Foreign Bodies | |
| |
|
Occult incorporated ferromagnetic foreign bodies are dangerous, in particular those close to the eyes. The patient's history may help because many patients do not remember such accidents. In case of doubt, X-rays should be taken prior to MR imaging. | | | | • For this and other aspects of MRI safety see our InfoSheet about MRI Safety. | | | • Patient-related information is collected in our MRI Patient Information.
| | | | • Share the entry 'Foreign Bodies': | | | | |
| | | Searchterm 'Foreign Bodies' was also found in the following service: | | | | |
| | |
| |
|
It is important to remember when working around a superconducting magnet that the magnetic field is always on. Under usual working conditions the field is never turned off. Attention must be paid to keep all ferromagnetic items at an adequate distance from the magnet. Ferromagnetic objects which came accidentally under the influence of these strong magnets can injure or kill individuals in or nearby the magnet, or can seriously damage every hardware, the magnet itself, the cooling system, etc..
See MRI resources Accidents.
The doors leading to a magnet room should be closed at all times except when entering or exiting the room. Every person working in or entering the magnet room or adjacent rooms with a magnetic field has to be instructed about the dangers. This should include the patient, intensive-care staff, and maintenance-, service- and cleaning personnel, etc..
The 5 Gauss limit defines the 'safe' level of static magnetic field exposure. The value of the absorbed dose is fixed by the authorities to avoid heating of the patient's tissue and is defined by the specific absorption rate.
Leads or wires that are used in the magnet bore during imaging procedures, should not form large-radius wire loops. Leg-to-leg and leg-to-arm skin contact should be prevented in order to avoid the risk of burning due to the generation of high current loops if the legs or arms are allowed to touch. The patient's skin should not be in contact with the inner bore of the magnet.
The outflow from cryogens like liquid helium is improbable during normal operation and not a real danger for patients.
The safety of MRI contrast agents is tested in drug trials and they have a high compatibility with very few side effects. The variations of the side effects and possible contraindications are similar to X-ray contrast medium, but very rare. In general, an adverse reaction increases with the quantity of the MRI contrast medium and also with the osmolarity of the compound.
See also 5 Gauss Fringe Field, 5 Gauss Line, Cardiac Risks, Cardiac Stent, dB/dt, Legal Requirements, Low Field MRI, Magnetohydrodynamic Effect, MR Compatibility, MR Guided Interventions, Claustrophobia, MRI Risks and Shielding. | | | | | | | | | • View the DATABASE results for 'MRI Safety' (42).
| | | • View the NEWS results for 'MRI Safety' (13).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | MRI Safety Resources | | | | |
| | | |
| |
|
Image Guidance
The higher the frequency, the larger will be the amount of heat developed. The more ionic the biochemical environment in the tissue, the more energy that will be deposited as heat.
This effect is well known for homogeneous model systems, but the complex structure of various human tissues makes detailed theoretical calculations very difficult, if not impossible.
By scanning problems, it is important to verify the transmission frequency. If the RF transmitted into the patient was, e.g. 5000 Hz lower than the resonance frequency of the protons, no protons was excited, and no signal returns. | | | | • View the DATABASE results for 'Radio Frequency' (72).
| | | • View the NEWS results for 'Radio Frequency' (2).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Tumor specific MRI contrast agents are in development to provide better delineation and progression information for various tumors. Clinical oncology has a need for contrast agents that can identify tumors and metastases at a size of 100,000 cells rather than 1,000,000,000 cells. This level of sensitivity requires excellent tumor targeting of imaging agents and a high MRI signal.
Tumor specific agents accumulate at pathological tissues by passive or active targeting mechanisms. Passive targeting agents use e.g., the natural defense mechanisms in which phagocytic cells remove foreign particles from the body. Active targeting is based on a ligand-directed, site-specific accumulation of contrast agents. The availability of macromolecular contrast agents such as feruglose and ultrasmall superparamagnetic iron oxide ( USPIO), which permit the assessment of tissue permeability, may also improve the detection of tumor grade, tumor type, and response to drugs that target angiogenesis.
See also Monoclonal Antibodies, Metalloporphyrins, Nitroxides and Ferrioxamine. | | | | • View the DATABASE results for 'Tumor Specific Agents' (6).