Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Fractional Echo' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Fractional Echo' found in 1 term [] and 1 definition [], (+ 1 Boolean[] results
1 - 3 (of 3)     
Result Pages : [1]
MRI Resources 
Nerve Stimulator - Portals - Safety pool - Breast MRI - MRI Accidents - Implant and Prosthesis pool
 
Fractional Echo
 
Fractional echo (also called asymmetric or partial echo) is used to shorten the echo time in a sequence, by acquiring partial echoes in the frequency direction. The reduction of echo time is possible because if the first part of the echo is not received, the dephasing lobe of the frequency encoding gradient is not to be on for quite as long, and this saves time.

See also Partial Fourier Technique, Read Conjugate Symmetry, Single Side View and acronyms for 'fractional echo' from different manufacturers.
spacer
 
• Share the entry 'Fractional Echo':  Facebook  Twitter  LinkedIn  
 
Further Reading:
  News & More:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
MRI Resources 
Intraoperative MRI - Spine MRI - Image Quality - Distributors - MR Guided Interventions - Stimulator pool
 
Partial Echo
 
(PE) The partial echo technique (also called fractional echo) is used to shorten the minimum echo time. By the acquisition of only a part of k-space data this technique benefits (like all partial Fourier techniques) from the complex conjugate symmetry between the k-space halves (this is called Hermitian symmetry).
The dephasing gradient in the frequency direction is reduced, and the duration of the readout gradient and the data acquisition window are shortened. Partial echo gives a better SNR at a given TE when a smaller FOV or thinner slices are selected, allows a longer sampling time, and a larger water fat shift (WFS, see also bandwidth) due to a lower gradient amplitude. The resolution is not affected. This is often used in gradient echo sequences (e.g. FLASH, Contrast Enhanced Magnetic Resonance Angiography) to reduce the echo time and yields a lower gradient moment. The disadvantage of using a partial echo can be a lower SNR, although this may be partly offset by the reduced echo time.
Also called Fractional Echo, Read Conjugate Symmetry, Single Side View.

See also Partial Fourier Technique and acronyms for 'partial echo' from different manufacturers.
spacer

• View the DATABASE results for 'Partial Echo' (4).Open this link in a new window

 
Further Reading:
  Basics:
Method and apparatus for subterranean formation flow imaging
   by www.google.com    
MRI Resources 
Case Studies - Jobs pool - Abdominal Imaging - Software - Patient Information - Portals
 
Inhomogeneity
 
Inhomogeneity is the degree of lack of homogeneity, for example the fractional deviation of the local magnetic field from the average value of the field. Inhomogeneities of the static magnetic field, produced by the scanner as well as by object susceptibility, is unavoidable in MRI. The large value of gyromagnetic coefficient causes a significant frequency shift even for few parts per million field inhomogeneity, which in turn causes distortions in both geometry and intensity of the MR images.
Manufacturers try to make the magnetic field as homogeneous as possible, especially at the core of the scanner. Even with an ideal magnet, a little inhomogeneity is always left and is caused in addition by the susceptibility of the imaging object. The geometrical distortion (displacement of the pixel locations) are important e.g., for some cases as stereotactic surgery. Displacements up to 3 to 5 mm have been reported. The second problem is the undesired changes in the intensity or brightness of pixels, which may cause problems in determining different tissues and reduce the maximum achievable image resolution.
mri safety guidance
Image Guidance
General strategies for reducing field inhomogeneity induced artifacts:
Increasing the strength of the gradient magnetic field.
Decreasing the echo time.
Improving the image resolution. Phase encoding. Postprocessing.
spacer

• View the DATABASE results for 'Inhomogeneity' (21).Open this link in a new window

 
Further Reading:
  News & More:
Why non-magnetic capacitors matter in medical imaging
Wednesday, 19 February 2020   by www.medicaldesignandoutsourcing.com    
Implementation of Dual-Source RF Excitation in 3 T MR-Scanners Allows for Nearly Identical ADC Values Compared to 1.5 T MR Scanners in the Abdomen
Wednesday, 29 February 2012   by www.plosone.org    
MRI Resources 
Raman Spectroscopy - Education pool - Spectroscopy pool - Safety Products - Bioinformatics - Calculation
 
     1 - 3 (of 3)     
Result Pages : [1]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]