Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'Susceptibility Artifact' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'Susceptibility Artifact' found in 1 term [] and 15 definitions [], (+ 5 Boolean[] results
1 - 5 (of 21)     next
Result Pages : [1]  [2 3 4]  [5]
Searchterm 'Susceptibility Artifact' was also found in the following services: 
spacer
News  (1)  Resources  (1)  
 
Susceptibility ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
DESCRIPTION
Signal dropout, bright spots, spatial distortion
REASON
HELP
Remove the metal, do not take a gradient echo sequence, take a short echo time
Materials with magnetic susceptibility cause this artifact. There are in general three kinds of materials with magnetic susceptibility: ferromagnetic materials (iron, nickel etc.) with a strong influence and paramagnetic/diamagnetic (aluminium, platinum etc./gold, water, most organic compounds etc.) materials with a minimal/non influence on magnetic fields. In MRI, susceptibility artifacts are caused for example by medical devices in or near the magnetic field or by implants of the patient. These materials with magnetic susceptibility distort the linear magnetic field gradients, which results in bright areas (misregistered signals) and dark areas (no signal) nearby the magnetic material.
mri safety guidance
Image Guidance
Use a spin echo or a fast spin echo sequence, because gradient echo sequences are more sensitve to susceptibility artifacts. A high bandwidth (small water fat shift) and a short echo time help also to reduce this artifact.
In some cases it is even beneficial to use a gradient echo sequence, e.g. a cavernom contains some iron-rich haemosiderin, which also causes a signal void on gradient echo sequences and for this purpose increases the diagnostic image quality.
spacer
 
• Share the entry 'Susceptibility Artifact':  Facebook  Twitter  LinkedIn  
 
• Related Searches:
    • PRinciples of Echo Shifting using a Train of Observations
    • Magnetization
    • Metal Artifact
    • Homogeneity
    • Echo Planar Imaging
 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
Susceptibility Artifacts
   by www.mritutor.org    
  News & More:
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
Ultrashort echo time (UTE) MRI of the spine in thalassaemia
February 2004   by bjr.birjournals.org    
MRI Resources 
Service and Support - MRI Physics - Cochlear Implant - Patient Information - Pregnancy - Case Studies
 
Metal ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Metal, susceptibility
DESCRIPTION
Signal dropout, bright spots
REASON
HELP
Remove the metal
Ferromagnetic metal will cause a magnetic field inhomogeneity, which in turn causes a local signal void, often accompanied by an area of high signal intensity, as well as a distortion of the image. They create their own magnetic field and dramatically alter precession frequencies of protons in the adjacent tissues. Tissues adjacent to ferromagnetic components become influenced by the induced magnetic field of the metal hardware rather than the parent field and, therefore, either fail to precess or do so at a different frequency and hence do not generate useful signal. Two components contribute to susceptibility artifact, induced magnetism in the ferromagnetic component itself and induced magnetism in protons adjacent to the component.
Artifacts from metal may have varied appearances on MRI scans due to different type of metal or configuration of the piece of metal. The biocompatibility of metallic alloys, stainless steel, cobalt chrome and titanium alloy is based on the presence of a constituent element within the alloy that has the ability to form an adherent oxide coating that is stable, chemically inert and hence biocompatible. In relation to imaging titanium alloys are less ferromagnetic than both cobalt and stainless steel, induce less susceptibility artifact and result in less marked image degradation.
mri safety guidance
Image Guidance
Remove the metal when possible or take a not so sensitive sequence (a SE or another sequence with a rephasing 180° pulse).

See also Susceptibility Artifact.
spacer

• View the DATABASE results for 'Metal Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Metal-Induced Artifacts in MRI
   by www.ajronline.org    
Metal Artefact Reduction
Thursday, 9 June 2011   by www.revisemri.com    
  News & More:
Multiacquisition with variable resonance image combination T2 (MAVRIC SL T2) for postoperative cervical spine with artificial disc replacement
Friday, 11 November 2022   by www.nature.com    
Modeling of Active Shimming of Metallic Needles for Interventional MRI
Monday, 29 June 2020   by pubmed.ncbi.nlm.nih.gov    
MRI Resources 
Education - Abdominal Imaging - NMR - Devices - Directories - Used and Refurbished MRI Equipment
 
Negative Oral Contrast AgentsForum -
related threadsInfoSheet: - Contrast Agents - 
Intro, Overview, 
Characteristics, 
Types of, 
etc.MRI Resource Directory:
 - Contrast Agents -
 
Categories of negative oral contrast agents:
Negative oral contrast media are usually based on superparamagnetic particles and act by inducing local field inhomogeneities, which results in shortening of both T1 and T2 relaxation times. Superparamagnetic contrast agents have predominant T2 weighted effects. Biphasic contrast media are agents that have different signal intensities on different sequences, depending on the concentration at which they are used.
Suitable materials for oral contrast agents should have little or no absorption by the stomach or intestines, complete excretion, no motion or susceptibility artifacts, affordability, and uniform marking of the gastrointestinal tract. Benefits of negative oral contrast agents are the reduction of ghosting artifacts caused by the lack of signal. Superparamagnetic iron oxides produce also in low concentrations a noticeable signal loss; but can generate susceptibility artifacts especially in gradient echo sequences. Perfluorochemicals do not dilute in the bowel because they are not miscible with water.
High cost, poor availability, and limited evaluations of side effects are possible disadvantages.
Negative oral contrast agents are used e.g., in MRCP, where the ingestion of 600-900 ml of SPIO cancels out the signal intensity of the lumen (in addition after the injection of a gadolinium-based contrast medium, the enhancement of the inflammatory tissues is clearer seen), and in MR abdominal imaging of Crohn's disease in combination with mannitol.
mri safety guidance
Contrast Guidance
Blueberry or pineapple juices are useable for examinations of the pancreas (MRCP, upper abdominal imaging) as cheep contrast agents, because of the content of magnetic substances (e.g. manganese).

See also Ferristene, Ferumoxsil, Oral Magnetic Particles, Gastrointestinal Imaging.
spacer

• View the DATABASE results for 'Negative Oral Contrast Agents' (7).Open this link in a new window

 
Further Reading:
  Basics:
Components of Oral Contrast Agent for Magnetic Resonance Imaging(.pdf)
   by www.ffcr.or.jp    
Searchterm 'Susceptibility Artifact' was also found in the following services: 
spacer
News  (1)  Resources  (1)  
 
Burst Pulse SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Burst pulse sequences are fast imaging sequences capable of image acquisition in less than 100 ms. Basically a train of low flip angle pulses generates a long train of echoes. The complete sequence is performed with the application of a constant read gradient. Phase encoding may be implemented using short phase encoding gradients between echoes.
The advantage of this sequence type is that it is less demanding on gradient speed than other fast techniques (e.g. echo planar imaging EPI) and it produces images, which are substantially free of susceptibility artifacts.
The disadvantage is that the technique is less sensitive than competing methods.
spacer
MRI Resources 
Equipment - Quality Advice - Lung Imaging - Pathology - Jobs pool - Claustrophobia
 
Cardiac StentForum -
related threadsMRI Resource Directory:
 - Safety -
 
The MRI safety of cardiac stents is dependent of the material, the examined part of the body and the used field strength. A susceptibility artifact is expected also in low magnetic fields, but less.
mri safety guidance
MRI Safety Guidance
Most of the used materials are non-magnetic, for this case there is no risk for movement caused through the magnetic field. If the cardiac stent is outside the region of the radio frequency pulse, also the risk of e.g. heating is low.
spacer

• View the DATABASE results for 'Cardiac Stent' (4).Open this link in a new window

 
Further Reading:
  Basics:
Guidelines for the Management of Patients with Coronary Artery Stents Referred for MRI Procedures
Tuesday, 1 January 2019   by www.ismrm.org    
MRI Resources 
MRI Reimbursement - Journals - Sequences - Societies - Safety pool - Coils
 
     1 - 5 (of 21)     next
Result Pages : [1]  [2 3 4]  [5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Personalized protocols (age, gender, body habitus, etc.) lead to :
more automated planning 
improved patient comfort 
better diagnostics 
optimized image quality 
nothing 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]