Aliasing ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Aliasing, backfolding, foldover, phase wrapping, wrap around
DESCRIPTION
Image wrap around
Aliasing is an artifact that occurs in MR images when the scanned body part is larger than field of view (FOV). As a consequence of the acquired k-space frequencies not being sampled densely enough, whereby portions of the object outside of the desired FOV get mapped to an incorrect location inside the FOV. The cyclical property of the Fourier transform fills the missing data of the right side with data from behind the FOV of the left side and vice versa. This is caused by a too small number of samples acquired in, e.g. the frequency encoding direction, therefore the spectrums will overlap, resulting in a replication of the object in the x direction.
Aliasing in the frequency direction can be eliminated by twice as fast sampling of the signal or by applying frequency specific filters to the received signal.
A similar problem occurs in the phase encoding direction, where the phases of signal-bearing tissues outside of the FOV in the y-direction are a replication of the phases that are encoded within the FOV. Phase encoding gradients are scaled for the field of view only, therefore tissues outside the FOV do not get properly phase encoded relative to their actual position and 'wraps' into the opposite side of the image.
mri safety guidance
Image Guidance
Use a larger FOV, RFOV or 3D Volume, apply presaturation pulses to the undesired tissue, adjust the position of the FOV, or select a small coil which will only receive signal from objects inside or near the coil. The number of phase encoding steps must be increased in phase direction, unfortunately resulting in longer scan times.
When this is not possible it can be corrected by oversampling the data. Aliasing is eliminated by Oversampling in frequency direction. No Phase Wrap (Foldover Suppression) options typically correct the phase encoding by doubling the field of view, doubling the number of phase encodes (to keep resolution constant) and halving the number of averages (to keep scan time constant) then discarding the additional data and processing the image within the desired field of view (but this is more time consuming).
Tissue outside this doubled area can be folded nevertheless into the image as phase wrap. In this case combine more than 2 number of excitations / number of signal averages with foldover suppression.
See also Aliasing, Foldover Suppression, Oversampling, and Artifact Reduction - Aliasing.
spacer
 
• Related Searches:
    • Backfolding Artifact
    • Oversampling
    • Wrap Around Artifact
    • Foldover Artifact
    • Phase Wrapping Artifact
Searchterm 'Spectrum' was also found in the following services: 
spacer
Radiology  (7) Open this link in a new windowUltrasound  (9) Open this link in a new window
Apodization
 
The multiplication of acquired MR data by a function smoothly tapering off at higher spatial frequencies so as to reduce ringing artifacts near edges in the corresponding image or spectrum due to truncation and Gibbs phenomenon. It is a form of filtering.
spacer

• View the DATABASE results for 'Apodization' (2).Open this link in a new window

MRI Resources 
Corporations - Devices - General - MRI Reimbursement - Calculation - Used and Refurbished MRI Equipment
 
BandwidthForum -
related threads
 
(BW) Bandwidth is a measure of frequency range, the range between the highest and lowest frequency allowed in the signal. For analog signals, which can be mathematically viewed as a function of time, bandwidth is the width, measured in Hertz of a frequency range in which the signal's Fourier transform is nonzero.
•
The receiver (or acquisition) bandwidth (rBW) is the range of frequencies accepted by the receiver to sample the MR signal. The receiver bandwidth is changeable (see also acronyms for 'bandwidth' from different manufacturers) and has a direct relationship to the signal to noise ratio (SNR) (SNR = 1/squareroot (rBW). The bandwidth depends on the readout (or frequency encoding) gradient strength and the data sampling rate (or dwell time).
Bandwidth is defined by BW = Sampling Rate/Number of Samples.
A smaller bandwidth improves SNR, but can cause spatial distortions, also increases the chemical shift. A larger bandwidth reduces SNR (more noise from the outskirts of the spectrum), but allows faster imaging.
•
The transmit bandwidth refers to the RF excitation pulse required for slice selection in a pulse sequence. The slice thickness is proportional to the bandwidth of the RF pulse (and inversely proportional to the applied gradient strength). Lowering the pulse bandwidth can reduce the slice thickness.
mri safety guidance
Image Guidance
A higher bandwidth is used for the reduction of chemical shift artifacts (lower bandwidth - more chemical shift - longer dwell time - but better signal to noise ratio). Narrow receive bandwidths accentuate this water fat shift by assigning a smaller number of frequencies across the MRI image. This effect is much more significant on higher field strengths. At 1.5 T, fat and water precess 220 Hz apart, which results in a higher shift than in Low Field MRI.
Lower bandwidth (measured in Hz) = higher water fat shift (measured in pixel shift).

See also Aliasing, Aliasing Artifact, Frequency Encoding, and Chemical Shift Artifact.
spacer

• View the DATABASE results for 'Bandwidth' (19).Open this link in a new window

 
Further Reading:
  Basics:
Bandwidth
   by en.wikipedia.org    
  News & More:
Automated Quality Assurance for Magnetic Resonance Image with Extensions to Diffusion Tensor Imaging(.pdf)
   by scholar.lib.vt.edu    
A Real-Time Navigator Approach to Compensating for Motion Artifacts in Coronary Magnetic Resonance Angiography
   by www.cs.nyu.edu    
Searchterm 'Spectrum' was also found in the following services: 
spacer
News  (8)  Resources  (18)  Forum  (1)  
 
Baseline
 
A generally smooth background curve, only the data above the baseline/background will be included in the calculation. Either the integrals or peak heights of the resonance spectral lines in the spectrum are measured. For BOLD imaging it is a non-activated image, in contrast to an activated image.
spacer

• View the DATABASE results for 'Baseline' (9).Open this link in a new window

 
Further Reading:
  News & More:
Study is first to find that CPAP therapy restores brain tissue in adults with sleep apnea
Monday, 7 June 2010   by www.eurekalert.org    
Searchterm 'Spectrum' was also found in the following services: 
spacer
Radiology  (7) Open this link in a new windowUltrasound  (9) Open this link in a new window
Baseline Correction
 
Postprocessing of the spectrum to suppress baseline deviations from zero that may be superimposed on desired spectral lines. These deviations may be due either to various instrumental effects or to very broad spectral lines.
spacer

• View the DATABASE results for 'Baseline Correction' (3).Open this link in a new window

MRI Resources 
Hospitals - Service and Support - Implant and Prosthesis pool - Jobs - MRI Technician and Technologist Schools - NMR
 
previous      6 - 10 (of 34)     next
Result Pages : [1]  [2 3 4 5 6 7]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]