Magnetic Resonance - Technology Information Portal Welcome to MRI Technology
Info
  Sheets

Out-
      side
 



 
 'refocusing pulse' 
SEARCH FOR    
 
  2 3 5 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Result : Searchterm 'refocusing pulse' found in 0 term [] and 6 definitions [], (+ 15 Boolean[] results
previous     16 - 20 (of 21)     next
Result Pages : [1 2]  [3 4 5]
Searchterm 'refocusing pulse' was also found in the following service: 
spacer
Forum  (1)  
 
Spoiler Gradient PulseInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
Magnetic field gradient pulse applied to effectively remove transverse magnetization by producing a rapid variation of its phase along the direction of the gradient. This is done after the echo so that transverse magnetization is destroyed prior to the next excitation pulse, to spoil any remaining xy-magnetization or to refocus the xy-magnetization.
For example, when used to remove the unwanted signal resulting from an imperfect 180° refocusing RF pulse, a corresponding compensating gradient pulse may be applied prior to the refocusing RF pulse in order to avoid spoiling the desired transverse magnetization resulting from the initial excitation. Also called homospoil pulse.
spacer
 
• Related Searches:
    • Homospoil
    • Saturation
    • Refocusing
    • Coherent Gradient Echo
    • Radio Frequency Spoiling
 
Further Reading:
  News & More:
Clinical evaluation of a speed optimized T2 weighted fast spin echo sequence at 3.0 T using variable flip angle refocusing, half-Fourier acquisition and parallel imaging
Wednesday, 25 October 2006
MRI Resources 
Corporations - Equipment - Homepages - Claustrophobia - Stent - MR Guided Interventions
 
Dephasing Gradient
 
Magnetic field gradient pulse used to create spatial variation of phase of transverse magnetization. For example, it may be applied prior to signal detection in the presence of a magnetic field gradient with opposite polarity (or of the same polarity if separated by a refocusing RF pulse) so that the resulting gradient echo signal will represent a more complete sampling of the Fourier transformation of the desired image.

See also Spoiler Gradient Pulse.
spacer

• View the DATABASE results for 'Dephasing Gradient' (6).Open this link in a new window

 
Further Reading:
  Basics:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
MRI Resources 
Health - Movies - Service and Support - Pacemaker - Developers - Process Analysis
 
Hahn Echo
 
Production of spin echo by repeated RF pulses. First observed using equal (90°) RF pulses, now commonly used to describe refocusing of transverse magnetization by a 180° RF pulse. By choosing long echo delay times, the spins in a Hahn echo first dephase for a long time, then rephase, which makes the Hahn pulse sequence more susceptible to diffusion effects.
spacer

• View the DATABASE results for 'Hahn Echo' (2).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic resonance imaging
   by www.scholarpedia.org    
Searchterm 'refocusing pulse' was also found in the following service: 
spacer
Forum  (1)  
 
Refocused Gradient Echo SequenceInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.MRI Resource Directory:
 - Sequences -
 
Refocused GRE sequences use a refocusing gradient in the phase encoding direction during the end module to maximize (refocus) remaining xy- (transverse) magnetization at the time when the next excitation is due, while the other two gradients are, in any case, balanced.
When the next excitation pulse is sent into the system with an opposed phase, it tilts the magnetization in the α direction. As a result the z-magnetization is again partly tilted into the xy-plane, while the remaining xy-magnetization is tilted partly into the z-direction.
Companies use different acronyms to describe certain techniques.

Different terms for these gradient echo pulse sequences
R-GRE Refocused Gradient Echo,
FAST Fourier Acquired Steady State,
FFE Fast Field echo,
FISP Fast Imaging with Steady State Precession,
F-SHORT SHORT Repetition Technique Based on Free Induction Decay,
GFEC Gradient Field Echo with Contrast,
GRASS Gradient Recalled Acquisition in Steady State,
ROAST Resonant Offset Averaging in the Steady State,
SSFP Steady State Free Precession.
STERF Steady State Technique with Refocused FID

In this context, 'contrast' refers to the pulse sequence, it does not mean enhancement with a contrast agent.
spacer

• View the DATABASE results for 'Refocused Gradient Echo Sequence' (9).Open this link in a new window

MRI Resources 
MRI Training Courses - Jobs - Examinations - Implant and Prosthesis pool - Coils - Used and Refurbished MRI Equipment
 
Phase Cycling
 
Techniques of signal excitation in which the phases of the exciting or refocusing RF pulses are systematically varied and the resulting signals are then suitably combined in order to reduce or eliminate certain artifacts.
spacer
MRI Resources 
Crystallography - MR Guided Interventions - Collections - NMR - Software - Liver Imaging
 
previous      16 - 20 (of 21)     next
Result Pages : [1 2]  [3 4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 23 November 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]