Oversampling
 
Oversampling is the increase in data to avoid aliasing and wrap around artifacts. Aliasing is the incorrectly mapping of tissue signal from outside the FOV to a location inside the FOV. This is caused by the fact, that the acquired k-space frequency data is not sampled density enough.
Oversampling in frequency direction, done by increasing the sampling frequency, prevents this aliasing artifact. The proper frequency based on the sampling theorem (Shannon sampling theorem/Nyquist sampling theorem) must be at least twice the frequency of each frequency component in the incoming signal. All frequency components above this limit will be aliased to frequencies between zero and half of the sampling frequency and combined with the proper signal information, which creates the artifact. Oversampling creates a larger field of view, more data needs to be stored and processed, but this is for modern MRI systems not a real problem. Oversampling in phase direction (no phase wrap), to eliminate wrap around artifacts, by increasing the number of phase encoding steps, results in longer scan/processing times.
spacer

• View the DATABASE results for 'Oversampling' (10).Open this link in a new window

 
Further Reading:
  Basics:
The Basics of MRI
   by www.cis.rit.edu    
The Scientist and Engineer's Guide to Digital Signal Processing
   by www.dspguide.com    
Oxygen Mapping
 
The activation of the brain is mirrored in the local changes in metabolism and hemodynamics (brit. haemodynamics). By using heavily T2 weighted scans activated brain areas show an increase in signal intensity as oxyhaemoglobin (brit. oxyhaemoglobin) is diamagnetic and deoxyhemoglobin (brit. deoxyhaemoglobin) is paramagnetic. Oxygen mapping will benefit from high field strength.

See also Blood Oxygenation Level Dependent Contrast, Haemoglobin, Perfusion Imaging, Diamagnetism, Paramagnetism.
spacer

• View the DATABASE results for 'Oxygen Mapping' (2).Open this link in a new window

 
Further Reading:
  News & More:
Potential and Limitations of Oxygen-17 MR Perfusion Measurements
Monday, 1 March 2004   by www.case.edu    
Gold Acupuncture Needle MRI Pain Discovery
Friday, 3 January 2014   by www.healthcmi.com    
Brain-imaging technique could offer invaluable prognostic data
Thursday, 16 February 2012   by www.theengineer.co.uk    
MRI Resources 
Case Studies - Open Directory Project - Pregnancy - Functional MRI - Education pool - MRI Technician and Technologist Career
 
previous       26 - 27 (of 27)     
O--Of   Of-ON   OP-Op   Or-Os   Os-Ou    Ov-Ox    
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]