Brain imaging, magnetic resonance imaging of the head or skull, cranial magnetic resonance tomography (MRT), neurological MRI - they describe all the same radiological imaging technique for medical diagnostic.
Magnetic resonance imaging of the human brain includes the anatomic description and the detection of lesions. Special techniques like diffusion weighted imaging, functional magnetic resonance imaging ( fMRI) and spectroscopy provide also information about the function and chemical metabolites of the brain.
MRI provides detailed pictures of brain and nerve tissues in multiple planes without obstruction by overlying bones. Brain MRI is the procedure of choice for most brain disorders. It provides clear images of the brainstem and posterior brain, which are difficult to view on a CT scan. It is also useful for the diagnosis of demyelinating disorders (disorders such as multiple sclerosis (MS) that cause destruction of the myelin sheath of the nerve).
With this noninvasive procedure also the evaluation of blood flow and the flow of cerebrospinal fluid (CSF) is possible. Different MRA methods, also without contrast agents can show a venous or arterial angiogram. MRI can distinguish tumors, inflammatory lesions, and other pathologies from the normal brain anatomy. However, MRI scans are also used instead other methods to avoid the dangers of interventional procedures like angiography (DSA - digital subtraction angiography) as well as of repeated exposure to radiation as required for computed tomography (CT) and other X-ray examinations.
A ( birdcage) bird cage coil achieves uniform excitation and reception and is commonly used to study the brain. Usually a brain MRI procedure includes FLAIR, T2 weighted and T1 weighted sequences in two or three planes. See also Fetal MRI, Fluid Attenuation Inversion Recovery ( FLAIR), Perfusion Imaging and High Field MRI. See also Arterial Spin Labeling. | | | | | | | | | | | | | Further Reading: | | Basics:
|
|
News & More:
| |
MRI Reveals Significant Brain Abnormalities Post-COVID Monday, 21 November 2022 by neurosciencenews.com | | |
Combining genetics and brain MRI can aid in predicting chances of Alzheimer's disease Wednesday, 29 June 2022 by www.sciencedaily.com | | |
Roundup: How Even Mild COVID Can Affect the Brain; This Many Daily Steps Improves Longevity; and More Friday, 11 March 2022 by baptisthealth.net | | |
A low-cost and shielding-free ultra-low-field brain MRI scanner Tuesday, 14 December 2021 by www.nature.com | | |
Large International Study Reveals Spectrum of COVID-19 Brain Complications Tuesday, 9 November 2021 by www.itnonline.com | | |
Brain MRI-Based Subtypes of MS Predict Disability Progression, Treatment Response Thursday, 13 May 2021 by www.neurologyadvisor.com | | |
New MRI method improves detection of disease changes in the brain's network Thursday, 11 June 2020 by www.compute.dtu.dk | | |
New NeuroCOVID Classification System Uses MRI to Categorize Patients Friday, 12 June 2020 by www.diagnosticimaging.com | | |
New MRI technique can 'see' molecular changes in the brain Thursday, 5 September 2019 by medicalxpress.com | | |
Talking therapy or medication for depression: Brain scan may help suggest better treatment Monday, 27 March 2017 by www.newsnation.in | | |
MRI identifies brain abnormalities in chronic fatigue syndrome patients Wednesday, 29 October 2014 by www.eurekalert.org | | |
MRIs Useful in Tracking Depression in MS Patients Tuesday, 1 July 2014 by www.hcplive.com | | |
Contrast agent linked with brain abnormalities on MRI Tuesday, 17 December 2013 by www.sciencecodex.com | | |
MRIs Reveal Signs of Brain Injuries Not Seen in CT Scans Tuesday, 18 December 2012 by www.sciencedaily.com | | |
Iron Deposits in the Brain May Be Early Indicator of MS Wednesday, 13 November 2013 by www.healthline.com | | |
Migraine Sufferers Have Thicker Brain Cortex Tuesday, 20 November 2007 by www.medicalnewstoday.com |
|
| |
| | | | | |
| |
|
The company changed its name from Advanced Magnetics, Inc. to AMAG Pharmaceuticals, Inc. in July 2007.
AMAG Pharmaceuticals, Inc., a biopharmaceutical company, developed and manufactured organ-specific diagnostic contrast agents that provide clearer images during magnetic resonance imaging ( MRI) tests used to detect tumors and other abnormalities.
The company had two MRI related products on the market: Feridex I.V. (for the diagnosis of liver lesions) and GastroMARK (used for bowel and abdominal MR imaging). In November 2008, AMAG Pharmaceuticals, Inc. decided to discontinue the manufacturing of Feridex. The development of Combidex as a contrast agent for lymph disease has also been stopped.
The Company has now two commercial products: Feraheme® and GastroMARK®.
Feraheme® is the trade name of Ferumoxytol (formerly Code 7228) and is indicated for the treatment of iron deficiency anemia. Feraheme® is also being developed as a diagnostic agent for vascular-enhanced magnetic resonance imaging ( MRI) to assess peripheral arterial disease.
MRI Contrast Agents:
Contact Information
MAIL
AMAG Pharmaceuticals, Inc.
61 Mooney St.
Cambridge, MA 02138
USA
| | | | • View the DATABASE results for 'AMAG Pharmaceuticals, Inc.' (10).
| | | • View the NEWS results for 'AMAG Pharmaceuticals, Inc.' (7).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
An image artifact is a structure not normally present but visible as a result of a limitation or malfunction in the hardware or software of the MRI device, or in other cases a consequence of environmental influences as heat or humidity or it can be caused by the human body (blood flow, implants etc.). The knowledge of MRI artifacts (brit. artefacts) and noise producing factors is important for continuing maintenance of high image quality. Artifacts may be very noticeable or just a few pixels out of balance but can give confusing artifactual appearances with pathology that may be misdiagnosed.
Changes in patient position, different pulse sequences, metallic artifacts, or other imaging variables can cause image distortions, which can be reduced by the operator; artifacts due to the MR system may require a service engineer.
Many types of artifacts may occur in magnetic resonance imaging. Artifacts in magnetic resonance imaging are typically classified as to their basic principles, e.g.:
•
Physiologic (motion, flow)
•
Hardware (electromagnetic spikes, ringing)
Several techniques are developed to reduce these artifacts (e.g. respiratory compensation, cardiac gating, eddy current compensation) but sometimes these effects can also be exploited, e.g. for flow measurements.
See also the related poll result: ' Most outages of your scanning system are caused by failure of'
| | | | • View the DATABASE results for 'Artifact' (166).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'Magnetic Resonance Imaging' was also found in the following services: | | | | |
| | |
| |
|
(MR mammography) Magnetic resonance imaging of the breast is particularly useful in evaluation of newly diagnosed breast cancer, in women whose breast tissue is mammographically very dense and for screening in women with a high lifetime risk of breast cancer because of their family history or genetic disposition.
Breast MRI can be performed on all standard whole body magnets at a field strength of 0.5 T - 1.5 Tesla. Powerful gradient strengths over 15 mT/m will help to improve the balance between spatial resolution, scanning speed, and volume coverage. The use of a dedicated bilateral breast coil is obligatory.
Malignant lesions release angiogenic factors that increase local vessel density and vessel permeability. Breast cancer is detectable due to the strong enhancement in dynamic breast imaging that peaks early (about 1-2 min.) after contrast medium injection. If breast cancer is suspected, a breast biopsy may be necessary to secure the diagnosis. See also Magnetic Resonance Imaging MRI, Biopsy and MR Guided Interventions.
Requirements in breast MRI procedures:
•
Both breasts must be measured without gaps.
•
For the best possible detection of enhancement fat signal should be eliminated either by image subtraction or by
spectrally selective fat saturation.
•
Thin slices are necessary to assure absence of partial
volume effects.
•
Imaging should be performed with a spatial
resolution in plane less than 1 mm.
For Ultrasound Imaging (USI) see Breast Ultrasound at Medical-Ultrasound-Imaging.com.
See also the related poll result: ' MRI will have replaced 50% of x-ray exams by' | | | | | | | | | | | • View the DATABASE results for 'Breast MRI' (13).
| | | • View the NEWS results for 'Breast MRI' (41).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
Technology advances in breast cancer screenings lead to early diagnosis Friday, 6 October 2023 by ksltv.com | | |
Are synthetic contrast-enhanced breast MRI images as good as the real thing? Friday, 18 November 2022 by healthimaging.com | | |
Abbreviated breast MRI protocols not as cost-effective as promised, new study shows Wednesday, 20 July 2022 by healthimaging.com | | |
Deep learning poised to improve breast cancer imaging Thursday, 24 February 2022 by www.eurekalert.org | | |
Pre-Operative Breast MRI Can Help Identify Patients Likely to Experience Nipple-Sparing Mastectomy Risks Wednesday, 7 April 2021 by www.diagnosticimaging.com | | |
Breast cancer screening recalls: simple MRI measurement could avoid 30% of biopsies Monday, 1 March 2021 by www.eurekalert.org | | |
A Comparison of Methods for High-Spatial-Resolution Diffusion-weighted Imaging in Breast MRI Tuesday, 25 August 2020 by pubs.rsna.org | | |
Pre-Operative Breast MRI Diagnoses More Cancers in Women with DCIS Thursday, 9 July 2020 by www.diagnosticimaging.com | | |
Breast MRI and tumour biology predict axillary lymph node response to neoadjuvant chemotherapy for breast cancer Thursday, 26 December 2019 by cancerimagingjournal.biomedcentral.com | | |
Breast MRI Coding Gets an Overhaul in 2019 Wednesday, 9 January 2019 by www.aapc.com | | |
How accurate are volumetric software programs when compared to breast MRI? Thursday, 27 July 2017 by www.radiologybusiness.com | | |
Additional Breast Cancer Tumors Found on MRI After Mammography May Be Larger, More Aggressive Wednesday, 9 December 2015 by www.oncologynurseadvisor.com | | |
Preoperative MRI May Overdiagnose Contralateral Breast Cancer Wednesday, 2 December 2015 by www.cancertherapyadvisor.com | | |
BI-RADS and breast MRI useful in predicting malignancy Wednesday, 30 May 2012 by www.oncologynurseadvisor.com |
|
| |
| | | | | |
| |
|
(EPI) Echo planar imaging is one of the early magnetic resonance imaging sequences (also known as Intascan), used in applications like diffusion, perfusion, and functional magnetic resonance imaging. Other sequences acquire one k-space line at each phase encoding step. When the echo planar imaging acquisition strategy is used, the complete image is formed from a single data sample (all k-space lines are measured in one repetition time) of a gradient echo or spin echo sequence (see single shot technique) with an acquisition time of about 20 to 100 ms.
The pulse sequence timing diagram illustrates an echo planar imaging sequence from spin echo type with eight echo train pulses. (See also Pulse Sequence Timing Diagram, for a description of the components.)
In case of a gradient echo based EPI sequence the initial part is very similar to a standard gradient echo sequence. By periodically fast reversing the readout or frequency encoding gradient, a train of echoes is generated.
EPI requires higher performance from the MRI scanner like much larger gradient amplitudes. The scan time is dependent on the spatial resolution required, the strength of the applied gradient fields and the time the machine needs to ramp the gradients.
In EPI, there is water fat shift in the phase encoding direction due to phase accumulations. To minimize water fat shift (WFS) in the phase direction fat suppression and a wide bandwidth (BW) are selected. On a typical EPI sequence, there is virtually no time at all for the flat top of the gradient waveform. The problem is solved by "ramp sampling" through most of the rise and fall time to improve image resolution.
The benefits of the fast imaging time are not without cost. EPI is relatively demanding on the scanner hardware, in particular on gradient strengths, gradient switching times, and receiver bandwidth. In addition, EPI is extremely sensitive to image artifacts and distortions. | | | | • View the DATABASE results for 'Echo Planar Imaging' (19).
| | | • View the NEWS results for 'Echo Planar Imaging' (1).