Free Induction Decay
 
(FID) A free induction decay curve is generated as excited nuclei relax. The amplitude of the FID signal becomes smaller over time as net magnetization returns to equilibrium. If transverse magnetization of the spins is produced, e.g. by a 90° pulse, a transient MR signal will result that will decay toward zero with a characteristic time constant T2 (or T2*); this decaying signal is the free induction decay.
The signal peaks of the echoes fall onto this T2 decay curve, while at each echo the signals arise and decay with T2*. The typical T2 relaxation times being of the order of 5-200 ms in the human body. The first part of the FID is not observable (named the 'receiver dead time') caused by residual effects of the powerful exciting radio frequency pulse on the electronics of the receiver.
spacer
 
• Share the entry 'Free Induction Decay':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • T2 Relaxation
    • Spin
    • Excitation
    • Net Magnetization Vector
    • T2 Time
 
Further Reading:
  Basics:
Free induction decay
   by en.wikipedia.org    
  News & More:
Magnetic resonance imaging
   by www.scholarpedia.org    
MRI Resources 
Chemistry - Implant and Prosthesis pool - IR - DICOM - Breast Implant - Veterinary MRI
 
SHORT Repetition Technique Based on Free Induction DecayInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
spacer

• View the DATABASE results for 'SHORT Repetition Technique Based on Free Induction Decay' (2).Open this link in a new window

MRI Resources 
MRI Technician and Technologist Jobs - Case Studies - Mobile MRI - Process Analysis - Pediatric and Fetal MRI - Lung Imaging
 
Contrast Enhanced Gradient Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
Contrast enhanced GRE sequences provide T2 contrast but have a relatively poor SNR. Repetitive RF pulses with small flip angles together with appropriate gradient profiles lead to the superposition of two resonance signals.
The first signal is due to the free induction decay FID observed after the first and all ensuing RF excitations.
The second is a resonance signal obtained as a result of a spin echo generated by the second and all addicted RF-pulses.
Hence it is absent after the first excitation, it is a result of the free induction decay of the second to last RF-excitation and has a TE, which is almost 2TR. For this echo to occur the gradients have to be completely symmetrical relative to the half time between two RF-pulses, a condition that makes it difficult to integrate this pulse sequence into a multiple slice imaging technique. The second signal not only contains echo contributions from free induction decay, but obviously weakened by T2-decay. Since the echo is generated by a RF-pulse, it is truly T2 rather than T2* weighted. Correspondingly it is also less sensitive to susceptibility changes and field inhomogeneities.
Companies use different acronyms to describe certain techniques.
Different terms (see also acronyms) for these gradient echo pulse sequences:
CE-FAST Contrast Enhanced Fourier Acquired Steady State,
CE-FFE Contrast Enhanced Fast Field Echo,
CE-GRE Contrast Enhanced Gradient-Echo,
DE-FGR Driven Equilibrium FGR,
FADE FASE Acquisition Double Echo,
PSIF Reverse Fast Imaging with Steady State Precession,
SSFP Steady State Free Precession,
T2 FFE Contrast Enhanced Fast Field Echo (T2 weighted).

In this context, 'contrast enhanced' refers to the pulse sequence, it does not mean enhancement with a contrast agent.
spacer

• View the DATABASE results for 'Contrast Enhanced Gradient Echo Sequence' (4).Open this link in a new window

MRI Resources 
Examinations - Developers - Jobs - Veterinary MRI - Knee MRI - Bioinformatics
 
Coherent Gradient EchoInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Coherent gradient echo sequences can measure the free induction decay (FID), generated just after each excitation pulse or the echo formed prior to the next pulse. Coherent gradient echo sequences are very sensitive to magnetic field inhomogeneity. An alternative to spoiling is to incorporate residual transverse magnetization directly into the longitudinal steady state. These GRE sequences use a refocusing gradient in the phase encoding direction during the end module to maximize remaining transverse (xy) magnetization at the time when the next excitation is due, while the other two gradients are, in any case, balanced.
When the next excitation pulse is sent into the system with an opposed phase, it tilts the magnetization in the -a direction. As a result the z-magnetization is again partly tilted into the xy-plane, while the remaining xy-magnetization is tilted partly into the z-direction.
A fully refocused sequence with a properly selected and uniform f would yield higher signal, especially for tissues with long T2 relaxation times (high water content) so it is used in angiographic, myelographic or arthrographic examinations and is used for T2* weighting. The repetition time for this sequence has to be short. With short TR, coherent GE is also useable for breath hold and 3D technique. If the repetition time is about 200 msec there's no difference between spoiled or unspoiled GE. T1 weighting is better with spoiled techniques.
The common types include GRASS, FISP, FAST, and FFE.
The T2* component decreases with long TR and short TE. The T1 time is controlled by flip angle. The common TR is less than 50 ms and the common TE less than 15 ms
Other types have stronger T2 dependence but lower SNR. They include SSFP, CE-FAST, PSIF, and CE-FFE-T2.
Examples of fully refocused FID sequences are TrueFISP, bFFE and bTFE.
spacer

• View the DATABASE results for 'Coherent Gradient Echo' (6).Open this link in a new window

MRI Resources 
Musculoskeletal and Joint MRI - Software - Examinations - Manufacturers - Liver Imaging - Homepages
 
Fourier TransformationMRI Resource Directory:
 - Process Analysis -
 
(FT) The Fourier transformation is a mathematical procedure to separate out the frequency components of a signal from its amplitudes as a function of time, or the inverse Fourier transformation (IFT) calculates the time domain from the frequency domain. The FT is used to generate the spectrum from the free induction decay or spin echo in the pulse MR technique and is essential to most MR imaging techniques. The Fourier transformation can be generalized to multiple dimensions, e.g. to relate an image to its corresponding k-space representation, or to include chemical shift information in some chemical shift imaging techniques. Fourier transformation analysis allows spatial information to be reconstructed from the raw data.
spacer

• View the DATABASE results for 'Fourier Transformation' (39).Open this link in a new window

 
Further Reading:
  Basics:
Fourier Transform Imaging of Spin Vortex Eigenmodes
Friday, 13 August 2004   by www.physik.uni-regensburg.de    
MR Image Reconstruction from Raw Data
   by dukemil.egr.duke.edu    
The Scientist and Engineer's Guide to Digital Signal Processing
MRI Resources 
Breast MRI - Contrast Agents - Online Books - Veterinary MRI - MRI Technician and Technologist Career - Absorption and Emission
 
     1 - 5 (of 8)     next
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]