Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Artifact' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Artifact' found in 62 terms [] and 104 definitions []
1 - 5 (of 166)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13]  [14 15 16 17 18 19 20 ... ]
Searchterm 'Artifact' was also found in the following services: 
spacer
News  (6)  Resources  (10)  Forum  (22)  
 
ArtifactForum -
related threadsInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
An image artifact is a structure not normally present but visible as a result of a limitation or malfunction in the hardware or software of the MRI device, or in other cases a consequence of environmental influences as heat or humidity or it can be caused by the human body (blood flow, implants etc.). The knowledge of MRI artifacts (brit. artefacts) and noise producing factors is important for continuing maintenance of high image quality. Artifacts may be very noticeable or just a few pixels out of balance but can give confusing artifactual appearances with pathology that may be misdiagnosed.
Changes in patient position, different pulse sequences, metallic artifacts, or other imaging variables can cause image distortions, which can be reduced by the operator; artifacts due to the MR system may require a service engineer.
Many types of artifacts may occur in magnetic resonance imaging. Artifacts in magnetic resonance imaging are typically classified as to their basic principles, e.g.:
Physiologic (motion, flow)
Hardware (electromagnetic spikes, ringing)
Inherent physics (chemical shift, susceptibility, metal)

Several techniques are developed to reduce these artifacts (e.g. respiratory compensation, cardiac gating, eddy current compensation) but sometimes these effects can also be exploited, e.g. for flow measurements.

See also the related poll result: 'Most outages of your scanning system are caused by failure of'
spacer
 
• Share the entry 'Artifact':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Artifact by Patient Movement
    • Signal to Noise Ratio
    • Bandwidth
    • Black Boundary Artifact
    • Aliasing Artifact
 
Further Reading:
  Basics:
ARTEFACT VERSUS ARTIFACT
Saturday, 26 January 2002   by www.worldwidewords.org    
  News & More:
MRI results affected by movement? MIT researchers have an AI-powered solution
Friday, 25 August 2023   by healthimaging.com    
Magnetic eyelashes: A new source of MRI artifacts
Wednesday, 24 July 2019   by medicalxpress.com    
On the Horizon - Next Generation MRI
Wednesday, 23 October 2013   by thefutureofthings.com    
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
DC ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
spacer

• View the DATABASE results for 'DC Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Image Artefacts
   by www.fmrib.ox.ac.uk    
MRI Resources 
Raman Spectroscopy - Corporations - Movies - Non-English - Stent - Quality Advice
 
Flow ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
DESCRIPTION
Vascular ghosts (ghosting artifact), anomalous intensities in images
REASON
Movement of body fluids
HELP
Flow compensation, presaturation, triggering
Flow effects in MRI produce a range of artifacts, e.g. intravascular signal void by time of flight effects; turbulent dephasing and first echo dephasing, caused by flowing blood.
Through movement of the hydrogen nuclei (e.g. blood flow), there is a location change between the time these nuclei experience a radio frequency pulse and the time the emitted signal is received (because the repetition time is asynchronous with the pulsatile flow).
The blood flow occasionally produces intravascular high signal intensities due to flow related enhancement, even echo rephasing and diastolic pseudogating. The pulsatile laminar flow within vessels often produces a complex multilayered band that usually propagates outside the head in the phase encoded direction. Blood flow artifacts should be considered as a special subgroup of motion artifacts.
mri safety guidance
Image Guidance
Artifacts can be reduced by reduction of phase shifts with flow compensation (gradient moment nulling), suppression of the blood signal with saturation pulses parallel to the slices, synchronization of the imaging sequence with the heart cycle (cardiac triggering) or can be flipped 90° by swapping the phase//frequency encoding directions.

See also Flow Related Enhancement and Flow Effects.
 
Images, Movies, Sliders:
 Knee MRI Sagittal T1 003  Open this link in a new window
 
spacer

• View the DATABASE results for 'Flow Artifact' (6).Open this link in a new window

 
Further Reading:
  News & More:
MRI measure of blood flow over atherosclerotic plaque may detect dangerous plaque
Friday, 5 April 2013   by www.sciencecodex.com    
Advanced Visualization Techniques Could Change the Paradigm for Diagnosis and Treatment of Heart Disease
Thursday, 31 May 2012   by www.sciencedaily.com    
Searchterm 'Artifact' was also found in the following services: 
spacer
News  (6)  Resources  (10)  Forum  (22)  
 
Gibbs ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Gibbs, Gibbs ringing, truncation
DESCRIPTION
Edge ringing, syrinx-like stripe
REASON
Sharp changes in intensity (incomplete digitization of the echo)
HELP
More samples
The Gibbs or ringing artifact appears as a series of lines in the MR image parallel to abrupt and intense changes in the object at this location. This artifact does not occur visibly on smooth objects. This artifact is caused by the Gibbs phenomenon, an overshoot or ringing of Fourier series occurring at discontinuities.
In the spinal cord, a small syrinx can be simulated by the Gibbs phenomenon. Gibbs artifacts are also seen in other regions, for example the brain//skull interface.
Fine lines visible in an image may be due to undersampling of the high spatial frequencies, respectively incomplete digitization of the echo.
With more encoding steps the Gibbs artifacts is less intense and narrower. Therefore, e.g. the artifact is more intense in the 256 point dimension of a 256x512 acquisition matrix.
mri safety guidance
Image Guidance
This problem can only be resolved by smoothing filters (LanczosSigmaFactor, 2-D Exponential Filtering, Gegenbauer Reconstruction etc.) or with a higher acquisition matrix and/or a smaller FOV, to smooth the object.

See also Gibbs Phenomenon and Apodization.
spacer

• View the DATABASE results for 'Gibbs Artifact' (4).Open this link in a new window

 
Further Reading:
  Basics:
Lanczos sigma Factor
   by mathworld.wolfram.com    
  News & More:
Troubleshooting the ACR MRI Accreditation Phantom Tests
   by www.aapm.org    
Searchterm 'Artifact' was also found in the following services: 
spacer
Radiology  (4) Open this link in a new windowUltrasound  (60) Open this link in a new window
Metal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Metal, susceptibility
DESCRIPTION
Signal dropout, bright spots
REASON
HELP
Remove the metal
Ferromagnetic metal will cause a magnetic field inhomogeneity, which in turn causes a local signal void, often accompanied by an area of high signal intensity, as well as a distortion of the image. They create their own magnetic field and dramatically alter precession frequencies of protons in the adjacent tissues. Tissues adjacent to ferromagnetic components become influenced by the induced magnetic field of the metal hardware rather than the parent field and, therefore, either fail to precess or do so at a different frequency and hence do not generate useful signal. Two components contribute to susceptibility artifact, induced magnetism in the ferromagnetic component itself and induced magnetism in protons adjacent to the component.
Artifacts from metal may have varied appearances on MRI scans due to different type of metal or configuration of the piece of metal. The biocompatibility of metallic alloys, stainless steel, cobalt chrome and titanium alloy is based on the presence of a constituent element within the alloy that has the ability to form an adherent oxide coating that is stable, chemically inert and hence biocompatible. In relation to imaging titanium alloys are less ferromagnetic than both cobalt and stainless steel, induce less susceptibility artifact and result in less marked image degradation.
mri safety guidance
Image Guidance
Remove the metal when possible or take a not so sensitive sequence (a SE or another sequence with a rephasing 180° pulse).

See also Susceptibility Artifact.
spacer

• View the DATABASE results for 'Metal Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Metal-Induced Artifacts in MRI
   by www.ajronline.org    
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
  News & More:
Multiacquisition with variable resonance image combination T2 (MAVRIC SL T2) for postoperative cervical spine with artificial disc replacement
Friday, 11 November 2022   by www.nature.com    
Modeling of Active Shimming of Metallic Needles for Interventional MRI
Monday, 29 June 2020   by pubmed.ncbi.nlm.nih.gov    
MRI Resources 
Process Analysis - Health - Developers - Jobs - MRCP - Safety Products
 
     1 - 5 (of 166)     next
Result Pages : [1 2 3 4 5 6 7 8 9 10 11 12 13]  [14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]