Magnetic ForcesMRI Resource Directory:
 - MRI Accidents -
 
Forces can result from the interaction of magnetic fields. Pulsed magnetic field gradients can interact with the main magnetic field during the MRI scan, to produce acoustic noise through the gradient coil.
Magnetic fields attract ferromagnetic objects with forces, which can be a lethal danger if one is hit by an unrestrained object in flight. One could also be trapped between the magnet and a large unrestrained ferromagnetic object or the object could damage the MRI machine.
Access control and personnel awareness are the best preventions of such accidents. The attraction mechanism for ferromagnetic objects is that the magnetic field magnetizes the iron. This induced magnetization reacts with the gradient of the magnetic field to produce an attraction toward the strongest area of the field. The details of this interaction are very dependent on the shape and composition of the attracted object. There is a very rapid increase of force as one approaches a magnet. There is also a torque or twisting force on objects, e.g. a long cylinder (such as a pen or an intracranial aneurysm clip) will tend to align along the magnet's field lines. The torque increases with field strength while the attraction increases with field gradient.
Depending on the magnetic saturation of the object, attraction is roughly proportional to object mass. Motion of conducting objects in magnetic fields can induce eddy currents that can have the effect of opposing the motion.

See also Duty Cycle.

See also the related poll result: 'Most outages of your scanning system are caused by failure of'
spacer
• For this and other aspects of MRI safety see our InfoSheet about MRI Safety.
• Patient-related information is collected in our MRI Patient Information.

 
Further Reading:
  Basics:
How strong are magnets?
   by my.execpc.com    
Magnetic Field of the Strongest Magnet
2003   by hypertextbook.com    
  News & More:
Imaging chain faces regulators after inmate, guard get stuck to MRI machine
Friday, 1 December 2023   by healthimaging.com    
Measuring magnetic force field distributions in microfluidic devices: Experimental and numerical approaches
Saturday, 2 December 2023   by analyticalsciencejournals.onlinelibrary.wiley.com    
Two stuck to MRI machine for 4 hrs
Tuesday, 11 November 2014   by www.mumbaimirror.com    
New imaging project for new applications in cancer diagnostics
Monday, 27 March 2017   by www.news-medical.net    
MRI Safety Resources 
Pacemaker - Safety Products - Guidance - Safety pool - Nerve Stimulator
 
ContraindicationsForum -
related threadsMRI Resource Directory:
 - Safety -
 
The principal contraindications of the MRI procedure are mostly related to the presence of metallic implants in a patient. The risks of MRI scans increase with the used field strength. In general, implants are becoming increasingly MR safe and an individual evaluation is carried out for each case.
mri safety guidance
MRI Safety Guidance
Some patients should not be examined in MRI machines, or come closer than the 5 Gauss line to the system.

Absolute Contraindications for the MRI scan:
•
electronically, magnetically, and mechanically activated implants
•
ferromagnetic or electronically operated active devices like automatic cardioverter defibrillators
•
metallic splinters in the eye
•
ferromagnetic haemostatic clips in the central nervous system (CNS)

Patients with absolute contraindications should not be examined or only with special MRI safety precautions. Patients with an implanted cardiac pacemaker have been scanned on rare occasions, but pacemakers are generally considered an absolute contraindication. Relative contraindications may pose a relative hazard, and the type and location of an implant should be assessed prior to the MRI examination.

Relative Contraindications for the MRI scan:
•
other pacemakers, e.g. for the carotid sinus
•
lead wires or similar wires (MRI Safety risk)
•
prosthetic heart valves (in high fields, if dehiscence is suspected)
•
haemostatic clips (body)
•

Osteosynthesis material is usually anchored so well in the patients that no untoward effect will result. Another effect on metal parts in the patient's body is the heating of these parts through induction. In addition, image quality may be severely degraded. The presence of other metallic implants such as surgical clips etc. should be made known to the MRI operators. Most of these materials are non-magnetic, but if magnetic, they can pose a hazard.

See also MRI safety, Pregnancy, Claustrophobia and Tattoos.
Radiology-tip.comradRadiation Safety,  As Low As Reasonably Achievable
spacer
Medical-Ultrasound-Imaging.comUltrasound Safety
spacer

• View the DATABASE results for 'Contraindications' (11).Open this link in a new window

 
Further Reading:
  Basics:
MRI in Patients with Implanted Devices: Current Controversies
Monday, 1 August 2016   by www.acc.org    
Newer Heart Devices Safe During MRI
Monday, 23 August 2004   by www.hospimedica.com    
Physics of MRI Safety
   by www.aapm.org    
FDA Releases New Guidance On Establishing Safety, Compatibility Of Passive Implants In MR Environments
Tuesday, 16 December 2014   by www.meddeviceonline.com    
  News & More:
Women with permanent make-up tattoos suffer horrific facial burns after going in for MRI scans - which create an electric current in the ink
Monday, 4 July 2016   by www.dailymail.co.uk    
Positive diagnosis for neural therapeutic implants
Tuesday, 19 April 2016   by medicalxpress.com    
Codman Neuro develops new MRI-resistant programmable valve for treatment of hydrocephalus
Tuesday, 22 September 2015   by www.news-medical.net    
MRI Resources 
Manufacturers - Functional MRI - Research Labs - Contrast Enhanced MRI - RIS - Devices
 
Metal Detector CoilMRI Resource Directory:
 - Coils -
 
Metal detector coils can be used to find hidden ferromagnetic materials inside of clothes or the body of persons before entering a scanning room. Magnetic fields attract ferromagnetic objects with forces, which can be dangerous for humans and MRI machines.
spacer
Searchterm 'Ferromagnetic' was also found in the following services: 
spacer
News  (7)  Resources  (3)  Forum  (1)  
 
Cochlear ImplantMRI Resource Directory:
 - Safety -
 
mri safety guidance
MRI Safety Guidance
Not necessarily a contraindication, but the examination may damage or impair it. Interference in electronic cochlear implants and ferromagnetic mechanical stapedial replacements has been reported.
spacer

• View the DATABASE results for 'Cochlear Implant' (2).Open this link in a new window

 
Further Reading:
  Basics:
Pain, magnet displacement in MRI in patients with cochlear implants
Thursday, 20 November 2014   by medicalxpress.com    
Pediatric and Adult Cochlear Implantation1
2003   by radiographics.rsnajnls.org    
  News & More:
FDA Approves MED-EL's SYNCHRONY Cochlear Implant
Friday, 23 January 2015   by www.businesswire.com    
Hazards: Scanners Can Damage Hearing Implants
Wednesday, 10 December 2008   by www.nytimes.com    
MRI Resources 
DICOM - Manufacturers - Guidance - Open Directory Project - MRI Reimbursement - Directories
 
Field Inhomogeneity ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Field inhomogeneity
DESCRIPTION
Image distortion signal loss
REASON
HELP
Larger FOV, oversampling
A disturbance of the field homogeneity, because of magnetic material (inside or outside the patient), technical problems or scanning at the edge of the field.
When images were obtained in a progression from the center to the edge of the coil, the homogeneity of the field observed by the imaged volume, changes when the distance from the center of the volume increase. The same problem appears by scanning at a distance from the isocenter in left-right direction or too large field of view.
There are different types of bad image quality, the images are noisy, distorted or the fat suppression doesn't work because of badly set shim currents.
E.g. by using an IR sequence, changes in the T1 recovery rates of the tissues are involved. The inversion time at the center of the imaged volume is appropriate to suppress fat, but at the edge of the coil the same inversion time is sufficient to suppress water. Since the inversion time is not changed, the T1 recovery rates will increase.
mri safety guidance
Image Guidance
Take a smaller imaging volume (and for fat suppression a volume shimming), take care that the imaged region is at the center of the coil and that no magnetic material is inside the imaging volume.
spacer

• View the DATABASE results for 'Field Inhomogeneity Artifact' (3).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Libraries - Non-English - Education pool - Resources - Pregnancy - Patient Information
 
previous      6 - 10 (of 22)     next
Result Pages : [1]  [2 3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]