' started by: 'ranran zhao' on Sun. 6 Jun.04]
Category:
Applications and Examinations |
Black Blood MRA |
The black blood effect is a natural high contrast between blood and myocardium caused by the flow void of blood. Black blood techniques are for example used to demonstrate fatty infiltration of the right ventricular wall with right ventricular dysplasia. Black blood MRA is usual based on fast spin echo or inversion recovery sequences and seldom used as cine imaging. The advantage of this type of sequence consists of being less sensitive to respiratory and cardiac motion. Bright blood cine imaging is the standard method, but there are manufacturer with different techniques, e.g. Toshiba has a black blood cine technique.
|