Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Gibbs Artifact' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Gibbs Artifact' found in 1 term [] and 3 definitions [], (+ 2 Boolean[] results
1 - 5 (of 6)     next
Result Pages : [1]  [2]
MRI Resources 
Mobile MRI Rental - Software - DICOM - Research Labs - Directories - MRI Technician and Technologist Career
 
Gibbs ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Gibbs, Gibbs ringing, truncation
DESCRIPTION
Edge ringing, syrinx-like stripe
REASON
Sharp changes in intensity (incomplete digitization of the echo)
HELP
More samples
The Gibbs or ringing artifact appears as a series of lines in the MR image parallel to abrupt and intense changes in the object at this location. This artifact does not occur visibly on smooth objects. This artifact is caused by the Gibbs phenomenon, an overshoot or ringing of Fourier series occurring at discontinuities.
In the spinal cord, a small syrinx can be simulated by the Gibbs phenomenon. Gibbs artifacts are also seen in other regions, for example the brain//skull interface.
Fine lines visible in an image may be due to undersampling of the high spatial frequencies, respectively incomplete digitization of the echo.
With more encoding steps the Gibbs artifacts is less intense and narrower. Therefore, e.g. the artifact is more intense in the 256 point dimension of a 256x512 acquisition matrix.
mri safety guidance
Image Guidance
This problem can only be resolved by smoothing filters (LanczosSigmaFactor, 2-D Exponential Filtering, Gegenbauer Reconstruction etc.) or with a higher acquisition matrix and/or a smaller FOV, to smooth the object.

See also Gibbs Phenomenon and Apodization.
spacer
 
• Share the entry 'Gibbs Artifact':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • Truncation Artifact
    • Ringing Artifact Reduction
    • Number of Excitations
    • Pixel
    • Acquisition Matrix
 
Further Reading:
  Basics:
Lanczos sigma Factor
   by mathworld.wolfram.com    
  News & More:
Troubleshooting the ACR MRI Accreditation Phantom Tests
   by www.aapm.org    
MRI Resources 
Spectroscopy - MRA - Cardiovascular Imaging - Universities - Crystallography - Portals
 
Gibbs Phenomenon
 
In mathematics, the Gibbs phenomenon (also known as ringing artifacts, named after the American physicist J. Willard Gibbs) is the peculiar manner in which the Fourier series of a piecewise continuously differentiable periodic function f behaves at a jump discontinuity: the nth partial sum of the Fourier series has large oscillations near the jump, which might increase the maximum of the partial sum above that of the function itself. The overshoot does not die out as the frequency increases, but approaches a finite limit.
In MRI, artifactual ripples parallel to abrupt and intense changes are caused by the Fourier transformation.

See Gibbs Artifact, Truncation Artifact, Ringing Artifact Reduction.
spacer

• View the DATABASE results for 'Gibbs Phenomenon' (5).Open this link in a new window

 
Further Reading:
  Basics:
Gibbs phenomenon
   by en.wikipedia.org    
MRI Resources 
Stimulator pool - Jobs pool - Breast Implant - MRI Centers - Lung Imaging - Implant and Prosthesis pool
 
Gibbs RingingInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
spacer

• View the DATABASE results for 'Gibbs Ringing' (3).Open this link in a new window

MRI Resources 
Portals - Libraries - Blood Flow Imaging - MRI Training Courses - Pregnancy - Sequences
 
Truncation ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
DESCRIPTION
Edge ringing, syrinx-like stripe
REASON
Sharp changes in intensity (incomplete digitization of the echo)
HELP
Take more samples
A data truncation artifact may occur when the interface between high and low signal intensities is encountered in one imaging plane. The 2D-FT techniques transform the MR signal to spatial intensity image data with frequency and phase information encoding each axis in the plane of the scan. This artifact is found in both frequency and phase axes. Artifactual ripples adjacent to edges in an image or sharp features in a spectrum, caused by omission of higher frequency terms in Fourier transformation, particularly with the use of zero filling to replace unsampled higher frequencies.
Complex shapes are specified by series of sine and cosine waves of various frequencies, phase and amplitude. Some shapes are more difficult to encode than others. The most difficult shapes to represent with Fourier series of terms are waveforms with instantaneous transitions, tissue discontinuities or edges. The low-frequency components of the series describe the overall shape of the step function. Higher frequency components are needed to describe the corners if the step function more accurately. If not enough samples are taken, these areas cannot be accurately represented. The truncation of the infinite data series results in a ringing artifact because of the inability to accurately approximate this tissue discontinuity with a shorter truncated data set. Therefore, the ringing that occurs at all tissue boundaries on MR is called truncation artifact.
mri safety guidance
Image Guidance
This problem can be easily resolved by taking more samples - a higher acquisition matrix and/or a smaller FOV. See Gibbs Artifact and Gibbs Phenomenon.
spacer

• View the DATABASE results for 'Truncation Artifact' (2).Open this link in a new window

 
Further Reading:
  News & More:
Magnetic Resonance Imaging (MRI)
2003   by www.hull.ac.uk    
MRI Resources 
Journals - Corporations - Safety pool - Societies - Safety Training - Jobs pool
 
Ringing Artifact Reduction
 
Multiplication of acquired MR data by a function smoothly tapering off at higher spatial frequencies so as to reduce ringing artifacts near edges in the corresponding image or spectrum due to truncation and Gibbs phenomenon. It is a form of filtering.
spacer

• View the DATABASE results for 'Ringing Artifact Reduction' (2).Open this link in a new window

MRI Resources 
Shielding - Raman Spectroscopy - Artifacts - Pregnancy - Equipment - MRI Training Courses
 
     1 - 5 (of 6)     next
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



How AI will impact MRI :
only diagnostics 
saving time 
reducing cost 
makes planning obsolete 
reduce human knowledge 
not at all 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]