Result : Searchterm 'Body Coil' found in 1 term [] and 31 definitions []
previous     11 - 15 (of 32)     next
Result Pages : [1]  [2 3 4 5 6 7]
Searchterm 'Body Coil' was also found in the following services: 
spacer
News  (2)  Resources  (2)  Forum  (2)  
 
MAGNETOM ESSENZA
 
www.healthcare.siemens.com/magnetic-resonance-imaging/0-35-to-1-5t-mri-scanner/magnetom-essenza/ From Siemens Medical Systems;
Received FDA clearance in 2007.
The MAGNETOM Essenza is designed to combine high system performance with simple installation and power requirements to provide optimal operating costs for limited budgets. The standard system has up to 25 integrated coil elements and 8 independent radio frequency channels. Tim allows the combination of up to 4 different coils that reduce patient and coil repositioning.
The 1.5 Tesla system is designated for a complete range of clinical applications, including neurology, orthopedics, body imaging, angiography, cardiology, breast imaging, oncology and pediatric MRI.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Ultra-short bore
Head, spine, torso/ body coil, neurovascular, cardiac, neck, and multi-purpose flex coils. Peripheral vascular, breast, shoulder, knee, wrist, foot//ankle, TMJ optional.
CHANNELS (min. / max. configuration)
8, 16
Chemical shift imaging and single volume spectroscopy
IMAGING TECHNIQUES
iPAT, mSENSE and GRAPPA (image, k-space), noncontrast angiography, radial motion compensation
MINIMUM TR
3-D GRE: 1.58 (256 matrix)
MINIMUM TE
3-D GRE: 0.5 (256 matrix)
FOV
0.5 - 45 cm
BORE DIAMETER
or W x H
At isocenter: 60 cm
TABLE CAPACITY
200 kg
MAGNET WEIGHT (gantry included)
4350 kg in operation
DIMENSION H*W*D (gantry included)
145 x 226 x 216 cm
5-GAUSS FRINGE FIELD
2.5 m / 4.0 m
CRYOGEN USE
Zero boil off rate, approx. 10 years
COOLING SYSTEM
Water; single cryogen, 2 stage refrigeration
up to 100 T/m/s
30 mT/m, 300 msec to 10 mT/m
Passive, active; first order standard second order optional
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; 45 kVA
spacer
MRI Resources 
Portals - Journals - Used and Refurbished MRI Equipment - Directories - NMR - Artifacts
 
MAGNETOM Prisma
 
www.healthcare.siemens.com/magnetic-resonance-imaging/3t-mri-scanner/magnetom-prisma From Siemens Medical Systems; Received FDA clearance in 2013.
The MAGNETOM Prisma is the 3T PowerPack for exploration that offers most demanding clinical and research challenges of today and the future. The latest parallel transmit technology, TimTX TrueShape, enables zooming into specific body regions for enhanced image quality. Furthermore, the Tim 4G integrated coil technology offers remarkable imaging flexibility and supports complex examinations across the whole body.
Onsite upgrades to the MAGNETOM Prisma for customers who have already installed the 3 Tesla MAGNETOM Trio are possible.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Ultra-short bore
3 Tesla
Head, spine, torso/ body coil, neurovascular, cardiac, neck, shoulder, knee, wrist, foot//ankle and multi-purpose flex coils. Peripheral vascular, breast, shoulder.
CHANNELS (min. / max. configuration)
64, 128
IMAGING TECHNIQUES
iPAT, mSENSE and GRAPPA (image, k-space), noncontrast angiography, radial motion compensation, Dixon
FOV
0.5 - 50 cm
BORE DIAMETER
or W x H
At isocenter: L-R 60 cm
TABLE CAPACITY
250 kg
MAGNET WEIGHT (gantry included)
13000 kg
DIMENSION H*W*D (gantry included)
173 x 230 x 222 cm
5-GAUSS FRINGE FIELD
2.6 m / 4.6 m
CRYOGEN USE
Zero boil off rate, refill approx. 10 years
COOLING SYSTEM
Water
up to 200 T/m/s
MAX. AMPLITUDE
80 mT/m
Passive, active; first order, second order
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground;
spacer
MRI Resources 
MRI Technician and Technologist Career - Collections - Education - Fluorescence - Supplies - Breast MRI
 
MAGNETOM Skyra
 
www.healthcare.siemens.com/magnetic-resonance-imaging/0-35-to-1-5t-mri-scanner/magnetom-skyra/ From Siemens Medical Systems; Received FDA clearance in 2010.
MAGNETOM Skyra is a top-of-the-line, patient friendly wide bore 3 Tesla MRI system.
The system is equipped with the Tim 4G and Dot system (Total imaging matrix and Day optimizing throughput), to enhance both productivity and image quality with the complete range of advanced applications for clinical routine and research. Tim 4G features lighter, trimmer MRI coils that take up less space inside the magnet but deliver a high coil element density with increased signal to noise ratio and the possibility to use high iPAT factors.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Open bore
3 Tesla
Head, spine, torso/ body coil, neurovascular, cardiac, neck, shoulder, knee, wrist, foot//ankle and multi-purpose flex coils. Peripheral vascular, breast, shoulder.
CHANNELS (min. / max. configuration)
48, 64, 128
Chemical shift imaging, single voxel spectroscopy
IMAGING TECHNIQUES
iPAT, mSENSE and GRAPPA (image, k-space),CAIPIRINHA (k-space), noncontrast angiography, plaque imaging, radial motion compensation, Dixon
MINIMUM TR
3D T1 spoiled GRE: 0.95 (256 matrix)
MINIMUM TE
3D T1 spoiled GRE: 0.22 (256 matrix), Ultra-short TE
FOV
0.5 - 50 cm
BORE DIAMETER
or W x H
At isocenter: L-R 70 cm, A-P (with table) 55 cm
TABLE CAPACITY
250 kg
MAGNET WEIGHT (gantry included)
5768 kg
DIMENSION H*W*D (gantry included)
173 x 231 x 219 cm
5-GAUSS FRINGE FIELD
2.6 m / 4.6 m
CRYOGEN USE
Zero boil off rate, approx. 10 years
COOLING SYSTEM
Water; single cryogen, 2 stage refrigeration
up to 200 T/m/s
MAX. AMPLITUDE
45 mT/m
3 linear with 20 coils, 5 nonlinear 2nd-order
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; 110 kVA
spacer
Searchterm 'Body Coil' was also found in the following services: 
spacer
News  (2)  Resources  (2)  Forum  (2)  
 
MAGNETOM Spectra
 
www.healthcare.siemens.com/magnetic-resonance-imaging/3t-mri-scanner/magnetom-spectra From Siemens Medical Systems;
Received FDA clearance in 2012.
The MAGNETOM Spectra is a cost-optimized high field MRI system with Tim 4G and Dot technologies. The system consumes less energy compared to other 3 Tesla scanners. The magnet-cooling helium is contained in a closed loop, which prevents the gas from escaping and reduces the need for refills. TimTX includes innovative techniques in the radio frequency excitation hardware as well as new application and processing features enabling uniform RF distribution in all body regions.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Short bore
3 Tesla
Head, spine, torso/ body coil, neurovascular, neck and multi-purpose flex coils. Peripheral vascular, breast, shoulder, knee, wrist, foot//ankle, endorectal optional.
CHANNELS
24
Chemical shift imaging, single voxel spectroscopy
IMAGING TECHNIQUES
iPAT, mSENSE and GRAPPA (image, k-space), noncontrast angiography, radial motion compensation
FOV
0.5 - 45 cm
BORE DIAMETER
or W x H
At isocenter: L-R 60 cm
TABLE CAPACITY
200 kg
MAGNET WEIGHT
7200 kg
DIMENSION H*W*D (gantry included)
173 x 231 x 219 cm
5-GAUSS FRINGE FIELD
2.6 m / 4.6 m
CRYOGEN USE
Zero boil off rate, approx. 10 years
COOLING SYSTEM
Water; single cryogen, 2 stage refrigeration
up to 125 T/m/s
MAX. AMPLITUDE
33 mT/m
Passive, active; first order standard, second order optional
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; connection value with chiller 100 kvA /without chiller 60 kVA
spacer
MRI Resources 
Contrast Enhanced MRI - Spectroscopy pool - Anatomy - Brain MRI - Musculoskeletal and Joint MRI - Manufacturers
 
MAGNETOM Trioâ„¢ (TIM System)InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.
 
www.medical.siemens.com/webapp/wcs/stores/servlet/ProductDisplay?storeId=10001&langId=-11&catalogId=-11&catTree=100001%2C12786%2C12754&level=0&productId=145739 From Siemens Medical Systems;
The Magnetom Trio™, a 3T whole body MRI system with Tim (total imaging matrix technology), targets clinical applications such as abdominal, cardiac, spine, whole body and orthopedics. TIM enables flexible coil combinations for high resolution imaging of large anatomical areas without the need to change coils.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Short bore
CP head, CP integrated body coil, neck matrix, breast array, 8-channel knee array, CP wrist, shoulder array, CP extremity
Yes/Mutli-nuclear MRS
PULSE SEQUENCES
SE, IR, 2D/3D TurboSE, Turbo IR, Dark-Fluid IR, True IR, 2D/3D MEDIC, 2D/3D GRE FLASH, 2D/3D GRE FISP, 2D/3D PSIF, 2D TurboFLASH, 3D MP-RAGE, 3D TurboFLASH, 2D/3D TOF angiography, MTC, TONE with 3D TOF MRA, GMR, LOTA
IMAGING MODES
Single, multislice, volume study, multi angle, multi oblique
SINGLE/MULTI SLICE
178 images/sec at 256 x 256 at 100% FOV
40 cm
Min 2D/3D: 0.1/0.05 mm
1024 x 1024 full screen display
MEASURING MATRIX
64 x 64 to 1024 x 1024
10 micrometer
60 cm
MAGNET WEIGHT
12000 kg
H*W*D
221 x 244 x 213 cm
POWER REQUIREMENTS
380/400/420/440/480 V
COOLING SYSTEM TYPE
Single cryogen, 2 stage refrig.
0.12 L/hr helium
STRENGTH
40 - 45 mT/m
200 mT/m/ms
5-GAUSS FRINGE FIELD
3.4 m / 5.9 m
Passive, act.; 1st order std./2nd opt.
spacer
MRI Resources 
Guidance - PACS - MRA - Lung Imaging - Pregnancy - Artifacts
 
previous      11 - 15 (of 32)     next
Result Pages : [1]  [2 3 4 5 6 7]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]