FDA cleared and CE Mark 2011.
The Biograph mMR has a fully-integrated design for simultaneous PET/MRI imaging. The dedicated hardware includes solid-state, avalanche photodiode PET detector and adapted, PET-compatible MR coils.
The possibility of truly simultaneous operation allows the acquisition of several magnetic resonance imaging ( MRI) sequences during the positron emission tomography (PET) scan, without increasing the examination time.
See also Hybrid Imaging.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Simultaneous PET/MRI
26 cm (typical overlap 23%)
A-P 45, R-L 50, H-F 50 cm
PET RING DIAMETER
65.6 cm
PATIENT SCAN RANGE
199 cm
HORIZONTAL SPEED
200 mmsec
PET DETECTOR
Solid state, 4032 avalanche photo diodes
DETECTOR SCINTILLATION MATERIAL
LSO, 28672 crystals
CRYSTAL SIZE
4 x 4 x 20 mm
DIMENSION H*W*D (gantry included)
335 x 230 x 242 cm (finshed covers)
COOLING SYSTEM
PET system: water; MRI system: water
Aautomatic, patient specific shim; active shim 3 linear and 5 non-linear channels (seond order)
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; Total system 110kW
| | | | | • Share the entry 'Biograph mMR': | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
| | | | • View the DATABASE results for 'Coil' (193).
| | | • View the NEWS results for 'Coil' (11).
| | | | Further Reading: | News & More:
|
|
3T MRI-Compatible Small Body Array Gets FDA Nod Monday, 15 January 2024 by www.diagnosticimaging.com | | |
High-field MRI Coils – that work, superbly, even at 750 MHz
by www.dotynmr.com | | |
New superconducting coil improves MRI performance Wednesday, 20 July 2016 by www.eurekalert.org | | |
First European fully superconductive coil reaching a magnetic field of 25 Tesla produced Wednesday, 8 June 2016 by www.news-medical.net | | |
ScanMed Introduces a Groundbreaking New Orbit and Mandible Array to the MRI Market Wednesday, 26 March 2014 by www.digitaljournal.com | | |
MRI coil optimized for imaging of infants gets FDA clearance Tuesday, 17 April 2012 by www.medcitynews.com | | |
Low-profile MRI coil company inching toward commercialization Wednesday, 21 December 2011 by www.medcitynews.com | | |
High-Resolution Uniform MR Imaging of Finger Joints Using a Dedicated RF Coil at 3 Tesla Sunday, 31 January 2010 by www.ncbi.nlm.nih.gov | | |
DOTmed Industry Sector Report: MRI Coil Sales & Service Companies Thursday, 13 December 2007 by www.dotmed.com |
|
| |
| | | | | |
| |
|
| | | | • View the DATABASE results for 'Coil Diameter' (3).
| | | | |
| | | Searchterm 'MRI Coil' was also found in the following services: | | | | |
| | |
| |
|
| | | | • View the DATABASE results for 'Coil Loading' (2).
| | | | |
| | | | | |
| |
|
| | | | • View the DATABASE results for 'Element' (40).
| | | | |
| | | | |
| |
| | 1 - 5 (of 27) nextResult Pages : [1 2] [3 4 5 6] |
| |
|
| |
| Look Ups |
| |