Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 MRI Database 
SEARCH FOR    
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
previous     26 - 30 (of 113)     next 
C--Ca   Ca-Ca   Ca-Ca   Ca-Ce   Ce-Ch   Ch-Ch     Ch-Ci   Ci-Cl   Cl-Cl   Cl-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Cr   Cr-Cr   Cr-Cr   Cr-Cy   
MRI Resources 
Knee MRI - Guidance - Absorption and Emission - Liver Imaging - Hospitals - Functional MRI
 
Chemical Shift ArtifactInfoSheet: - Artifacts -
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Chemical shift, black boundary, spatial misregistration, relief
DESCRIPTION
Black or bright band
During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
mri safety guidance
Image Guidance
For artifact reduction helps a smaller water fat shift (higher bandwidth), a higher matrix, an in phase TE or a spin echo technique. Since the misregistration offset is present in the read out axis the patient may be rescanned with this axis parallel to the fat-water interface. Steeper gradient may be employed to reduce the chemical shift offset in mm. Another strategy is to employ specialized pulse sequences such as fat saturation or inversion recovery imaging. Fat suppression techniques eliminate chemical shift artifacts caused by the lack of fat signal.

See also Black Boundary Artifact and Magnetic Resonance Spectroscopy.
spacer

• View the DATABASE results for 'Chemical Shift Artifact' (7).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
  News & More:
What is chemical shift artefact? Why does it occur? How many Hz at 1.5 T?
   by www.revisemri.com    
Abdominal MRI at 3.0 T: The Basics Revisited
Wednesday, 20 July 2005   by www.ajronline.org    
Chemical Shift ImagingInfoSheet: - Sequences -
Intro, Overview, 
Types of, 
etc.MRI Resource Directory:
 - Spectroscopy pool -
 
(CSI) Chemical shift imaging is an extension of MR spectroscopy, allowing metabolite information to be measured in an extended region and to add the chemical analysis of body tissues to the potential clinical utility of Magnetic Resonance. The spatial location is phase encoded and a spectrum is recorded at each phase encoding step to allow the spectra acquisition in a number of volumes covering the whole sample. CSI provides mapping of chemical shifts, analog to individual spectral lines or groups of lines.
Spatial resolution can be in one, two or three dimensions, but with long acquisition times od full 3D CSI. Commonly a slice-selected 2D acquisition is used. The chemical composition of each voxel is represented by spectra, or as an image in which the signal intensity depends on the concentration of an individual metabolite. Alternatively frequency-selective pulses excite only a single spectral component.
There are several methods of performing chemical shift imaging, e.g. the inversion recovery method, chemical shift selective imaging sequence, chemical shift insensitive slice selective RF pulse, the saturation method, spatial and chemical shift encoded excitation and quantitative chemical shift imaging.

See also Magnetic Resonance Spectroscopy.
spacer

• View the DATABASE results for 'Chemical Shift Imaging' (6).Open this link in a new window

 
Further Reading:
  Basics:
1H MR Spectroscopy and Chemical Shift Imaging of the In Vivo Brain at 7 Tesla
Sunday, 26 November 2006   by tobias-lib.uni-tuebingen.de    
MRI evaluation of fatty liver in day to day practice: Quantitative and qualitative methods
Wednesday, 3 September 2014   by www.sciencedirect.com    
  News & More:
Spin echoes, CPMG and T2 relaxation - Introductory NMR & MRI from Magritek
2013   by www.azom.com    
mDIXON being developed to simplify and accelerate liver MRI
September 2010   by incenter.medical.philips.com    
MRI Resources 
Stimulator pool - Spine MRI - Collections - Process Analysis - Shielding - Veterinary MRI
 
Chemical Shift Reference
 
A compound with respect to whose frequency the chemical shifts of other compounds can be compared. The standard can be either internal or external to the sample. Because of the need for possible corrections due to differential magnetic susceptibility between an external standard and the sample being measured, the use of an internal standard is generally preferred.
spacer
Chemical Shift Selective Imaging SequenceInfoSheet: - Sequences -
Intro, Overview, 
Types of, 
etc.MRI Resource Directory:
 - Spectroscopy pool -
 
(CHESS) A sequence for water suppression in proton MR spectroscopy and for water or fat suppression in MR imaging. This technique uses a frequency-selective 90° pulse to selectively excite the water signal, followed by a spoiler gradient to dephase the resulting magnetization. The gradients may be repeated several times in different directions to increase its effectiveness.

See also Chemical Shift Imaging and Chemical Shift.
spacer

• View the DATABASE results for 'Chemical Shift Selective Imaging Sequence' (2).Open this link in a new window

Chemical Shift Spatial OffsetInfoSheet: - Artifacts -
Case Studies, 
Reduction Index, 
etc.
 
Image artifact of apparent spatial offset of regions with different chemical shifts along the direction of the frequency encoding gradient; a similar effect may be found in the slice selection direction.

See Chemical Shift Artifact.
spacer
 
Further Reading:
  Basics:
1H MR Spectroscopy and Chemical Shift Imaging of the In Vivo Brain at 7 Tesla
Sunday, 26 November 2006   by tobias-lib.uni-tuebingen.de    
MRI Resources 
Mobile MRI - Chemistry - Pacemaker - Supplies - Implant and Prosthesis pool - Databases
 
previous       26 - 30 (of 113)      next 
C--Ca   Ca-Ca   Ca-Ca   Ca-Ce   Ce-Ch    Ch-Ch    Ch-Ci   Ci-Cl   Cl-Cl   Cl-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Co   Co-Cr   Cr-Cr   Cr-Cr   Cr-Cy   
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]