Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Artifact' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Artifact' found in 62 terms [] and 104 definitions []
previous     46 - 50 (of 166)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13]  [14 15 16 17 18 19 20 ... ]
Searchterm 'Artifact' was also found in the following services: 
spacer
News  (6)  Resources  (10)  Forum  (22)  
 
Phase Mismapping (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
spacer
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
Artifact by Patient MovementInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Motion, movement
DESCRIPTION
Blurring, ghosting
REASON
Patient movement
HELP
Fast scan techniques
Patient movement during the scans are often an imaging problem. Artifacts from patient movement are widely varied due to a dependence when during k-space filling the motion occurs. When the patient moving causes only in the last few seconds of the scan at that time the outside edges of K-space were being filled, and as a result the artifact does not overly affect the image (there are only fine lines).
image guidance
Image Guidance
A good cooperation between the patient and the operator is the best way to avoid these artifacts, in difficult cases a sedative may help. If a compliance of the patient is not possible (e.g. pain, stroke, or consciousness), choose fast scan methods like gradient echo or single shot technique.

See also Motion Artifact and Phase Encoded Motion Artifact.
spacer
 
Further Reading:
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
MRI results affected by movement? MIT researchers have an AI-powered solution
Friday, 25 August 2023   by healthimaging.com    
MRI Resources 
Brain MRI - MRI Reimbursement - Libraries - MR Guided Interventions - Pacemaker - Developers
 
Field Inhomogeneity ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Field inhomogeneity
DESCRIPTION
Image distortion signal loss
REASON
HELP
Larger FOV, oversampling
A disturbance of the field homogeneity, because of magnetic material (inside or outside the patient), technical problems or scanning at the edge of the field.
When images were obtained in a progression from the center to the edge of the coil, the homogeneity of the field observed by the imaged volume, changes when the distance from the center of the volume increase. The same problem appears by scanning at a distance from the isocenter in left-right direction or too large field of view.
There are different types of bad image quality, the images are noisy, distorted or the fat suppression doesn't work because of badly set shim currents.
E.g. by using an IR sequence, changes in the T1 recovery rates of the tissues are involved. The inversion time at the center of the imaged volume is appropriate to suppress fat, but at the edge of the coil the same inversion time is sufficient to suppress water. Since the inversion time is not changed, the T1 recovery rates will increase.
mri safety guidance
Image Guidance
Take a smaller imaging volume (and for fat suppression a volume shimming), take care that the imaged region is at the center of the coil and that no magnetic material is inside the imaging volume.
spacer

• View the DATABASE results for 'Field Inhomogeneity Artifact' (3).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Searchterm 'Artifact' was also found in the following services: 
spacer
News  (6)  Resources  (10)  Forum  (22)  
 
Machine Imperfection ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Machine imperfection, data error
DESCRIPTION
Striped ghosts with a shift of half the field of view
REASON
Non-uniform sampling, phase differences
HELP
Data correction
Machine imperfection-based artifacts manifest themselves due to the fact that the odd k-space lines are acquired in a different direction than the even k-space lines. Slight differences in timing result in shifts of the echo in the acquisition window. By the shift theorem, such shifts in the time domain data then produce linear phase differences in the frequency domain data.
Without correction, such phase differences in every second line produce striped ghosts with a shift of half the field of view, so-called Nyquist ghosts. Shifts in the applied magnetic field can also produce similar (but constant in amplitude) ghosts.
This artifact is commonly seen in an EPI image and can arise from both, hardware and sample imperfections.
A further source of machine-based artifact arises from the need to acquire the signal as quickly as possible. For this reason the EPI signal is often acquired during times when the gradients are being switched. Such sampling effectively means that the k-space sampling is not uniform, resulting in ringing artifacts in the image.
mri safety guidance
Image Guidance
Such artifacts can be minimized by careful setup of the spectrometer and/or correction of the data. For this reasons reference data are often collected, either as a separate scan or embedded in the imaging data. The non-uniform sampling can be removed by knowing the form of the gradient switching. It is possible to regrid the data onto a uniform k-space grid.
spacer

• View the DATABASE results for 'Machine Imperfection Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
Magic Angle Effect (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Magic angle
DESCRIPTION
Increase of the T2 time, bright signal in tendons
REASON
Angle about 55° to the main magnetic field
HELP
Angle not about 55°
The magic angle is a precisely defined angle, the value is approximately 54.7°. Hence, two nuclei with a dipolar coupling vector at an angle of approximately 54.7° to a strong external magnetic field have zero dipolar coupling.
Magic angle spinning is a technique in solid-state NMR spectroscopy, which employs this principle to remove or reduce dipolar couplings, thereby increasing spectral resolution. In MRI, the magic angle effect visualizes as bright spots through an increased T2 time on short echo time (TE) images, for e.g. collagen fibers of tendons and ligaments, which are oriented at the magic angle of approximately 54.7° to the magnetic field.
mri safety guidance
Image Guidance
Take care that tendons and ligaments are not oriented at about a 54.7° angle to the main magnetic field.
spacer
 
Further Reading:
  Basics:
Magic angle
   by en.wikipedia.org    
Magic Angle Effects
   by www.mritutor.org    
MRI Resources 
Coils - Contrast Enhanced MRI - Mass Spectrometry - Distributors - RIS - Stent
 
previous      46 - 50 (of 166)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13]  [14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]