| Info Sheets |
| | | | | | | | | | | | | | | | | | | | | | | | |
| Out- side |
| | | | |
|
| | | | | | | Searchterm 'View' was also found in the following services: | | | | |
| | |
| |
|
| | | | | Further Reading: | | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'View' was also found in the following services: | | | | |
| | |
| |
|
Quick Overview
REASON
Motion, heartbeat, respiration
HELP
Triggering, breath hold, pharmaceuticals to reduce bowel motion
Ghosting artifacts are in the most cases caused by movements (e.g., respiratory motion, bowel motion, arterial pulsations, swallowing, and heartbeat) and appear in the phase encoding direction.
Image Guidance
| | | | • View the DATABASE results for 'Ghosting Artifact' (5).
| | | | Further Reading: | Basics:
|
|
| |
| | | | | |
| |
|
Quick Over viewPlease note that there are different common names for this MRI artifact.
DESCRIPTION
Edge ringing, syrinx-like stripe
The Gibbs or ringing artifact appears as a series of lines in the MR image parallel to abrupt and intense changes in the object at this location. This artifact does not occur visibly on smooth objects. This artifact is caused by the Gibbs phenomenon, an overshoot or ringing of Fourier series occurring at discontinuities.
In the spinal cord, a small syrinx can be simulated by the Gibbs phenomenon. Gibbs artifacts are also seen in other regions, for example the brain//skull interface.
Fine lines visible in an image may be due to undersampling of the high spatial frequencies, respectively incomplete digitization of the echo.
With more encoding steps the Gibbs artifacts is less intense and narrower. Therefore, e.g. the artifact is more intense in the 256 point dimension of a 256x512 acquisition matrix.
Image Guidance
| | | | • View the DATABASE results for 'Gibbs Artifact' (4).
| | | | Further Reading: | Basics:
|
|
News & More:
| |
| |
| | | Searchterm 'View' was also found in the following services: | | | | |
| | |
| |
|
(HS) A method in which approximately one half of the acquisition matrix in the phase encoding direction is acquired. Half scan is possible because of symmetry in acquired data. Since negative values of phase encoded measurements are identical to corresponding positive values, only a little over half (more than 62.5%) of a scan actually needs to be acquired to replicate an entire scan.
This results in a reduction in scan time at the expense of signal to noise ratio. The time reduction can be nearly a factor of two, but full resolution is maintained.
Half scan can be used when scan times are long, the signal to noise ratio is not critical and where full spatial resolution is required. Half scan is particularly appropriate for scans with a large field of view and relatively thick slices; and, in 3D scans with many slices.
In some fast scanning techniques the use of Half scan enables a shorter TE thus improving contrast. For this reason, the Half scan parameter is located in the contrast menu.
More information about scan time reduction; see also partial fourier technique. | | | | • View the DATABASE results for 'Half Scan' (4).
| | | | |
| | | Searchterm 'View' was also found in the following services: | | | | |
| | | | | | |
| | | |
|
| |
| Look Ups |
| |