Opposed Phase ImageInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
An image in which the signal from two spectral components (such as fat and water) is 180° out of phase and leads to destructive interference in a voxel.
Since fat precesses slower than water, based on their chemical shift, their signals will decay and precess in the transverse plane at different frequencies. When the phase of the TE becomes opposed (180°), their combined signal intensities subtract with each other in the same voxel, producing a signal void or dark band at the fat/water interface of the tissues being examined.
Opposed phase gradient echo imaging for the abdomen is a lipid-type tissue sensitive sequence particularly for the liver and adrenal glands, which puts a signal intensity around abnormal water-based tissues or lesions that are fatty. Due to the increased sensitivity of opposed phase, the tissue visualization increases the lesion-to-liver contrast and exhibits more signal intensity loss in tissues containing small amounts of lipids compared to a spin echo T1 with fat suppression. Using an opposed phase gradient echo also provides the ability to differentiate various pathologies in the brain, including lipids, methaemoglobin, protein, calcifications and melanin.

See also Out of Phase, and Dixon.
 
Images, Movies, Sliders:
 MRI Liver Out Of Phase  Open this link in a new window
    
 
spacer
 
• Related Searches:
    • Black Boundary Artifact
    • Chemical Shift Imaging
    • Out of Phase
    • T2*
    • Field Echo with Echo Time set for Water and Fat Signals in Opposition
 
Further Reading:
  News & More:
Adrenal Myelolipoma
Tuesday, 19 June 2001   by www.emedicine.com    
Iron overload: accuracy of in-phase and out-of-phase MRI as a quick method to evaluate liver iron load in haematological malignancies and chronic liver disease
Friday, 1 June 2012   by www.ncbi.nlm.nih.gov    
Searchterm 'Suppression' was also found in the following services: 
spacer
Radiology  (1) Open this link in a new windowUltrasound  (4) Open this link in a new window
Phase Contrast AngiographyMRI Resource Directory:
 - MRA -
 
(PCA) With this method images of the blood flow-velocity (or any other movement of tissue) are produced. The MRI signal contains both amplitude and phase information. The phase information can be used with subtraction of images with and without a velocity encoding gradient. The signal will be directly proportional to the velocity because of the relation between blood flow-velocity and signal intensity.
This is the strength of PCA, complete suppression of stationary tissue (no velocity - no signal), the direct velocity of flow is being imaged, while in TOF (Inflow) angiography, tissue with short T1 (fat or methaemoglobin) might be visualized.
The strength of the gradient determines the sensitivity to flow. It is set by setting the aliasing or encoding velocity (VENC). Unfortunately, phase sensitization can only be acquired along one axis at a time. Therefore, phase contrast angiographic techniques tend to be 4 times slower than TOF techniques with the same matrix.

See also Phase Contrast Sequence, Magnetic Resonance Angiography, Contrast Enhanced Magnetic Resonance Angiography, Flow Effects and Flow Quantification.
 
Images, Movies, Sliders:
 PCA-MRA 3D Brain Venography Colored MIP  Open this link in a new window
    

 
spacer

• View the DATABASE results for 'Phase Contrast Angiography' (8).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance angiography: current status and future directions
Wednesday, 9 March 2011   by www.jcmr-online.com    
  News & More:
MR–ANGIOGRAPHY(.pdf)
MRI Resources 
Process Analysis - Spectroscopy pool - Artifacts - MR Myelography - Colonography - Functional MRI
 
Rotating Delivery of Excitation OffresonanceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(RODEO) The software package Spiral RODEO™ MRI (including a patented fat and ductal tissue suppression technology), developed from Aurora Imaging Technology Inc, improves the spatial resolution and sensitivity for cancer detection in breast MRI examinations.

See also Aurora® 15T Dedicated Breast MRI System.
spacer

• View the DATABASE results for 'Rotating Delivery of Excitation Offresonance' (2).Open this link in a new window

Searchterm 'Suppression' was also found in the following services: 
spacer
News  (2)  Resources  (2)  Forum  (7)  
 
Sample Imperfection (Artifact)InfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Sample imperfection
DESCRIPTION
Shifts of the signal in the phase encoding direction
REASON
Distorting the k-space trajectory, reduced bandwidth
HELP
Fat suppression, more excitations
Artifacts either by distorting the k-space trajectory (i.e. due to imperfect shimming) or as a consequence of the reduced bandwidth in the phase encode direction, commonly with EPI sequences.
While a standard spin warp-based sequence has an infinitely large bandwidth in the phase encode direction (about 1 or 2 kH), the bandwidth in EPI is related to the time between the gradient echoes (about a millisecond).
Hence even small frequency offsets can result in significant shifts of the signal in the phase encoding direction. Segmentation can introduce ghosting if there are significant difference in the amplitude and phase of the signal. This can be a particular problem when trying to acquire the segments in rapid succession.
mri safety guidance
Image Guidance
Suitable choices of excitation schemes and/or subsequent correction can help to reduce this artifact. The signal from fat can easily be offset by a large fraction of the FOV, and must be suppressed. The effect of frequency offsets can be reduced by collecting data with more than one excitation, which effectively increases the bandwidth in the phase encoding direction.
spacer
Searchterm 'Suppression' was also found in the following services: 
spacer
Radiology  (1) Open this link in a new windowUltrasound  (4) Open this link in a new window
Saturation Pulses
 
Sequence of RF (and gradient) pulses designed to produce saturation, typically in a selected region or set of regions, most often by the use of selective excitation followed by a spoiler pulse. Similar to some spectral suppression techniques. Can be used to reduce signal from flowing blood by saturating regions upstream from region being imaged.
spacer

• View the DATABASE results for 'Saturation Pulses' (7).Open this link in a new window

MRI Resources 
Quality Advice - Pathology - Lung Imaging - Non-English - Calculation - Sequences
 
previous      41 - 45 (of 49)     next
Result Pages : [1 2]  [3 4 5 6 7 8 9 10]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 3 December 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]