marc jupin
Thu. 26 Mar.15, 09:03
[Start of: 'receiver gain in siemens trio' 0 Reply]
Category:
Sequences and Imaging Parameters |
receiver gain in siemens trio |
Hi, we measured a phantom with two receiver gain option: LOW and HIGH (available from parameters).
We would like to calculate the difference in signal intensity, but unfortunately the rda files headers do not contain any information about the receiver gain value.
Does anyone have a clue how I could find the receiver gain value?
Thanks.
|
| | Reply to this thread (login or register first) | |
Anna Lena Golay
Wed. 25 Jan.12, 20:59
[Reply (3 of 7) to: 'Imaging optic neuritis' started by: 'Karen Lesley' on Wed. 18 Jan.12]
Category:
General |
Imaging optic neuritis |
A 2-3 mm STIR sequence through the optic nerves may show the characteristic high-signal intensity of optic neuritis. Gadolinium enhanced T1-weighted fat-saturated (to suppress the bright signal of the orbital fat tissue) images of the orbits show the inflammation of the optic nerve. White matter lesions, which denote a higher risk of developing MS, are typical imaged with FLAIR and T2-weighted images (hyperintense lesions), or show enhancement of T1-weighted images postcontrast.
|
| View the whole thread | | |
|
Azza Ahmed
Thu. 8 Jul.10, 10:42
[Start of: 'exceeding slew rates' 0 Reply]
Category:
Sequences and Imaging Parameters |
exceeding slew rates |
Can someone please let me know what happens to the MRI signal intensity if any of the gradient's slew rate was exceeded?
Azza AhmedrnPhD student
|
| | Reply to this thread (login or register first) | |
Reader Mail
Tue. 6 Mar.07, 22:48
[Start of: 'fluid PD vs T2 weighting' 1 Reply]
Category:
Sequences and Imaging Parameters |
fluid PD vs T2 weighting |
Can anyone explain me, why - considering SE sequences with the same TR - fluid in a PD weighting (short TE) should have less signal than in a T2 weighting (long TE)? eg:
Fig 10, p 1397; radiographics resident tutorial: contrast mechanisms in spin echo MR imaging; Radiographics 1994; 14; pp 1389-1404
I would expect that the signal intensity of any tissue (regardless of it Proton density of T2-relaxation time) should decrease with increasing echo time.
Thank you
|
| View the whole thread | Reply to this thread (login or register first) | |
Reader Mail
Mon. 30 Oct.06, 16:25
[Reply (1 of 2) to: 'interleaved slices and mean intensity difference in adjacent slices' started by: 'Reader Mail' on Wed. 4 Oct.06]
Category:
General |
interleaved slices and mean intensity difference in adjacent slices |
With interleaved slice acquisition for example in each TR first the odd slices, and then the even slices are acquired. Differences in signal intensity may occur by changes between these acquisitions (like movement), especially with a long TR. In particular, for fMRI sequences with interleaved slice acquisition a slice time correction is recommended to get accurate measurements.
However, I think "large" signal intensity differences with interleaved slices (except fMRI) are rarely seen with usual sequences and sufficient processing (interpolation) algorithms.
|
| View the whole thread | |
| |