Crosstalk (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Crosstalk, cross excitation
DESCRIPTION
Anomalous intensities in an image
REASON
HELP
Slice gap minimum 10%
Crosstalk is an artifact introduced into images by interference between adjacent slices of a scan, caused by a slice profile that is not ideal due to the constraints of the measurement technology. If the slice distances are too small, there is cross talk between the slices, which can affect T1 contrast.
mri safety guidance
Image Guidance
This artifact can be eliminated by limiting the minimum spacing (for the most sequences a minimum gap 10% and for IR sequences 20%) between the slices. Crosstalk can also be reduced by selection of interleaved slices (so a slice gap will not be necessary), but interleaved data acquisition can produce large mean intensity differences between adjacent slices.
spacer
 
Further Reading:
  Basics:
MRI GLOSSARY
   by fonar.com    
Slice-overlap Artifacts
   by www.mritutor.org    
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
Instability ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Virtually any instability or fluctuations in the system will cause artifacts of this type, of greater or lesser intensity.

See Motion Artifact.
spacer
MRI Resources 
Services and Supplies - Homepages - NMR - Spectroscopy pool - Breast Implant - Supplies
 
Wrap Around ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
A wrap around artifact (also called backfolding artifact or aliasing artifact) is produced by inadequate sampling or digitization. Wrap around artifacts result from digitizing fewer than two samples per period in a periodic function. Aliasing can occur in MR imaging whenever the scanned area extends beyond the field of view. These areas extending beyond the field of view boundaries are aliased back into the image to appear at artifactual locations.
 
Images, Movies, Sliders:
 Breast MRI Transverse T1 Pre Contrast 001  Open this link in a new window
    
 
spacer

• View the DATABASE results for 'Wrap Around Artifact' (3).Open this link in a new window

 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Artifact' was also found in the following services: 
spacer
News  (6)  Resources  (10)  Forum  (22)  
 
Eddy Current ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.
 
Quick Overview
Artifact Information
NAME
Eddy current
DESCRIPTION
Image distortion
REASON
Induced by changes in a magnetic field
Eddy currents can cause artifacts in images and may seriously degrade overall magnet performance.
mri safety guidance
Image Guidance
Common uses to reduce the influence of eddy currents on gradient fields are eddy current compensation and shielded gradient coils (active or passive). If persistent, it must be addressed by a service representative to verify the magnet performance.

See also Eddy Current Compensation.
spacer

• View the DATABASE results for 'Eddy Current Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Generic Eddy Current Compensation for Rapid Magnetic Resonance Imaging(.pdf)
   by www.switt.ch    
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
Phase Effect ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
spacer
 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
MRI Resources 
Safety pool - Societies - Knee MRI - Journals - Pathology - MRI Technician and Technologist Career
 
previous      21 - 25 (of 166)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13]  [14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]