Motion Artifact | |
| |
|
Quick Overview
Please note that there are different common names for this artifact.
NAME
Motion, phase encoded motion, instability, smearing
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic
Patient motion is the largest physiological effect that causes artifacts, often resulting from involuntary movements (e.g. respiration, cardiac motion and blood flow, eye movements and swallowing) and minor subject movements.
Movement of the object being imaged during the sequence results in inconsistencies in phase and amplitude, which lead to blurring and ghosting. The nature of the artifact depends on the timing of the motion with respect to the acquisition. Causes of motion artifacts can also be mechanical vibrations, cryogen boiling, large iron objects moving in the fringe field (e.g. an elevator), loose connections anywhere, pulse timing variations, as well as sample motion. These artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
| | | | | • Share the entry 'Motion Artifact': | | | | | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'Motion Artifact' was also found in the following service: | | | | |
| | |
| |
|
Quick Overview
HELP
cardiac and respiratory synchronization
Movement of the heart causes blurring and ghosting in the images. The artifacts appear in the phase encoding direction, independent of the direction of the motion.
Image Guidance
These artifacts can be reduced by using cardiac synchronization: triggering, gating or retrospective triggering. Maximum reduction can be achieved by using triggering in combination with flow compensation, respiratory triggering or breath hold and regional saturation techniques.
See also Motion Artifact. | | | | | | • View the DATABASE results for 'Cardiac Motion Artifact' (2).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | | | |
| |
|
Quick Overview
Please note that there are different common names for this artifact.
REASON
Movement of the imaged object
HELP
Compensation techniques, more averages, anti spasmodic, presaturation
This artifact is caused by movements of the patient or organic processes taking place in the body of the patient.
The artifact appears as bright noise, repeating densities or ghosting in the phase encoding direction.
Image Guidance
| | | | • View the DATABASE results for 'Phase Encoded Motion Artifact' (5).
| | | | |
| | | Searchterm 'Motion Artifact' was also found in the following services: | | | | |
| | |
| |
|
| | | | • View the DATABASE results for 'Motion Artifact Suppression Technique' (3).
| | | | Further Reading: | News & More:
|
|
| |
| | | Searchterm 'Motion Artifact' was also found in the following service: | | | | |
| | |
| |
|
(SMART) Movement of the chest and abdominal wall also causes blurring and ghosting in the images due to variation of signal during data collection. Using SMART averaging is one way of reducing these artifacts but is not always possible or effective enough. | | | | • View the DATABASE results for 'Shimadzu Motion Artifact Reduction Technique' (2).
| | | | |
| | | | |
| | 1 - 5 (of 24) nextResult Pages : [1] [2 3 4 5] |
| |
|
| |
| Look Ups |
| |