Frequency Offset
 
The difference between given signal frequency and a reference frequency.
spacer
 
• Share the entry 'Frequency Offset':  Facebook  Twitter  LinkedIn  
MRI Resources 
MRA - Pathology - MR Guided Interventions - Cochlear Implant - MRI Reimbursement - Nerve Stimulator
 
Sample Imperfection (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Sample imperfection
DESCRIPTION
Shifts of the signal in the phase encoding direction
REASON
Distorting the k-space trajectory, reduced bandwidth
HELP
Fat suppression, more excitations
Artifacts either by distorting the k-space trajectory (i.e. due to imperfect shimming) or as a consequence of the reduced bandwidth in the phase encode direction, commonly with EPI sequences.
While a standard spin warp-based sequence has an infinitely large bandwidth in the phase encode direction (about 1 or 2 kH), the bandwidth in EPI is related to the time between the gradient echoes (about a millisecond).
Hence even small frequency offsets can result in significant shifts of the signal in the phase encoding direction. Segmentation can introduce ghosting if there are significant difference in the amplitude and phase of the signal. This can be a particular problem when trying to acquire the segments in rapid succession.
mri safety guidance
Image Guidance
Suitable choices of excitation schemes and/or subsequent correction can help to reduce this artifact. The signal from fat can easily be offset by a large fraction of the FOV, and must be suppressed. The effect of frequency offsets can be reduced by collecting data with more than one excitation, which effectively increases the bandwidth in the phase encoding direction.
spacer
MRI Resources 
Examinations - MRI Centers - Raman Spectroscopy - Diffusion Weighted Imaging - Used and Refurbished MRI Equipment - Implant and Prosthesis
 
Adiabatic Radio Frequency Pulses
 
Adiabatic RF pulses are amplitude and frequency modulated pulses that are insensitive to the effects of B1-inhomogeneity and frequency offset (conventional RF pulses used in MRI are only amplitude modulated). Due to an extended application time adiabatic RF pulses are mostly used in NMR imaging applications.
spacer

• View the DATABASE results for 'Adiabatic Radio Frequency Pulses' (2).Open this link in a new window

MRI Resources 
Devices - Spine MRI - General - Software - Coils - Mobile MRI Rental
 
Flow Sensitive Alternating Inversion RecoveryInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(FAIR) In this sequence 2 inversion recovery images are acquired, one with a nonselective and the other with a slice selective inversion pulse. The z-magnetization in the first sequence is independent of flow. Inflowing spins give z-magnetization from second pulse. A major signal loss in FAIR is the T1 relaxation of tagged blood in transit to the imaging slice. Sharper edges of the inversion pulse give narrow spacing between the inversion edge and the 1st slice because reduced transit time gives lower T1 relaxation induced signal loss. The difference of the images in a consequence contains information proportional to flow (blood partition coefficient). Standard adiabatic inversion RF pulse does not have good slice-profile, because of power/SAR limitation. A c-shaped frequency offset corrected inversion (FOCI) RF pulse can help to increase the signal.
Perfusion imaging, e.g. myocardial, using tissue water as endogenous contrast is suggested.
spacer
MRI Resources 
Education - Universities - Devices - Corporations - Artifacts - Jobs
 
Chemical Shift ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Chemical shift, black boundary, spatial misregistration, relief
DESCRIPTION
Black or bright band
During frequency encoding, fat protons precess slower than water protons in the same slice because of their magnetic shielding. Through the difference in resonance frequency between water and fat, protons at the same location are misregistrated (dislocated) by the Fourier transformation, when converting MRI signals from frequency to spatial domain. This chemical shift misregistration cause accentuation of any fat-water interfaces along the frequency axis and may be mistaken for pathology. Where fat and water are in the same location, this artifact can be seen as a bright or dark band at the edge of the anatomy.
Protons in fat and water molecules are separated by a chemical shift of about 3.5 ppm. The actual shift in Hertz (Hz) depends on the magnetic field strength of the magnet being used. Higher field strength increases the misregistration, while in contrast a higher gradient strength has a positive effect. For a 0.3 T system operating at 12.8 MHz the shift will be 44.8 Hz compared with a 223.6 Hz shift for a 1.5 T system operating at 63.9 MHz.
mri safety guidance
Image Guidance
For artifact reduction helps a smaller water fat shift (higher bandwidth), a higher matrix, an in phase TE or a spin echo technique. Since the misregistration offset is present in the read out axis the patient may be rescanned with this axis parallel to the fat-water interface. Steeper gradient may be employed to reduce the chemical shift offset in mm. Another strategy is to employ specialized pulse sequences such as fat saturation or inversion recovery imaging. Fat suppression techniques eliminate chemical shift artifacts caused by the lack of fat signal.

See also Black Boundary Artifact and Magnetic Resonance Spectroscopy.
spacer

• View the DATABASE results for 'Chemical Shift Artifact' (7).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
  News & More:
What is chemical shift artefact? Why does it occur? How many Hz at 1.5 T?
   by www.revisemri.com    
Abdominal MRI at 3.0 T: The Basics Revisited
Wednesday, 20 July 2005   by www.ajronline.org    
MRI Resources 
MR Guided Interventions - Mobile MRI Rental - Diffusion Weighted Imaging - Supplies - Spectroscopy pool - Online Books
 
     1 - 5 (of 12)     next
Result Pages : [1]  [2 3]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]