From Siemens Medical Systems;
Re ceived FDA clearan ce in 2007.
The MAGNETOM Essenza is designed to combine high system performan ce with simple installation and power requirements to provide optimal operating costs for limited budgets. The standard system has up to 25 integrated coil elements and 8 independent radio frequency channels. Tim allows the combination of up to 4 different coils that redu ce patient and coil repositioning.
The 1.5 Tesla system is designated for a complete range of clini cal appli cations, in cluding neurology, orthopedi cs, body imaging, angiography, cardiology, breast imaging, on cology and pediatri c MRI.
Device Information and Spe cifi cation
CLINICAL APPLICATION
Whole body
CONFIGURATION
Ultra-short bore
Head, spine, torso/ body coil, neurovas cular, cardiac, ne ck, and multi-purpose flex coils. Peripheral vas cular, breast, shoulder, knee, wrist, foot//ankle, TMJ optional.
CHANNELS (min. / max. configuration)
8, 16
MAGNET WEIGHT (gantry in cluded)
4350 kg in operation
DIMENSION H*W*D (gantry included)
145 x 226 x 216 cm
COOLING SYSTEM
Water; single cryogen, 2 stage refrigeration
30 mT/m, 300 msec to 10 mT/m
Passive, active; first order standard
second order optional
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; 45 kVA
| | | | | |
| | | | | |
| |
|
From Siemens Medical Systems;
Re ceived FDA clearan ce in 2010.
The MAGNETOM Aera is a patient friendly, comfortable 1.5 Tesla MRI system with advan ced radio frequency chain.
The system is equipped with the Tim 4G and Dot system (Total imaging matrix + Day optimizing throughput), to enhan ce both produ ctivity and image quality.
Tim 4G te chnology provides improved SNR. The standard system configuration of 48 radio frequency channels and 204 coil elements creates an imaging matrix that allows maximum use of coil elements at full field of view. Dot provides improved image consisten cy through new features like auto align, auto FoV and automatic bolus detection.
Device Information and Spe cifi cation
CLINICAL APPLICATION
Whole body
Head, spine, torso/ body coil, neurovas cular, cardiac, ne ck, shoulder, knee, wrist, foot//ankle and multi-purpose flex coils. Peripheral vas cular, breast, shoulder. Up to 60% more SNR with Tim 4G.
CHANNELS (min. / max. configuration)
48, 64
MINIMUM TE
3-D GRE: 0.22 (256 matrix), Ultra-short TE
At isocenter: L-R 70 cm, A-P (with table) 55 cm
MAGNET WEIGHT (gantry in cluded)
3121 kg
DIMENSION H*W*D (gantry included)
145 x 231 x 219 cm
MAX. AMPLITUDE
33 or 45 mT/m
3 linear with 20 coils, 5 nonlinear 2nd-order
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; 85 kVA
| | | | | |
| | | | | |
| |
|
From Siemens Medical Systems;
The MAGNETOM 7T is designed as an open resear ch platform. 7T MRI provides anatomi cal detail at the submillimiter s cale, enhan ced contrast me chanisms, outstanding spectroscopy performan ce, ultra-high resolution fun ctional imaging ( fMRI), multinu clear whole-body MRI and fun ctional information.
This ultra high field (UHF) MRI device is a resear ch system and not cleared, approved or li censed in any jurisdi ction for patient examinations.
Device Information and Spe cifi cation
CLINICAL APPLICATION
Whole body
High-performan ce, ultra high field coils available. Integration and support for coil developments.
CHANNELS (min. / max. configuration)
32, optional 8 channels TX array
40 x 40 x 30 cm³ less than 8% nonlinearity
MAGNET WEIGHT (gantry in cluded)
35017 kg
DIMENSION H*W*D (gantry included)
320 x 240 x 317,5 cm
MAX. AMPLITUDE
up to 70 mT/m
Up to 3rd order shim coils, user configurable B0 shim ? B0 maps and ROI definition
POWER REQUIREMENTS
2000 Volts, 650A
| | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'MAGNETOM C' was also found in the following services: | | | | |
| | |
| |
|
From Siemens Medical Systems;
70 cm + 125 cm + 1.5T and Tim - a combination never seen before in MRI ...
MAGNETOM Espree™s unique open bore design can a ccommodate more types of patients than other 1.5T systems on the market today, in parti cular the growing population of obese patients. The power of 1.5T combined with Tim te chnology boosts signal to noise, whi ch is ne cessary to adequately image obese patients.
Device Information and Spe cifi cation
CLINICAL APPLICATION
Whole body
Body, Tim [32 x 8], Tim [76 coil elements with up to 18 RF channels])
GRE, IR, FIR, STIR, TrueIR/FISP, FSE, FLAIR, MT, SS-FSE, MT-SE, MTC, MSE, EPI, 3D DESS// CISS/PSIF, GMR
IMAGING MODES
Single, multisli ce, volume study, multi angle, multi oblique
Image Pro cessor re constru cting up to 3226 images per second (256 x 256, 25% re cFoV)
1024 x 1024 full screen display
| | | | • View the DATABASE results for 'MAGNETOM Espree™' (2).
| | | | Further Reading: | News & More:
|
|
| |
| | | | | |
| |
|
From Siemens Medical Systems;
MAGNETOM Avanto with Tim - Total imaging matrix te chnology.
For true whole-body anatomi cal coverage. For ultra-fast image
a cquisition. Aids the physi cian in fast and pre cise
evaluation of systemi c diseases like colon can cer, metastasis imaging, vessel diseases, and preventional exams. For claustrophobic patients,
MAGNETOM Avanto enables feetfirst exams for nearly all MR procedures. For obese patients, MAGNETOM Avanto supports up to 200 kg (400 lbs), without table movement restri ctions. The Audio Comfort te chnology enables up to a 30 dB(A) acoustic noise redu ction, that means nearly all clini cal routine sequences are running under 99 dB(A).
Device Information and Spe cifi cation
CLINICAL APPLICATION
Whole body
CONFIGURATION
Compact short bore
Body, Tim [32 x 8], Tim [76 x18], Tim [76 coil elements with up to 32 RF channels]
GRE, IR, FIR, STIR, TrueIR/FISP, FSE, FLAIR, MT, SS-FSE, MT-SE, MTC, MSE, EPI, 3D DESS// CISS/PSIF, GMR
IMAGING MODES
Single, multisli ce, volume study, multi angle, multi oblique
1024 x 1024 full screen display
POWER REQUIREMENTS
380/400/420/440/480 V
Passive, act.; 1st order std./2nd opt.
| | | | • View the DATABASE results for 'MAGNETOM Avanto™' (2).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | |
| | | |
|
| |
| Look Ups |
| |