The portion of a MR spectrometer comprising the sample container and the RF coils, with some associated electronics. The RF coils may consist of separate receiver and transmitter coils in a crossed coil configuration, or, alternatively, a single coil to perform both functions.
The selective excitation of spins in only a limited region of space. This can be particularly useful for spectroscopy as well as imaging. Spatial localization of the signal source may be achieved through spatially selective excitation and the resulting signal may be analyzed directly for the spectrum corresponding to the excited region. It is usually achieved with selective excitation.
Typically, a single dimension of localization can be achieved with one selective RF excitation pulse (and a magnetic field gradient along a desired direction), while a localized volume (3D) can be excited with a stimulated echo produced with three selective RF pulses whose selective magnetic field gradients are mutually orthogonal, having a common intersection in the desired region. Similar 'crossed plane' excitation can be used with selective 180° refocusing pulses and conventional spin echoes.
A degree of spatial localization of excitation can alternatively be achieved with depth pulses, e.g. when using surface coils for excitation as well as signal detection. An indirect application of selective excitation for volume-selected spectroscopy is to use appropriate combinations of signals acquired after selective inversion of different regions, in order to subtract away the signal from undesired regions.