Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Spin Echo' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Spin Echo' found in 26 terms [] and 78 definitions []
previous     51 - 55 (of 104)     next
Result Pages : [1 2 3 4 5 6]  [7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
Searchterm 'Spin Echo' was also found in the following services: 
spacer
News  (1)  Forum  (10)  
 
RELAX 0.35T™InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.isoltech.co.kr/english/product/035t.htm From ISOL Technology
'RELAX is open type MRI system created by making up for the weakness of existing conventional MR systems and applying the strength and the application of the middle to high field MR without uncompromising the image quality.
RELAX offers you a premium mix of form, performance and functionality that are patient and user friendly beyond comparison. - New breed of MRI pursuing - patients comfort'
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Open
Head, C-spine, L-spine, TMJ, Knee, Shoulder, General purpose, Phased Array System: 4 digital receiver channels (Up to 12 channels)
SYNCHRONIZATION
ECG/peripheral: Optional/yes, respiratory gating
PULSE SEQUENCES
2D/3D Spin echo, 2D/3D Gradient echo, 2D/3D Fast spin echo, Inversion recovery, 2D/3D Fast gradient echo sequences, FLAIR/STIR, 2D/3D TOF
IMAGING MODES
3D volume imaging (MIP, MPR) MR Angiography package
30 cm
MAGNET TYPE
Permanent
40 cm diameter (patient)
MAGNET WEIGHT
13500 kg
H*W*D
170 x 210 x 130 cm
COOLING SYSTEM TYPE
Water-cooled coil and air-cooled amplifier
STRENGTH
25 - 30 mT/m
5-GAUSS FRINGE FIELD
lower than 2.4 m from the iso-center
Passive and active
spacer
MRI Resources 
Lung Imaging - PACS - MRI Technician and Technologist Schools - Open Directory Project - Coils - Health
 
Rapid Acquisition with Refocused EchoesInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
(RARE) If the image lines from multiple echoes are used for the same image, this results in the RARE pulse sequence. The sequence is similar to fast spin echo.

See Fast Spin Echo.
spacer

• View the DATABASE results for 'Rapid Acquisition with Refocused Echoes' (2).Open this link in a new window

 
Further Reading:
  Basics:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
MRI Resources 
Sequences - Mobile MRI - Blood Flow Imaging - Image Quality - Mass Spectrometry - MRA
 
Scan TimeForum -
related threads
 
(SCT) The total scan time is the time required to collect all data needed to generate the programmed images. The scan time is related to the used pulse sequence and dependent on the assemble of parameters like e.g., repetition time (TR), Matrix, number of signal averages (NSA), TSE- or EPI factor and flip angle.
For example, the total scan time for a standard spin echo or gradient echo sequence is number of repetitions x the scan time per repetition (means the product of repetition time (TR), number of phase encoding steps, and NSA).

See also Number of Excitations, Turbo Spin Echo Turbo Factor, Echo Planar Imaging Factor, Flip Angle and Image Acquisition Time.

See also acronyms for 'scan time parameters' from different manufacturers.
spacer

• View the DATABASE results for 'Scan Time' (47).Open this link in a new window


• View the NEWS results for 'Scan Time' (10).Open this link in a new window.
 
Further Reading:
  Basics:
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Sunday, 1 August 2004   by www.ajronline.org    
  News & More:
For MRI, time is of the essence A new generation of contrast agents could make for faster and more accurate imaging
Tuesday, 28 June 2011   by scienceline.org    
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
Searchterm 'Spin Echo' was also found in the following services: 
spacer
News  (1)  Forum  (10)  
 
Susceptibility ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
DESCRIPTION
Signal dropout, bright spots, spatial distortion
REASON
HELP
Remove the metal, do not take a gradient echo sequence, take a short echo time
Materials with magnetic susceptibility cause this artifact. There are in general three kinds of materials with magnetic susceptibility: ferromagnetic materials (iron, nickel etc.) with a strong influence and paramagnetic/diamagnetic (aluminium, platinum etc./gold, water, most organic compounds etc.) materials with a minimal/non influence on magnetic fields. In MRI, susceptibility artifacts are caused for example by medical devices in or near the magnetic field or by implants of the patient. These materials with magnetic susceptibility distort the linear magnetic field gradients, which results in bright areas (misregistered signals) and dark areas (no signal) nearby the magnetic material.
mri safety guidance
Image Guidance
Use a spin echo or a fast spin echo sequence, because gradient echo sequences are more sensitve to susceptibility artifacts. A high bandwidth (small water fat shift) and a short echo time help also to reduce this artifact.
In some cases it is even beneficial to use a gradient echo sequence, e.g. a cavernom contains some iron-rich haemosiderin, which also causes a signal void on gradient echo sequences and for this purpose increases the diagnostic image quality.
spacer

• View the DATABASE results for 'Susceptibility Artifact' (8).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Susceptibility Artifacts
   by www.mritutor.org    
  News & More:
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
Ultrashort echo time (UTE) MRI of the spine in thalassaemia
February 2004   by bjr.birjournals.org    
MRI Resources 
Pregnancy - Case Studies - Education - MRI Technician and Technologist Career - Education pool - Bioinformatics
 
Aurora® 1.5T Dedicated Breast MRI SystemInfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.
 
www.auroramri.com/mri/product/ From Aurora Imaging Technology, Inc.;
The Aurora® 1.5T Dedicated Breast MRI System with Bilateral SpiralRODEO™ is the first and only FDA approved MRI device designed specifically for breast imaging. The Aurora System, which is already in clinical use at a growing number of leading breast care centers in the US, Europe, got in December 2006 also the approval from the State Food and Drug Administration of the People's Republic of China (SFDA).
'Some of the proprietary and distinguishing features of the Aurora System include: 1) an ellipsoid magnetic shim that provides coverage of both breasts, the chest wall and bilateral axillary lymph nodes; 2) a precision gradient coil with the high linearity required for high resolution spiral reconstruction;; 3) a patient-handling table that provides patient comfort and procedural utility; 4) a fully integrated Interventional System for MRI guided biopsy and localization; and 5) the user-friendly AuroraCAD™ computer-aided image display system designed to improve the accuracy and efficiency of diagnostic interpretations.'
Device Information and Specification
CLINICAL APPLICATION
CONFIGURATION
Short bore compact
IMAGING MODES
Bi-lateral RODEO fat suppression, high resolution Rotating Delivery of Excitation Offresonance Spiral, integrated targeting SW compatible with major MR guided intervention equipment
TR
10 ms for gradient echo and less than 2,500 ms for T2 weighted spin echo
TE
From 5 ms for RODEO Plus to over 80 ms, 120 ms for T2 sequences
SINGLE/MULTI SLICE
Around 0.02 sec for a 256x256 image, 12.4 sec for a 512 x 512 x 32 multislice set
FOV
20 - 36 cm, max. elliptical 36 x 44 cm
MEASURING MATRIX
512 x 512
BORE DIAMETER
or W x H
64 cm diameter (gantry)
MAGNET WEIGHT
8,500 lbs
H*W*D
240 x 188 x 163 cm
POWER REQUIREMENTS
150A/120V-208Y/3 Phase//60 Hz/5 Wire
COOLING SYSTEM
Helium for magnet, distilled/de-ionized water for coil;
spacer

• View the DATABASE results for 'Aurora® 1.5T Dedicated Breast MRI System' (2).Open this link in a new window


• View the NEWS results for 'Aurora® 1.5T Dedicated Breast MRI System' (3).Open this link in a new window.
 
Further Reading:
  News & More:
Aurora Imaging Technology Announces Approval in China
   by salesandmarketingnetwork.com    
Aurora Imaging Technology, Inc. Release: Results of a Multi-Center Trial Demonstrates Superior Diagnostic Performance of the Aurora® 1.5T Dedicated Breast MRI System Over Whole-Body Breast MRI
Monday, 1 October 2012   by www.biospace.com    
MRI Resources 
Pregnancy - Sequences - Chemistry - Education - Implant and Prosthesis - Fluorescence
 
previous      51 - 55 (of 104)     next
Result Pages : [1 2 3 4 5 6]  [7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]