Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'View' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'View' found in 5 terms [] and 99 definitions []
previous     81 - 85 (of 104)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
Searchterm 'View' was also found in the following services: 
spacer
News  (56)  Resources  (52)  Forum  (50)  
 
Phased Array CoilInfoSheet: - Coils - 
Intro, 
Overview, 
etc.MRI Resource Directory:
 - Coils -
 
The phased array coils operate typically as receive only coils. In that case, the in the MRI device implemented body coil act as the transmitter and sends the radio frequency energy to generate the excitation pulses. State-of-the-art array coil systems include the use of 4 (up to 32) coils with separate receivers. This method is often referred to as a phased array system, although the signals are not added such that the signal phase information is included. The use of phased array coils allows the decreasing of the number of signal averages, which shortens the scan time by high SNR and resolution.
High-sensitivity RF surface coils and digital processing algorithms have been developed that speed up image acquisition and reconstruction during the MRI scan.
Fast parallel imaging techniques, for example sensitivity encoding (SENSE), 'Partially Parallel Imaging with Localized Sensitivity' (PILS), Simultaneous Acquisition of Spatial Harmonics (SMASH) or Array Spatial Sensitivity Encoding Technique (ASSET) use phased array multichannel coils to further improve spatial and temporal resolution. The sensitivity profile of a phased array coil element is measured by a separate low resolution 3D acquisition over the entire field of view in the case of a SENSE acquisition. For an mSENSE measurement, a self-calibration acquires some of the missing lines in the center of the k-space.
Also called linear array coil or synergy surface coil.

See also the related poll result: '3rd party coils are better than the original manufacturer coils'
spacer
 
• Related Searches:
    • Signal to Noise Ratio
    • Radio Frequency Coil
    • Transmitter
    • Sensitivity Encoding
    • Temporal Resolution
 
Further Reading:
  Basics:
Coil Arrays for Parallel MRI: Introduction and Overview.
   by www.mr.ethz.ch    
Searchterm 'View' was also found in the following services: 
spacer
Radiology  (40) Open this link in a new windowUltrasound  (50) Open this link in a new window
Picture Archiving and Communication SystemMRI Resource Directory:
 - PACS -
 
(PACS) A system used to communicate and archive medical imaging data, mostly images and associated textural data generated in a radiology department, and disseminated throughout the hospital. A PACS is usually based on the DICOM (Digital Imaging and Communications in Medicine) standard.
The main components in the PACS are:
•
acquisition devices where the images are acquired,
•
short and longer term archives for storage of digital and textural data,
•
a database and database management,
•
diagnostic and review workstations,
•
software to run the system,
•
a communication network linking the system components,
•
interfaces with other networks (hospital and radiological information systems).

Acquisition devices, which acquire their data in direct digital format, like a MRI system, are most easily integrated into a PACS.
Short term archives need to have rapid access, such as provided by a RAID (redundant array of independent disks), whereas long term archives need not have such rapid access and can be consigned, e.g. to optical disks or a magnetic.
High speed networks are necessary for rapid transmission of imaging data from the short term archive to the diagnostic workstations. Optical fiber, ATM (asynchronous transfer mode), fast or switched Ethernet, are examples of high speed transmission networks, whereas demographic textural data may be transmitted along conventional Ethernet.
Sophisticated software is a major element in any hospital-wide PACS. The software concepts include: preloading or prefetching of historical images pertinent to current examinations, worklists and folders to subdivide the vast mass of data acquired in a PACS in a form, which is easy and practical to access, default display protocols whereby images are automatically displayed on workstation monitors in a prearranged clinically logical order and format, and protocols radiologists can rapidly report worklists of undictated examinations, using a minimum of computer manipulation.
spacer

• View the DATABASE results for 'Picture Archiving and Communication System' (5).Open this link in a new window


• View the NEWS results for 'Picture Archiving and Communication System' (1).Open this link in a new window.
 
Further Reading:
  Basics:
Healthcare IT Yellow Pages PACS / Image Management Directory
   by www.health-infosys-dir.com    
MRI Resources 
Pathology - Breast Implant - MRI Reimbursement - Collections - Software - MRI Technician and Technologist Schools
 
Quadrature Detector
 
The quadrature detector is a part of the receiver that converts the high-frequency MRI signal to a lower frequency. This phase sensitive detector or demodulator detects the components of the signal in phase with a reference signal and 90° out of phase with the reference signal. The modulated signal contains i.e. the frequency range across the field of view encoded by the frequency encoding gradient. This may be performed by either analog or digital means.
spacer

• View the DATABASE results for 'Quadrature Detector' (7).Open this link in a new window

Searchterm 'View' was also found in the following services: 
spacer
News  (56)  Resources  (52)  Forum  (50)  
 
Quadrupole ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Quadrupole artifact
DESCRIPTION
Signal loss, intensity variations
REASON
B1 disturbance
HELP
Fat suppression (SPIR or FatSat) is very critical to the magnetic field homogeneity. Eddy currents in the patient results in B1 disturbance from left to right and from anterior to posterior. The artifact is seen as signal intensity variations with SPIR, like a signal intensity loss diagonal in the image. The short T1 inversion recovery (STIR) sequence is due to another type of fat suppression insensitive to this artifact.
mri safety guidance
Image Guidance
spacer
Searchterm 'View' was also found in the following services: 
spacer
Radiology  (40) Open this link in a new windowUltrasound  (50) Open this link in a new window
Radio Frequency Noise ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Please note that there are different common names for this artifact.
Artifact Information
DESCRIPTION
Static on the image
REASON
Electromagnetic emissions
HELP
Shielding, eliminate the factor of disturbance
RF noise, which often appears as static on the image, can be caused by a medical device located anywhere in the MR procedure room. RF noise is a result of excessive electromagnetic emissions from the device that interference with the proper operation of the MR scanner. The interference is attenuated and aliased in the frequency direction.
mri safety guidance
Image Guidance
In the first case try to eliminate the factor of disturbance, if the problem persists screen the magnet- and room shielding.

See also Signal to Noise Ratio.
spacer

• View the DATABASE results for 'Radio Frequency Noise Artifact' (3).Open this link in a new window

 
Further Reading:
  Basics:
MRI Artifact Gallery
   by chickscope.beckman.uiuc.edu    
MRI Resources 
Manufacturers - Societies - Online Books - Liver Imaging - Raman Spectroscopy - Education
 
previous      81 - 85 (of 104)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 ... ]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



Next big thing in MRI will be :
AI 
remote operator 
personalized protocols 
helium-free 
molecular MRI 
portable MRI 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 21 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]