Quick Overview Please note that there are different common names for this artifact.
DESCRIPTION
Black contours at boundaries
Image Guidance
| | | | | Further Reading: | Basics:
|
|
| |
| | | Searchterm 'Artifact' was also found in the following services: | | | | |
| | |
| |
|
Quick Overview
HELP
cardiac and respiratory synchronization
Movement of the heart causes blurring and ghosting in the images. The artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
These artifacts can be reduced by using cardiac synchronization: triggering, gating or retrospective triggering. Maximum reduction can be achieved by using triggering in combination with flow compensation, respiratory triggering or breath hold and regional saturation techniques.
See also Motion Artifact. | | | | | | • View the DATABASE results for 'Cardiac Motion Artifact' (2).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Quick Overview Please note that there are different common names for this artifact.
DESCRIPTION
Black or bright band
During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
Image Guidance
| | | | • View the DATABASE results for 'Chemical Shift Artifact' (7).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'Artifact' was also found in the following services: | | | | |
| | |
| |
|
Quick Overview
DESCRIPTION
Loss of contrast between two adjacent tissues
The Partial Volume Artifact is caused by imaging voxel containing two different tissues and therefore possessing a signal average of both tissues.
Image Guidance
| | | | • View the DATABASE results for 'Partial Volume Artifact' (4).
| | | | Further Reading: | Basics:
|
|
| |
| | | Searchterm 'Artifact' was also found in the following services: | | | | |
| | |
| |
|
| | | | • View the DATABASE results for 'Phase Wrapping Artifact' (2).
| | | | Further Reading: | Basics:
|
|
| |
| | | | |
| | | |
|
| |