| Info Sheets |
| | | | | | | | | | | | | | | | | | | | | | | | |
| Out- side |
| | | | |
|
| | | 'Magnetic Susceptibility' | |
Result : Searchterm 'Magnetic Susceptibility' found in 1 term [] and 15 definitions [], (+ 10 Boolean[] results
| previous 21 - 25 (of 26) nextResult Pages : [1] [2 3 4] [5 6] | | | | Searchterm 'Magnetic Susceptibility' was also found in the following services: | | | | |
| | |
| |
|
| | | | | | | | | Further Reading: | News & More:
|
|
| |
| | | | | |
| |
|
A pacemaker is a device for internal or external battery-operated cardiac pacing to overcome cardiac arrhythmias or heart block. All implanted electronic devices are susceptible to the electro magnetic fields used in magnetic resonance imaging. Therefore, the main magnetic field, the gradient field, and the radio frequency (RF) field are potential hazards for cardiac pacemaker patients.
The pacemaker's susceptibility to static field and its critical role in life support have warranted special consideration. The static magnetic field applies force to magnetic materials. This force and torque effects rise linearly with the field strength of the MRI machines. Both, RF fields and pulsed gradients can induce voltages in circuits or on the pacing lead, which will heat up the tissue around e.g. the lead tip, with a potential risk of thermal injury.
Regulations for pacemakers provide that they have to switch to the magnet mode in static magnetic fields above 1.0 mT. In MR imaging, the gradient and RF fields may mimic signals from the heart with inhibition or fast pacing of the heart. In the magnet mode, most of the current pacemakers will pace with a fix pulse rate because they do not accept the heartsignals. However, the state of an implanted pacemaker will be unpredictable inside a strong magnetic field. Transcutaneous controller adjustment of pacing rate is a feature of many units. Some achieve this control using switches activated by the external application of a magnet to open/close the switch. Others use rotation of an external magnet to turn internal controls. The fringe field around the MRI magnet can activate such switches or controls. Such activations are a safety risk.
Areas with fields higher than 0.5 mT ( 5 Gauss Limit) commonly have restricted access and/or are posted as a safety risk to persons with pacemakers.
A Cardiac pacemaker is because the risks, under normal circumstances an absolute contraindication for MRI procedures.
Nevertheless, with special precaution the risks can be lowered. Reprogramming the pacemaker to an asynchronous mode with fix pacing rate or turning off will reduce the risk of fast pacing or inhibition. Reducing the SAR value reduces the potential MRI risks of heating. For MRI scans of the head and the lower extremities, tissue heating also seems to be a smaller problem. If a transmit receive coil is used to scan the head or the feet, the cardiac pacemaker is outside the sending coil and possible heating is very limited. | | | | • View the DATABASE results for 'Cardiac Pacemaker' (6).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Categories of negative oral contrast agents:
Negative oral contrast media are usually based on superpara magnetic particles and act by inducing local field inhomogeneities, which results in shortening of both T1 and T2 relaxation times. Superparamagnetic contrast agents have predominant T2 weighted effects.
Biphasic contrast media are agents that have different signal intensities on different sequences, depending on the concentration at which they are used.
Suitable materials for oral contrast agents should have little or no absorption by the stomach or intestines, complete excretion, no motion or susceptibility artifacts, affordability, and uniform marking of the gastrointestinal tract.
Benefits of negative oral contrast agents are the reduction of ghosting artifacts caused by the lack of signal. Superpara magnetic iron oxides produce also in low concentrations a noticeable signal loss; but can generate susceptibility artifacts especially in gradient echo sequences. Perfluorochemicals do not dilute in the bowel because they are not miscible with water.
High cost, poor availability, and limited evaluations of side effects are possible disadvantages.
Negative oral contrast agents are used e.g., in MRCP, where the ingestion of 600-900 ml of SPIO cancels out the signal intensity of the lumen (in addition after the injection of a gadolinium-based contrast medium, the enhancement of the inflammatory tissues is clearer seen), and in MR abdominal imaging of Crohn's disease in combination with mannitol.
| | | | • View the DATABASE results for 'Negative Oral Contrast Agents' (7).
| | | | Further Reading: | Basics:
|
|
| |
| | | Searchterm 'Magnetic Susceptibility' was also found in the following services: | | | | |
| | |
| |
|
Diamagnetism occurs only by a substance in the presence of an externally applied magnetic field. Dia magnetic contrast agents are complexes in which the metal ion (e.g., Zn, Bi and Ca) is dia magnetic.
Potential dia magnetic materials in gastrointestinal MRI:
A suspension of clay minerals (Kaopectate with kaolin, a common over the counter drug) can be used as a negative oral contrast agent caused by the dia magnetic properties.
By using this preparation as a gastrointestinal contrast agent e.g., in pancreas MRI or MRCP, the absence of signal is clearly visible in the stomach and duodenum.
Barium sulfate commonly used as an X-ray contrast agent has also been tested for use in abdominal imaging. The dia magnetic properties of the barium particles are caused by a susceptibility effect around them, the resulting signal loss is strengthening by a replacement of water protons with barium.
See also Diamagnetism. | | | | • View the DATABASE results for 'Gastrointestinal Diamagnetic Contrast Agents' (7).
| | | | |
| | | | | |
| |
|
Ferromagnetism is a phenomenon by which a material can exhibit a spontaneous magnetization: a net magnetic moment in the absence of an external magnetic field. More recently: a material is ferro magnetic, only if all of its magnetic ions add a positive contribution to the net magnetization (for differentiation to ferri magnetic and antiferro magnetic materials). If some of the magnetic ions subtract from the net magnetization (if they are partially anti-aligned), then the material is ferri magnetic. If the ions anti-align completely so as to have zero net magnetization, despite the magnetic ordering, then it is an antiferromagnet. All of these alignment effects only occur at temperatures below a certain critical temperature, called the Curie temperature (for ferromagnets and ferrimagnets) or the Néel temperature (for antiferromagnets). Typical ferromagnetic materials are iron, cobalt, and nickel.
In MRI ferromagnetic objects, even very small ones, as implants or incorporations distort the homogeneity of the main magnetic field and cause susceptibility artifacts. | | | | • View the DATABASE results for 'Ferromagnetism' (7).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | | |
| | | |
|
| |
| Look Ups |
| |