Undersampling
 
Undersampling is the decrease in data to increase image acquisition speed (shorter scan times without loss of quality - increased productivity - reduced cost of equipment). There are different strategies to decrease data without losing quality (e.g. reduction of the FOV in one or more spatial directions - RFOV).
Reduction in data normally is associated with an increase in aliasing (degradation of the SNR through backfolding of the entire noise spectrum), or with other artifact caused by missing data, which results in fine lines.
spacer
 
• Share the entry 'Undersampling':  Facebook  Twitter  LinkedIn  
MRI Resources 
Directories - Knee MRI - Crystallography - IR - Online Books - MRA
 
Aliasing ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Aliasing, backfolding, foldover, phase wrapping, wrap around
DESCRIPTION
Image wrap around
Aliasing is an artifact that occurs in MR images when the scanned body part is larger than field of view (FOV). As a consequence of the acquired k-space frequencies not being sampled densely enough, whereby portions of the object outside of the desired FOV get mapped to an incorrect location inside the FOV. The cyclical property of the Fourier transform fills the missing data of the right side with data from behind the FOV of the left side and vice versa. This is caused by a too small number of samples acquired in, e.g. the frequency encoding direction, therefore the spectrums will overlap, resulting in a replication of the object in the x direction.
Aliasing in the frequency direction can be eliminated by twice as fast sampling of the signal or by applying frequency specific filters to the received signal.
A similar problem occurs in the phase encoding direction, where the phases of signal-bearing tissues outside of the FOV in the y-direction are a replication of the phases that are encoded within the FOV. Phase encoding gradients are scaled for the field of view only, therefore tissues outside the FOV do not get properly phase encoded relative to their actual position and 'wraps' into the opposite side of the image.
mri safety guidance
Image Guidance
Use a larger FOV, RFOV or 3D Volume, apply presaturation pulses to the undesired tissue, adjust the position of the FOV, or select a small coil which will only receive signal from objects inside or near the coil. The number of phase encoding steps must be increased in phase direction, unfortunately resulting in longer scan times.
When this is not possible it can be corrected by oversampling the data. Aliasing is eliminated by Oversampling in frequency direction. No Phase Wrap (Foldover Suppression) options typically correct the phase encoding by doubling the field of view, doubling the number of phase encodes (to keep resolution constant) and halving the number of averages (to keep scan time constant) then discarding the additional data and processing the image within the desired field of view (but this is more time consuming).
Tissue outside this doubled area can be folded nevertheless into the image as phase wrap. In this case combine more than 2 number of excitations / number of signal averages with foldover suppression.
See also Aliasing, Foldover Suppression, Oversampling, and Artifact Reduction - Aliasing.
spacer

• View the DATABASE results for 'Aliasing Artifact' (11).Open this link in a new window

MRI Resources 
Anatomy - Bioinformatics - Mobile MRI Rental - Health - Most Wanted - Safety pool
 
Backfolding ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
NAME
Backfolding, foldover, phase wrapping, wrap around
DESCRIPTION
Image wrap around
Backfolding always occurs due to wrong phase encoding caused by objects outside the planned FOV. Phase encoding gradients are scaled for the field of view only. Tissues outside the FOV do not get properly phase encoded relative to their actual position and 'wraps' into the opposite side of the image. The Backfolding artifact projects image contents which fall outside the imaging FOV back into the image; the back folded information thus reappearing on the other side of the image. In fact, information along the phase encoding direction can be viewed as projected onto a cylindrical screen with a circumference corresponding to the linear field of view dimension in the phase encoding direction.

See also Aliasing Artifact.
spacer

• View the DATABASE results for 'Backfolding Artifact' (2).Open this link in a new window

 
Further Reading:
  Basics:
Aliasing or wrap around artifacts
Thursday, 31 March 2011   by de.slideshare.net    
Searchterm 'Undersampling' was also found in the following service: 
spacer
News  (1)  
 
Generalized Autocalibrating Partially Parallel AcquisitionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(GRAPPA) GRAPPA is a parallel imaging technique to speed up MRI pulse sequences. The Fourier plane of the image is reconstructed from the frequency signals of each coil (reconstruction in the frequency domain).
Parallel imaging techniques like GRAPPA, auto-SMASH and VD-AUTO-SMASH are second and third generation algorithms using k-space undersampling. A model from a part of the center of k-space is acquired, to find the coefficients of the signals from each coil element, and to reconstruct the missing intermediary lines. The acquisition of these additional lines is a form of self-calibration, which lengthens the overall short scan time. The acquisition of these k-space lines provides mapping of the whole field as well as data for the image contrast.
Algorithms of the GRAPPA type work better than the SENSE type in heterogeneous body parts like thoracic or abdominal imaging, or in pulse sequences like echo planar imaging. This is caused by differences between the sensitivity map and the pulse sequence (e.g. artifacts) or an unreliable sensitivity map.
spacer

• View the DATABASE results for 'Generalized Autocalibrating Partially Parallel Acquisition' (2).Open this link in a new window

MRI Resources 
Resources - Safety Products - Case Studies - Spectroscopy pool - NMR - Homepages
 
Gibbs ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this MRI artifact.
Artifact Information
NAME
Gibbs, Gibbs ringing, truncation
DESCRIPTION
Edge ringing, syrinx-like stripe
REASON
Sharp changes in intensity (incomplete digitization of the echo)
HELP
More samples
The Gibbs or ringing artifact appears as a series of lines in the MR image parallel to abrupt and intense changes in the object at this location. This artifact does not occur visibly on smooth objects. This artifact is caused by the Gibbs phenomenon, an overshoot or ringing of Fourier series occurring at discontinuities.
In the spinal cord, a small syrinx can be simulated by the Gibbs phenomenon. Gibbs artifacts are also seen in other regions, for example the brain//skull interface.
Fine lines visible in an image may be due to undersampling of the high spatial frequencies, respectively incomplete digitization of the echo.
With more encoding steps the Gibbs artifacts is less intense and narrower. Therefore, e.g. the artifact is more intense in the 256 point dimension of a 256x512 acquisition matrix.
mri safety guidance
Image Guidance
This problem can only be resolved by smoothing filters (LanczosSigmaFactor, 2-D Exponential Filtering, Gegenbauer Reconstruction etc.) or with a higher acquisition matrix and/or a smaller FOV, to smooth the object.

See also Gibbs Phenomenon and Apodization.
spacer

• View the DATABASE results for 'Gibbs Artifact' (4).Open this link in a new window

 
Further Reading:
  Basics:
Lanczos sigma Factor
   by mathworld.wolfram.com    
  News & More:
Troubleshooting the ACR MRI Accreditation Phantom Tests
   by www.aapm.org    
MRI Resources 
DICOM - Abdominal Imaging - Fluorescence - Pathology - Mobile MRI Rental - Directories
 
     1 - 5 (of 6)     next
Result Pages : [1]  [2]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]