FDA cleared and CE Mark 2011.
The Biograph mMR has a fully-integrated design for simultaneous PET/MRI imaging. The dedicated hardware includes solid-state, avalanche photodiode PET detector and adapted, PET-compatible MR coils.
The possibility of truly simultaneous operation allows the acquisition of several magnetic resonance imaging ( MRI) sequences during the positron emission tomography (PET) scan, without increasing the examination time.
See also Hybrid Imaging.
Device Information and Specification
CLINICAL APPLICATION
Whole Body
CONFIGURATION
Simultaneous PET/MRI
26 cm (typical overlap 23%)
A-P 45, R-L 50, H-F 50 cm
PET RING DIAMETER
65.6 cm
PATIENT SCAN RANGE
199 cm
HORIZONTAL SPEED
200 mmsec
PET DETECTOR
Solid state, 4032 avalanche photo diodes
DETECTOR SCINTILLATION MATERIAL
LSO, 28672 crystals
CRYSTAL SIZE
4 x 4 x 20 mm
DIMENSION H*W*D (gantry included)
335 x 230 x 242 cm (finshed covers)
COOLING SYSTEM
PET system: water; MRI system: water
Aautomatic, patient specific shim; active shim 3 linear and 5 non-linear channels (seond order)
POWER REQUIREMENTS
380 / 400 / 420 / 440 / 460 / 480 V, 3-phase + ground; Total system 110kW
| |  | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
|  |  | Searchterm 'Resonance' was also found in the following services: | | | | |
|  |  |
| |
|
| | | |  | | • View the DATABASE results for 'Black Blood MRA' (6).
| | | | |
|  | |  |  |  |
| |
|
| |  | | • View the DATABASE results for 'Blood Pool Agents' (16).
| | | • View the NEWS results for 'Blood Pool Agents' (1).
| | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
|  |  | Searchterm 'Resonance' was also found in the following services: | | | | |
|  |  |
| |
|
When a group of spins is placed in a magnetic field, each spin aligns in one of the two possible orientations. The relative numbers of spins with different alignments will be given by the Boltzmann distribution.
Definition: if a system of particles, which are able to exchange energy in collisions is in thermal equilibrium, then the relative number ( population) of particles, N1 and N2, in two particular energy levels with corresponding energies, E1 and E2, is given by N1/N2 = exp [-(E1 - E2)/kT] where k is the Boltzmann constant and T is the absolute temperature.
For example, in NMR of protons at room temperature in a magnetic field of 0.25 tesla, the difference in relative numbers of spins aligned with the magnetic field and against the field is about one part in a million; the small excess of nuclei in the lower energy state is the basis of the net magnetization and the resonance phenomenon. | |  | | • View the DATABASE results for 'Boltzmann Distribution' (4).
| | | | |
|  |  | Searchterm 'Resonance' was also found in the following services: | | | | |
|  |  |
| |
|
'Next generation MRI system 1.5T CHORUS developed by ISOL Technology is optimized for both clinical diagnostic imaging and for research development.
CHORUS offers the complete range of feature oriented advanced imaging techniques- for both clinical routine and research. The compact short bore magnet, the patient friendly design and the gradient technology make the innovation to new degree of perfection in magnetic resonance.'
Device Information and Specification
CLINICAL APPLICATION
Whole body
Spin Echo, Gradient Echo, Fast Spin Echo,
Inversion Recovery ( STIR, Fluid Attenuated Inversion Recovery), FLASH, FISP, PSIF, Turbo Flash ( MPRAGE ),TOF MR Angiography, Standard echo planar imaging package (SE-EPI, GE-EPI), Optional:
Advanced P.A. Imaging Package (up to 4 ch.), Advanced echo planar imaging package,
Single Shot and Diffusion Weighted EPI, IR/FLAIR EPI
STRENGTH
20 mT/m (Upto 27 mT/m)