Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Slice' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Slice' found in 19 terms [] and 159 definitions []
previous     16 - 20 (of 178)     next
Result Pages : [1 2 3 4]  [5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
Searchterm 'Slice' was also found in the following services: 
spacer
News  (6)  Resources  (5)  Forum  (25)  
 
Multi Slice Off Resonance Fat SuppressionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
spacer
Searchterm 'Slice' was also found in the following services: 
spacer
Radiology  (24) Open this link in a new windowUltrasound  (9) Open this link in a new window
Contiguous Slice with Variable Echo MultiplanarInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(CSVEMP) MR imaging pulse sequence in which signals for multiple contiguous slices at multiple echoes are collected.
spacer
MRI Resources 
Safety Training - Intraoperative MRI - Homepages - Mass Spectrometry - Portals - Corporations
 
Multiple Overlapping Thin Slab (Slice) Acquisition
 
(MOTSA) This technique combines the best features of 2D time of flight angiography (2D TOF) and 3D TOF MRA. The MOTSA technique consists of multiple 2 cm thick 3D TOF slabs (which minimize saturation effects for through plane flow) combine to provide unlimited coverage similar to multiple 2D TOF slices. High resolution imaging of the carotid arteries is possible when image quality is of greater concern than acquisition time. Images with 1 mm (or less) spatial resolution in all three planes are required. The slabs typically overlap 25-40 to minimize the venetian blind artifact venetian blind artifact due to minimal saturation effects. MOTSA is an useful technique for the evaluation of vertebrobasilar ischemia and aneurysm scanning from the foramen magnum through the circle of Willis.
 
Images, Movies, Sliders:
 Circle of Willis, Time of Flight, MIP  Open this link in a new window
    
SlidersSliders Overview

 
spacer
Searchterm 'Slice' was also found in the following services: 
spacer
News  (6)  Resources  (5)  Forum  (25)  
 
Simultaneous Multislice Acquisition Using Rosette Trajectories
 
(SMART) The acronym for slice specific, multi angle, multi resolution, multi thickness scanning. This function allows the operator to individually customize the thickness, field of view and position of each slice in a multi angle study.
spacer
Searchterm 'Slice' was also found in the following services: 
spacer
Radiology  (24) Open this link in a new windowUltrasound  (9) Open this link in a new window
Crosstalk (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Crosstalk, cross excitation
DESCRIPTION
Anomalous intensities in an image
REASON
HELP
Slice gap minimum 10%
Crosstalk is an artifact introduced into images by interference between adjacent slices of a scan, caused by a slice profile that is not ideal due to the constraints of the measurement technology. If the slice distances are too small, there is cross talk between the slices, which can affect T1 contrast.
mri safety guidance
Image Guidance
This artifact can be eliminated by limiting the minimum spacing (for the most sequences a minimum gap 10% and for IR sequences 20%) between the slices. Crosstalk can also be reduced by selection of interleaved slices (so a slice gap will not be necessary), but interleaved data acquisition can produce large mean intensity differences between adjacent slices.
spacer

• View the DATABASE results for 'Crosstalk (Artifact)' (2).Open this link in a new window

 
Further Reading:
  Basics:
MRI GLOSSARY
   by fonar.com    
Slice-overlap Artifacts
   by www.mritutor.org    
MRI Resources 
Image Quality - Libraries - Service and Support - Raman Spectroscopy - Journals - Homepages
 
previous      16 - 20 (of 178)     next
Result Pages : [1 2 3 4]  [5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]