Signa Infinity 1.0T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/index.html From GE Healthcare;
the Signa Infinity Magnetic Resonance system is a short bore, high performance, whole-body imaging system operating at 1.0 Tesla. The system can image in any orthogonal or oblique plane (including single and double axis oblique), using a wide variety of pulse sequences.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head and body coil standard; all other coils optional; open architecture makes system compatible with a wide selection of coils
SYNCHRONIZATION
ECG/peripheral, respiratory gating
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography;; 2D/3D TOF, 2D/3D Phase Contrast;; 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, optional: EPI, 2D/3D Fiesta, FGRET, Spiral
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine
TR
4.4 msec to 12000 msec in increments of 1 msec
TE
1.0 to 2000 msec; increments of 1 msec
SINGLE/MULTI SLICE
Simultaneous scan and reconstruction;; up to 100 images/second with Reflex 100
1 cm to 48 cm continuous
2D 0.7 mm to 20 mm; 3D 0.1 mm to 5 mm
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
0.08 mm; 0.02 mm optional
MAGNET WEIGHT
3613 kg
H*W*D
172 x 208 x 216 cm
POWER REQUIREMENTS
480 or 380/415 V
COOLING SYSTEM TYPE
Closed-loop water-cooled gradient
Less than 0.03 L/hr liquid helium
STRENGTH
SmartSpeed 23 mT/m, HiSpeed Plus 33 mT/m
5-GAUSS FRINGE FIELD
4.0 m x 2.8 m axial x radial
Active
spacer
Searchterm 'Axial' was also found in the following services: 
spacer
Radiology  (7) Open this link in a new windowUltrasound  (17) Open this link in a new window
Signa Infinity 1.5T™ TwinSpeed with ExciteInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/s_excite15/index.html From GE Healthcare;
three dedicated MRI systems - a neuro imager, a cardiovascular system, and a whole body scanner - all in one system.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head and body coil standard; all other coils optional; open architecture makes system compatible with a wide selection of coils
Optional 2D/3D brain and prostate
SYNCHRONIZATION
ECG/peripheral, respiratory gating
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography: 2D/3D TOF, 2D/3D Phase Contrast;; 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, EPI, optional: 2D/3D Fiesta, FGRET, Spiral, Tensor
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine
TR
1.2 to 12000 msec in increments of 1 msec
TE
0.4 to 2000 msec in increments of 1 msec
SINGLE/MULTI SLICE
Simultaneous scan and reconstruction;; up to 200 or 400 images per second
1 cm to 48 cm continuous
2D 0.7 mm to 20 mm; 3D 0.1 mm to 5 mm
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
60 cm
MAGNET WEIGHT
3863 kg
H*W*D
172 x 208 x 216 cm
POWER REQUIREMENTS
480 or 380/415
COOLING SYSTEM TYPE
Closed-loop water-cooled gradient
Less than 0.03 L/hr liquid He
STRENGTH
Zoom 40 mT/m, whole body 23 mT/m
5-GAUSS FRINGE FIELD
4.0 m x 2.8 m axial x radial
Active
spacer

• View the DATABASE results for 'Signa Infinity 1.5T™ TwinSpeed with Excite' (2).Open this link in a new window

 
Further Reading:
  News & More:
Twin Gradient Technology - Potential Advantages For Diffusion Weighted MRI(.pdf)
   by www.paulrharvey.co.uk    
MRI Resources 
Movies - DICOM - Raman Spectroscopy - Online Books - Quality Advice - Pathology
 
Signa Infinity 1.5T™ with ExciteInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/s_excite15/index.html From GE Healthcare;
'EXCITE technology has the potential to open the door to new imaging techniques and clinical applications, leaping beyond conventional two and three-dimensional MRI to true 4D imaging that will improve the diagnosis of disease throughout the human body from head to foot.' Robert R. Edelman, M.D., Professor of Radiology at Northwestern University Medical School and Chairman, Department of Radiology, at Evanston Northwestern Healthcare.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
Head and body coil standard; all other coils optional; open architecture makes system compatible with a wide selection of coils
Optional 2D/3D brain and prostate
SYNCHRONIZATION
ECG/peripheral, respiratory gating, (SmartPrep, SmartStep)
PULSE SEQUENCES
Standard: SE, IR, 2D/3D GRE and SPGR, Angiography: 2D/3D TOF, 2D/3D Phase Contrast;; 2D/3D FSE, 2D/3D FGRE and FSPGR, SSFP, FLAIR, EPI, optional: 2D/3D Fiesta, FGRET, Spiral, Tensor
IMAGING MODES
Localizer, single slice, multislice, volume, fast, POMP, multi slab, cine
TR
1.3 to 12000 msec in increments of 1 msec
TE
0.4 to 2000 msec in increments of 1 msec
SINGLE/MULTI SLICE
Simultaneous scan and reconstruction;; up to 200 or 400 images per second
1 cm to 48 cm continuous
2D 0.7 mm to 20 mm; 3D 0.1 mm to 5 mm
1028 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
0.08 mm; 0.02 mm optional
MAGNET WEIGHT
3863 kg
H*W*D
172 x 208 x 216 cm
POWER REQUIREMENTS
480 or 380/415
COOLING SYSTEM TYPE
Closed-loop water-cooled gradient
CRYOGEN USE, L/hr
less than 0.03 L/hr liquid helium
STRENGTH
SmartSpeed 23 mT/m, HiSpeed Plus 33 mT/m, EchoSpeed Plus 33 mT/m
5-GAUSS FRINGE FIELD
4.0 m x 2.8 m axial x radial
Active
spacer

• View the DATABASE results for 'Signa Infinity 1.5T™ with Excite' (2).Open this link in a new window

Searchterm 'Axial' was also found in the following services: 
spacer
News  (2)  Resources  (2)  Forum  (13)  
 
Solenoid CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
A coil of wire wound in the form of a long cylinder. When a current is passed through the coil, the magnetic field within the coil is relatively uniform. Solenoid RF coils are commonly used when the static magnetic field is perpendicular to the long axis of the body.
The reason against a solenoidal coil is that the signal is missed when the axis of the solenoidal coil is coaxial with the horizontal magnetic field Bo. If the coil's axis can be put near a right angle (60° to 90°) with the principal field, so the loss in the signal is greatly dismissed, and good SNR can be achieved.
The two forms are single turn solenoid and multiple turn solenoids.
spacer
Searchterm 'Axial' was also found in the following services: 
spacer
Radiology  (7) Open this link in a new windowUltrasound  (17) Open this link in a new window
Upright™ MRIInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.fonar.com/standup.htm www.fonar.com/standup.htm www.fonar.com/standup.htm From FONAR Corporation;
in October of 2004, the company changed the product name of the Stand-Up™ MRI to the Upright™ MRI. The Indomitable™, Upright™ MRI is the only open MRI in the world that can perform positional MRI (pMRI), i.e. the Upright MRI™ scans patients in upright, weight-bearing positions, in addition to the conventional lie-down positions. The Upright™ MRI is the only device that can scan patients in the position their symptoms occur, in their position of pain. In early clinical reports independently confirm the effectiveness and potential of positional MRI. In October 2000, Fonar received permission to market the Indomitable™ from the FDA.
Device Information and Specification
CLINICAL APPLICATION
Whole body - weight-bearing MRI - position imaging (flexion, extension, bending, standing, sitting and recumbent scanning)
CONFIGURATION
Front-open and Top-open MRI
Solenoid ("Wrap-around") and planar RF receiver coils
SINGLE/MULTI SLICE
0.1 sec per slice
MULTISLICE
0.1 sec per slice
Min. 6.0 cm
Min. 0.8 mm (3DFT)
MEASURING MATRIX
128 - 512
MAGNET TYPE
Iron-frame electromagnet
45.72 cm horizontal
POWER REQUIREMENTS
380/400/415/440/480 V
COOLING SYSTEM TYPE
Water, closed-loop
CRYOGEN USE, L/hr
-
STRENGTH
12 mT/m
5 GAUSS FRINGE FIELD, radial/axial
active, passive
spacer

• View the DATABASE results for 'Upright™ MRI' (5).Open this link in a new window


• View the NEWS results for 'Upright™ MRI' (9).Open this link in a new window.
 
Further Reading:
  Basics:
Product Information - Fonar Corporation
  News & More:
Landmark Independent Study by UCLA School of Medicine Reports Comparison of Dynamic™ Upright® MRI With Static Upright MRI in More Than 1,000 Patients (1,302):
Thursday, 15 November 2007   by www.fonar.com    
FONAR Wins The Wall Street Journal Technology Innovation Award For 2007 For the FONAR UPRIGHT™ Multi-Position™ MRI
Monday, 1 October 2007   by www.fonar.com    
MRI Resources 
Homepages - Devices - Knee MRI - Quality Advice - Contrast Agents - Resources
 
previous      16 - 20 (of 20)     
Result Pages : [1]  [2 3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]