Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 MRI Database 
SEARCH FOR    
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
previous     156 - 160 (of 167)     next 
S--Sa   Sa-Sc   Sc-Se   Se-Se   Se-Se   Se-Se   Sh-Sh   Sh-SH   Sh-Sh   SH-Si   Si-Si   Si-Si   Si-Si   Si-Si   Si-Si   Sk-Sl   Sl-Sl   Sl-So   So-Sp   Sp-Sp   Sp-Sp   Sp-Sp   SP-Sp   Sp-Sp   Sp-Sp   Sp-Sp   Sp-St   St-St   St-St   St-Su   Su-Su   Su-Sw     Sy-Sy   Sy-Sy   
MRI Resources 
MRI Technician and Technologist Schools - Most Wanted - Image Quality - - Diffusion Weighted Imaging - Intraoperative MRI
 
Surface CoilForum -
there are related threadsInfoSheet: - Coils -
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A surface coil is essentially a loop of conducting material, such as copper tubing. The in-bore solenoidal sending coil is used as the transmitter of RF energy. This type of receiver coil is placed directly on or over the region of interest for increased magnetic sensitivity. The loop may form various shapes and be bent slightly to conform to the imaged body part. Surface coils have a good SNR for tissues adjacent to the coil and because the signal decrease with the distance, an eligibility homogeneity correction will equalize this over the field of view. A rule of thumb for surface coils is that the sensitivity decreases appreciably beyond a distance equal to the diameter of the coil.
The positioning of the coil is an important determinant of performance. As only the region close to the surface coil will contribute to the signal, there is an improvement in the SNR for these regions, compared to the use of receiver coils that surround the appropriate part of the body. These coils are specifically designed for localized body regions, and provide improved signal to noise ratios by limiting the spatial extent of the excitation or reception.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
spacer

• View the DATABASE results for 'Surface Coil' (81).Open this link in a new window

 
Further Reading:
  Basics:
System Architecture
2003   by www.revisemri.com    
Surface Coil NMRInfoSheet: - Coils -
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A surface coil placed over a region of interest will have an effective selectivity for a volume approximately subtended by the coil circumference and one radius deep from the coil center. Such a coil can be used for simple localization of sites for measurement of chemical shift spectra, especially of phosphorus, and blood flow studies. Some additional spatial selectivity can be achieved with magnetic field gradients.
spacer
MRI Resources 
Jobs - Crystallography - MR Myelography - Blood Flow Imaging - MRI Reimbursement - Image Quality
 
SusceptibilityForum -
there are related threads
 
spacer

• View the DATABASE results for 'Susceptibility' (45).Open this link in a new window

 
Further Reading:
  Basics:
Susceptibility-Weighted Imaging: Technical Aspects and Clinical Applications, Part 1
Friday, 30 January 2009   by www.ajnr.org    
Principles, Techniques, and Applications of T2*- based MR Imaging and Its Special Applications1
September 2009   by pubs.rsna.org    
Abdominal MRI at 3.0 T: The Basics Revisited
Wednesday, 20 July 2005   by www.ajronline.org    
  News & More:
MRI Reveals Significant Brain Abnormalities Post-COVID
Monday, 21 November 2022   by neurosciencenews.com    
Opportunities in Interventional and Diagnostic Imaging by Using High-performance Low-Field-Strength MRI
Tuesday, 1 October 2019   by pubs.rsna.org    
Ultrashort echo time (UTE) MRI of the spine in thalassaemia
February 2004   by bjr.birjournals.org    
Susceptibility ArtifactInfoSheet: - Artifacts -
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
DESCRIPTION
Signal dropout, bright spots, spatial distortion
REASON
HELP
Remove the metal, do not take a gradient echo sequence, take a short echo time
Materials with magnetic susceptibility cause this artifact. There are in general three kinds of materials with magnetic susceptibility: ferromagnetic materials (iron, nickel etc.) with a strong influence and paramagnetic/diamagnetic (aluminium, platinum etc./gold, water, most organic compounds etc.) materials with a minimal/non influence on magnetic fields. In MRI, susceptibility artifacts are caused for example by medical devices in or near the magnetic field or by implants of the patient. These materials with magnetic susceptibility distort the linear magnetic field gradients, which results in bright areas (misregistered signals) and dark areas (no signal) nearby the magnetic material.
mri safety guidance
Image Guidance
Use a spin echo or a fast spin echo sequence, because gradient echo sequences are more sensitve to susceptibility artifacts. A high bandwidth (small water fat shift) and a short echo time help also to reduce this artifact.
In some cases it is even beneficial to use a gradient echo sequence, e.g. a cavernom contains some iron-rich haemosiderin, which also causes a signal void on gradient echo sequences and for this purpose increases the diagnostic image quality.
spacer

• View the DATABASE results for 'Susceptibility Artifact' (8).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Susceptibility Artifacts
   by www.mritutor.org    
  News & More:
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
Ultrashort echo time (UTE) MRI of the spine in thalassaemia
February 2004   by bjr.birjournals.org    
Switchable CoilInfoSheet: - Coils -
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
An RF array coil consisting of several separately resonant elements, any one of which can be selected as the receiver coil at a particular time. Coils not in use are decoupled. Applications of switchable coils include imaging the whole spine without patient repositioning (where the coil elements may collectively be known as a ladder coil), imaging of bilateral structures such as TMJ or orbit using separate coils, or imaging using a coil with selectable field of view.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
spacer

• View the DATABASE results for 'Switchable Coil' (4).Open this link in a new window

MRI Resources 
Education pool - Pregnancy - Mobile MRI - Directories - Claustrophobia - General
 
previous       156 - 160 (of 167)      next 
S--Sa   Sa-Sc   Sc-Se   Se-Se   Se-Se   Se-Se   Sh-Sh   Sh-SH   Sh-Sh   SH-Si   Si-Si   Si-Si   Si-Si   Si-Si   Si-Si   Sk-Sl   Sl-Sl   Sl-So   So-Sp   Sp-Sp   Sp-Sp   Sp-Sp   SP-Sp   Sp-Sp   Sp-Sp   Sp-Sp   Sp-St   St-St   St-St   St-Su   Su-Su    Su-Sw    Sy-Sy   Sy-Sy   
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]