Inversion Recovery Sequence Timing Diagram (IR) The inversion recovery pulse sequence produces signals, which represent the longitudinal magnetization existing after the application of a 180° radio frequency pulse that rotates the magnetization Mz into the negative plane. After an inversion time (TI - time between the starting 180° pulse and the following 90° pulse), a further 90° RF pulse tilts some or all of the z-magnetization into the xy-plane, where the signal is usually rephased with a 180° pulse as in the spin echo sequence. During the initial time period, various tissues relax with their intrinsic T1 relaxation time.
In the pulse sequence timing diagram, the basic inversion recovery sequence is illustrated. The 180° inversion pulse is attached prior to the 90° excitation pulse of a spin echo acquisition. See also the Pulse Sequence Timing Diagram. There you will find a description of the components.
The inversion recovery sequence has the advantage, that it can provide very strong contrast between tissues having different T1 relaxation times or to suppress tissues like fluid or fat. But the disadvantage is, that the additional inversion radio frequency RF pulse makes this sequence less time efficient than the other pulse sequences.

Contrast values:
PD weighted: TE: 10-20 ms, TR: 2000 ms, TI: 1800 ms
T1 weighted: TE: 10-20 ms, TR: 2000 ms, TI: 400-800 ms
T2 weighted: TE: 70 ms, TR: 2000 ms, TI: 400-800 ms

See also Inversion Recovery, Short T1 Inversion Recovery, Fluid Attenuation Inversion Recovery, and Acronyms for 'Inversion Recovery Sequence' from different manufacturers.
 
Images, Movies, Sliders:
 Brain MRI Inversion Recovery  Open this link in a new window
    
 Knee MRI Sagittal STIR 002  Open this link in a new window
 Brain MRI Coronal FLAIR 001  Open this link in a new window
    
 
spacer
 
• Related Searches:
    • Flow Sensitive Alternating Inversion Recovery
    • Pulse, 180°
    • Double Inversion Recovery T1 Measurement
    • Saturation Pulses
    • Short T1 Inversion Recovery
 
Further Reading:
  Basics:
The equation for a repeated inversion recovery sequence
Contrast mechanisms in magnetic resonance imaging
2004   by www.iop.org    
  News & More:
FLAIR Vascular Hyperintensity: An Important MRI Marker in Patients with Transient Ischemic Attack
Thursday, 14 July 2022   by www.dovepress.com    
MRI Resources 
Societies - Safety Training - Corporations - Jobs - Services and Supplies - Manufacturers
 
Magnetic Field Mapping
 
The mapping of the magnetic field by measuring or imaging the spatial distribution of magnetic field strength, can be performed by scanning with a probe and handles a large range of field strengths, but is slow and tedious. Accurate field maps can be made by measuring the Larmor frequency as a function of position.
The field must be homogeneous enough to allow MR imaging to be performed, than the magnetic field can be mapped by different methods.
1. The adaptation of chemical shift imaging.
2. The faster one measures the change in signal phase in an image obtained with a gradient echo pulse sequence resulting from a change in echo time TE, which is proportional to the local field strength.
Also useful is a spin echo pulse sequence with data collection from two time locations of the readout gradient and the data acquisition interval, where each having a known shift of the acquisition center away from the spin echo.
spacer
 
Further Reading:
  News & More:
sPHENIX assembly update: magnet mapped, detectors prepared
Friday, 23 December 2022   by www.eurekalert.org    
MRI Resources 
PACS - - Functional MRI - Stimulator pool - MRCP - Pathology
 
Motion Compensation Pulse SequencesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Pulse sequences, designed to be insensitive to flow, e.g. at every even echo, a spin echo sequence is not flow sensitive. Velocity compensation is achieved by using gradients, which are either symmetrical around a 180° pulse and switched on twice as is the case for motion compensated spin echo pulse sequences, or two antisymmetrical gradient lobes without 180° pulse, which is the way to produce a velocity compensated gradient echo pulse sequence.
The signal of the second echo (and all other even echoes) is independent of the velocity of the object. Thus, velocity-based motion effects stemming from the entire voxel or from spins within a voxel (intravoxel incoherent motion) are suppressed with such pulse sequences.
If higher order motion is relevant, as it may be in turbulent jets across valves, acceleration and jerk effects can also be compensated for by the use of appropriate combinations of gradient- and radio frequency pulses.
With the increasingly stronger gradients, echo times in MR systems can be shortened to the point at which effects other than velocity effects hardly ever become relevant.
spacer

• View the DATABASE results for 'Motion Compensation Pulse Sequences' (2).Open this link in a new window

 
Further Reading:
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
Motion-compensation of Cardiac Perfusion MRI using a Statistical Texture Ensemble(.pdf)
June 2003   by www.imm.dtu.dk    
Searchterm 'Spin Echo' was also found in the following services: 
spacer
News  (1)  Forum  (10)  
 
Multiple Echo Imaging
 
Spin echo imaging using spin echoes acquired as a train. Typically a separate image is produced from each echo of the train.
spacer

• View the DATABASE results for 'Multiple Echo Imaging' (3).Open this link in a new window

MRI Resources 
Liver Imaging - Implant and Prosthesis - Education pool - Spectroscopy - Guidance - Stent
 
Number of EchoesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
The number of series of 180° RF rephasing pulses and their corresponding echoes for a fast or turbo spin echo pulse sequence is named echo train length (ETL) or turbo spin echo turbo factor (TSE TF).
spacer

• View the DATABASE results for 'Number of Echoes' (4).Open this link in a new window

MRI Resources 
MRI Technician and Technologist Schools - Pacemaker - Directories - Manufacturers - Equipment - Online Books
 
previous      46 - 50 (of 104)     next
Result Pages : [1 2 3 4 5 6]  [7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]