Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 

Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 




 
MRI Forum
'SAR'
SEARCH FORUM FOR   
 
Result: Searchterm 'SAR' found in 22 messages
Result Pages: 1 2 3 4 [5] 
More Results: Database  (62)  News Service  (34)  Resources  (14)  
Forum Overview
 bottom
Schmidt Marga

Sat. 17 Apr.04,
20:41

[Reply (1 of 2) to:
'Moire fringing'
started by: 'Kelly Wimpey'
on Tue. 6 Apr.04]


 
  Category: 
Artifacts

 
Moire fringing
Moire fringes are typical for gradient echo sequences, especially balanced sequences like for example true fisp, bFFE. The reasons are interferences and field inhomogeneities. High order shimming (volume shimming), a more suitable coil, a smaller bandwidth, and scanning not in the edges of the magnetic field will avoid or reduce this artifact. Also parallel imaging techniques can have a positive influence. If this artifact appears, also by using other sequences, a new field adjustment can be necessary (service).
Hope this will help.
 View the whole thread
John Doe

Thu. 27 Nov.03,
22:13

[Reply (1 of 5) to:
'Preparation'
started by: 'Renate'
on Thu. 27 Nov.03]


 
  Category: 
Patient Care and Comfort

 
Preparation
This is not usually necessary.
 View the whole thread

Result Pages : 1 2 3 4 [5] 
 top
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Look
      Ups






MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 18 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]