Wow this is a long subject, First you have to under stand how the 3 main gradients are used. (slice select, Freq, then phase)
So you are use 1 of the gradients for your phase encoding. You need to fill the k-space which depends on you matrix. Lets assume you are using 256 (phase). In order to fill K-space you have to have 256 aquistions of the signal. Meaning for each picture (or slice) the signal must be acquire(sampled) 256 times each at a differnt amplitude to put it in a different Line of k-space.
Example- If you are performing a SE(spin echo) you have 1 TR period and 1 TE period (TE=is when the signal is sampled) so you will have to do the TR period 256 times to fill K-space with diffent amplitudes.
There is alot more that I left out. If this did not answer your question let me know
|
|