Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Readout Gradient' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Readout Gradient' found in 1 term [] and 8 definitions [], (+ 10 Boolean[] results
previous     16 - 19 (of 19)     
Result Pages : [1]  [2]  [3 4]
MRI Resources 
Jobs - Jobs pool - Musculoskeletal and Joint MRI - Mobile MRI Rental - Implant and Prosthesis pool - Spectroscopy pool
 
FID Signal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
FID (with spin echo) signal
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A combination of B1 inhomogeneity, poor slice profile, and insufficient spoiler gradients between the refocusing pulse and the readout interval of a spin echo sequence results in a FID signal being detected along with the echo.
Since the FID is not phase encoded (normally the phase encoding occurs before the refocusing pulse), it is not dispersed along the phase encoding axis, but appears as a line across the center of the image.
mri safety guidance
Image Guidance
If the problem persists, it must be addressed by a service representative.
spacer
 
Further Reading:
  News & More:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Cardiovascular Imaging - RIS - Safety Products - Knee MRI - Corporations - General
 
Receiver turn on ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Receiver turn on
DESCRIPTION
Line across the center of the image
REASON
Combination of problems
HELP
Call the service
A receiver turn on artifact (not a true MR signal) appears similarly like a FID signal artifact, except that they extend into the signal-free region, while the FID (a real MR signal) is confined to the projection of the sample along the readout axis.
mri safety guidance
Image Guidance
FID artifacts cannot occur on gradient echo images, but receiver turn on is still possible on gradient echoes.

See also FID Signal Artifact.
spacer
 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Musculoskeletal and Joint MRI - Anatomy - Breast MRI - MRI Physics - Supplies - Chemistry
 
ViewForum -
related threads
 
The ensemble of raw data points collected during the signal readout. For example, during Cartesian sampling, normally used to refer to data collected with a fixed value of the phase encoding gradient. Also referred to as one line of k-space. In projection reconstruction, the line is radial oriented, while in spiral imaging, it is a spiral.
spacer

• View the DATABASE results for 'View' (160).Open this link in a new window


• View the NEWS results for 'View' (20).Open this link in a new window.
 
Further Reading:
  Basics:
Image Characteristics and Quality
   by www.sprawls.org    
MRI Resources 
Breast Implant - Devices - Societies - MRI Physics - Shielding - Collections
 
Phase Encoding
 
The process of locating a MR signal by altering the phase of spins in one dimension with a pulsed magnetic field gradient along that dimension prior to the acquisition of the signal.
If a gradient field is briefly switched on and then off again at the beginning of the pulse sequence right after the radio frequency pulse, the magnetization of the external voxels will either precess faster or slower relative to those of the central voxels.
During readout of the signal, the phase of the xy-magnetization vector in different columns will thus systematically differ. When the x- or y- component of the signal is plotted as a function of the phase encoding step number n and thus of time n TR, it varies sinusoidally, fast at the left and right edges and slow at the center of the image. Voxels at the image edges along the phase encoding direction are thus characterized by a higher 'frequency' of rotation of their magnetization vectors than those towards the center.
As each signal component has experienced a different phase encoding gradient pulse, its exact spatial reconstruction can be specifically and precisely located by the Fourier transformation analysis. Spatial resolution is directly related to the number of phase encoding levels (gradients) used. The phase encoding direction can be chosen, e.g. whenever oblique MR images are acquired or when exchanging frequency and phase encoding directions to control wrap around artifacts.
spacer

• View the DATABASE results for 'Phase Encoding' (73).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
MRI Resources 
Diffusion Weighted Imaging - Libraries - Open Directory Project - Examinations - Veterinary MRI - Cochlear Implant
 
previous      16 - 19 (of 19)     
Result Pages : [1]  [2]  [3 4]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]