Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 




 
MRI Forum
'Pulse Sequence'
SEARCH FORUM FOR   
 
Result: Searchterm 'Pulse Sequence' found in 12 messages
Result Pages: 1 2 [3] 
More Results: Database  (96)  News Service  (2)  Resources  (5)  
Forum Overview
 bottom
Steven Ford

Mon. 7 Mar.11,
16:07

[Reply (6 of 8) to:
'6-1.5T MAGNETS, DIFFERING GRADIENTS'
started by: 'Elise Gough'
on Wed. 23 Feb.11]


 
  Category: 
Applications and Examinations

 
6-1.5T MAGNETS, DIFFERING GRADIENTS
We maintain a lot of magnets. The leading cause of image quality problems is applications related. Nobody can possibly know all the nuances of pulse sequences by various vendors, software levels, etc. I've heard many times that a certain machine is no good, when in fact the sequences are inefficiently set up. Look there first.

Usually the techs welcome good training. If they don't want to be trained, then you have a problem, but it sounds like this issue is caused by other factors.

If there are differences in baseline quality between the machines, then compensate for that by other means. Signal starvation is easily remedied; keep the quality as consistent as you can and let time be the variable, if it comes down to that.




 View the whole thread
Oleg Borodin

Thu. 9 Aug.07,
13:46

[Start of:
'T1 measurements on 1.5T scaner'
0 Reply]


 
  Category: 
Basics and Physics

 
T1 measurements on 1.5T scaner
Dier colleges

I have same trouble with counting T1 of the gadolinium solution.

f.ex.
There are six fantom with gadolinium solution in 0.5, 1, 2, 4, 8 and 16 mmol/l. Fantoms scanned with pulse sequences SE with TR equal 50, 100, 200, 300, 400 ... 4000 ms. For each fantom calculated T1 by iteractive minimizing technique the equation: S(TR)=M0*(1-exp(-TR/T1))

At the result, I have six T1: 72, 170, 255, 403, 725, 1474 ms.

How can I calculate T1 and R1 of the pure solution of contrast media?
Reply to this thread
(login or register first)

Result Pages : 1 2 [3] 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]