Re: Calculation of Resolution and Tr
Tacq= 32s
Gradient strength= 10mT/m
Matrix size=256x256 pixels
dwell time= 0.03ms

How to calculate resolution and Tr?
Reply to this thread
(login or register first)
Steven Ford

Tue. 5 Aug.14,
19:57

[Reply (3 of 4) to:
'MRI artifact'
started by: 'Daniel Cronk'
on Mon. 16 Jun.14]


 
  Category: 
Artifacts

 
MRI artifact
No it is not an RF leak artifact. That would appear different. I would guess that there was one instance of motion. Note that the artifact is mostly on the high contrast areas (white against black) where the patient can move the most.

Also it may be that your acquisition matrix was a little on the low side, which would lead to larger pixel-y appearance that is more apparent when there's motion.
 
 

Steven Ford
Professional Imaging Services, Inc.
 View the whole thread
Steven Ford

Thu. 17 Mar.11,
03:16

[Reply (2 of 5) to:
'Building 3d Volumes from MRI DICOM'
started by: 'Robert Patten'
on Thu. 3 Mar.11]


 
  Category: 
General

 
Building 3d Volumes from MRI DICOM
In almost all cases, MRI images have a slice thickness far greater than their in-plane resolution, making multiplanar reconstruction meaningless. Also, most MRI images have a gap between the slices, which also renders multiplanar meaningless and impossible (unless you're happy with black bars where the slice gaps exist).

You can look at your images and the slice thickness is on the graphics. the slice gap is usually not displayed, but if you look at the slice location displayed on adjacent slices, and compare to the slice thickness, you can easily compute the slice gap.

The in-place resolution is (approx) the field of view divided by the acquisition matrix, both of which are also printed on the image somewhere.

By the way, if you have the dicom (.dcm) files, there is lots of data that's 'hidden' that you can access with a full featured dicom file editor.
 View the whole thread
David Brooks

Tue. 22 Dec.09,
02:35

[Start of:
'Scan equation I don't get, HELP'
2 Replies]


 
  Category: 
Basics and Physics

 
Scan equation I don't get, HELP
Here's the question:
The approximate scan time of a rapid spin echo sequence with a TR of 3500ms, a TE of 90ms, a 256x256 matrix, 1 exicitation, a 220mm FOV and an echo train length of 5 is______ minutes.

Formal they give is 3500msx1x256/5=?
a.3
b. 8
c. 11
d. 15

I get 179200. What am I missing in my studies and the math?
 View the whole threadReply to this thread
(login or register first)
Gonzalo Sanchez Benavides

Tue. 5 Jun.07,
07:53

[Start of:
'DTI 25 directions on GE signa excite 1.5'
4 Replies]


 
  Category: 
Sequences and Imaging Parameters

 
DTI 25 directions on GE signa excite 1.5
Hello,

I need some help with the DTI 25 gradient directions on GE signa excite.
I'm a neuropsychology researcher and i receive some data with this protocol, but the technician cannot tell me the matrix that i need to process it. It's somthing like this:

26
0 0 0
1 0 1
...

If anybody can help me i would be very grateful.

Thanks
 View the whole threadReply to this thread
(login or register first)

Result Pages : [1] 2 3 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 24 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]