In MRI, an echo is the emission of energy in form of an electromagnetic resonance signal of a nuclei after its excitation. At this point spins are back in phase again and the signal is measured. The desired number of echoes is selectable. Often until eight echoes are permissible for 2D or 3D scans using spin echo, inversion recovery or MIX techniques. Two echoes are permissible for all other techniques. A multi echo imaging sequence is needed for simultaneous measurement of T2 and density weighted images.
From ISOL Technology
'Ultra high field MR system, it's right close to you.
FORTE 3.0T is the new standard for the future ultra high field MR system.
If you are pushing the limits of your existing clinical MR scanner, the FORTE will surely take you to the next level of diagnostic imaging.
FORTE is the core leader of the medical technology in the 21st century. Proving effects of fMRI that cannot be measured with MRI less than 2.0T.'
(IR-FSPGR) This sequence uses an inversion pulse followed by a spoiled low flip angle and short TR gradientecho train to collect all partition data.
The 3D-IR-FSPGR uses centric ordering of Gz. The selected TI is also the effective TI. The centric reordering strategy can lead to image blurring if a steady state is not achieved quickly with low flip angles (less than 20°).